Brand Name: Alcon Laboratories, Inc recallZ-0305-2026 · Alcon Research LLC · initiated Aug 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alcon Research LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to incomplete seals in the pouch which provide the sterile barrier.
FDA root cause category: Process control
What the company is doing
On 08/05/2025, the firm sent via FedEx Overnight an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that for the Alcon Standalone Vitrectomy Consumables, there is potential for some pouches within impacted lots to have an incomplete seal. Customers are instructed to: locate and dispose of any affected lots of Alcon Standalone Vitrectomy Consumables remaining in your inventory. To comply with this Medical Device Recall and request the replacement of any unused product, please take the following steps: 1. Review inventory to determine if there are any unused affected product within their facility. See table on page 1 for affected Standalone Vitrectomy Consumable lots shipped to your location. 2. Segregate and dispose of any unused affected product from their inventory. 3. Call Alcon Customer Service to arrange for replacement of the affected inventory of Alcon Standalone Vitrectomy Consumables. 4. Respond to Alcon indicating understanding of these instructions even if they have zero (0) units remaining in inventory by completing the online response form at https://qrco.de/Vit-Seal or by completing and returning the attached Response Form and returning to Alcon via email: [email protected] or fax: 817-302-4337. 5. Forward this notification to all departments within their organization who may be in possession of this affected product; and any other organization to which this product may have been transferred. For questions contact, Alcon Customer Service at 1-800-862-5266 or an Alcon Sales Representative.
Recall details
- Recall number
- Z-0305-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 5, 2025
- Posted by the FDA
- Oct 24, 2025
- Recalling firm
- Alcon Research LLC
- Quantity
- 25,778
- Product code
- HQC
- FDA submissions
- K141065, K233876, K242184
- Distribution
- Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,
Product description
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065750918 PAK,I/A LINE,NGVS 8065750957 ASSY,SHIP,CONSTELLATION VFC 8065750972 ENDOILLUMATOR,RFID,23G 8065751014 CONST AUTO GAS FILL PAK 8065751014 CONST AUTO GAS FILL PAK 8065751441 ENDOILLUMINATOR,RFID,25G 8065751577 CHANDELIER,25GA,W/RFID Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Catalog Number: 8065750918 UDI-DI code:00380657509188 Lot/Batch Number: 1790L4 Catalog Number: 8065750957 UDI-DI code: 00380657509577 Lot/Batch Numbers: 1788D3 1788DC 1788D1 1790T0 179LNP 179LNT 179LNV 179LP2 179LPA 17A0P8 17A0PC 17A0PH 17C125 17CJEW 17CXT3 17D0RY 17DL3Y 17DL40 17DL41 17DL42 17DL46 17DPRT 17DPRU 17ERNK 17F2NR 17F55K Catalog Number: 8065750972 UDI-DI code: 00380657509720 Lot/Batch Numbers: 17C8T2 17C8T4 17E0TX Catalog Number: 8065751014 UDI-DI code: 00380657510146 Lot/Batch Numbers: 1789PV 1789PW 1788DD 1788DE 17C2MY 17C2NE 17DL3P 17DL3U 17DL3W 17DL3X 17E0WU 17E0X0 17F4D6 Catalog Number: 8065751441 UDI-DI code: 00380657514410 Lot/Batch Numbers: 17C6U5 17C359 17C35A Catalog Number: 8065751577 UDI-DI code: 00380657515776 Lot/Batch Numbers: 1789P4 1789P5 1789P6 179MCD 17C2RV 17C2RY 17C2T3 17DPN7 17DPN8
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Constellation 8065750957Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall138 reports
- Engauge 8065751577Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall14 reports
- Constellation 8065750918Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall12 reports
- Alcon 8065751014Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall11 reports
- Engauge 8065750972Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall1 report
- Engauge 8065751441Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97415).
- Z-0304-2026 · Open, ClassifiedBrand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product…
- Z-0306-2026 · Open, ClassifiedBrand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product…
- Z-0307-2026 · Open, ClassifiedBrand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product…
- Z-0308-2026 · Open, ClassifiedBrand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page