Skip to content

Alcon 8065751014Recalls & reported problems

Alcon Laboratories, Inc. · Auto Gas Fill Pak · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00380657510146), received between Sep 11, 2020 and Aug 8, 2024 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    Due to incomplete seals in the pouch which provide the sterile barrier.

    Z-0305-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Disconnection2
  • Detachment of Device or Device Component1
  • Device Markings/Labelling Problem1
  • Fitting Problem1
  • Gas/Air Leak1
  • Inability to Auto-Fill1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2024Radiofrequency Interference (RFI); Radiofrequency Interference (RFI)

    A CUSTOMER REPORTED THAT DURING SURGERY, THE AUTO GAS FILED DEVICE WAS EJECTED FROM OPHTHALMIC SYSTEM WITH SUFFICIENT FORCE TO IMPACT THE SURGICAL TECHNICIAN STANDING INFRONT OF THE SYSTEM. THE TECHNICIAN HAD MILD TO MODERATE SORE FOR FEW DAYS DUE TO THE IMPACT OF GAS EXPULSE FROM THE SYSTEM. THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT IMPACT AND PROCEDURE DETAILS WERE NOT REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19936108 ↗
  • MalfunctionApr 13, 2024Material Separation; Material Separation

    A PHYSICIAN REPORTED THAT THE TUBE CAME OFF DUE TO PRESSURE OF GAS DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY USING ANOTHER PRODUCT FROM DIFFERENT LOT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19102581 ↗
  • MalfunctionMar 8, 2024Disconnection; Fitting Problem; Disconnection; Fitting Problem

    BASED ON THE INFORMATION RECEIVED FOLLOWING SUBMISSION OF THE INITIAL REPORT, THIS EVENT ¿TUBE FOUND DISCONNECTED, SYRINGE PART WAS TOUCHING A STOPPER¿ DURING SURGERY DOES NOT MEET CRITERIA FOR REPORTING AS A REPORTABLE MALFUNCTION SINCE IT WAS ONLY ISSUE WITH AUTO GAS FILLING NOT WITH THE CASSETTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18866300 ↗
  • MalfunctionNov 15, 2023Suction Problem; Gas/Air Leak; Suction Problem; Gas/Air Leak

    A PHYSICIAN REPORTED THAT ASPIRATION FAILURE AND SUCTION ISSUE OCCURRED DURING SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE PROCEDURE TYPE WAS UNKNOWN. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18137786 ↗
  • MalfunctionOct 16, 2023No Flow; No Flow

    A CUSTOMER REPORTED THAT THE FLUID AIR EXCHANGE WOULD FILL WITH GAS BUT WOULD NOT EXCHANGE DURING SURGERY. THE PROCEDURE AND PATIENT HARM DETAILS WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17940053 ↗
  • MalfunctionOct 7, 2022Disconnection; Device Markings/Labelling Problem; Disconnection; Device Markings/Labelling Problem

    A PHYSICIAN REPORTED CONNECTING PART CAME OFF SEVERAL TIMES WHEN SUCKING GAS FROM THE CONSTELLATION DURING A SURGERY. A STICKER WITH SF6/C3F8 STATED WAS NOT INCLUDED. THE SURGERY TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER THE PRODUCT WAS REPLACED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15561832 ↗
  • MalfunctionMay 31, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    A PHYSICIAN REPORTED TUBE CAME OFF ON ITS OWN AT THE TIME OF THE PURGE DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER THE PRODUCT WAS REPLACED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14545948 ↗
  • MalfunctionFeb 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A CUSTOMER REPORTED THE CASSETTE LEAKED. THE PROCEDURE AND PATIENT IMPACT DETAILS WERE NOT PROVIDED. THERE WAS NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REPRESENTATIVE WHO INDICATED THE MALFUNCTION OCCURRED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER THE PRODUCT WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13591560 ↗
  • MalfunctionFeb 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A CUSTOMER REPORTED THE CASSETTE LEAKED. THE PROCEDURE AND PATIENT IMPACT DETAILS WERE NOT PROVIDED. THERE WAS NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REPRESENTATIVE WHO INDICATED THE MALFUNCTION OCCURRED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER THE PRODUCT WAS REPLACED..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13591530 ↗
  • MalfunctionFeb 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    A CUSTOMER REPORTED THE CASSETTES LEAKED. THE PROCEDURE AND PATIENT IMPACT DETAILS WERE NOT PROVIDED. THERE WAS NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REPRESENTATIVE WHO INDICATED THE MALFUNCTION OCCURRED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER THE PRODUCT WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13591140 ↗
  • MalfunctionSep 11, 2020Inability to Auto-Fill

    THE SYRINGE (AUTO GAS FILL PAK) WAS HOOKED UP TO THE CONSTELLATION SYSTEM AND "AUTO FILL" BUTTON WAS PRESSED. THE SYRINGE DID NOT FILL AS ANTICIPATED. A NEW PAK WAS OPENED AND USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10520718 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alcon 8065751014

Has the Alcon 8065751014 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0305-2026 (Class II), began on Aug 5, 2025: Due to incomplete seals in the pouch which provide the sterile barrier. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Alcon 8065751014?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Sep 11, 2020 and Aug 8, 2024. In those reports, the most common specific device problems are “Fluid/Blood Leak” (3), “Disconnection” (2) and “Detachment of Device or Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alcon 8065751014?

None of the reports are classified as deaths, 1 as an injury and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alcon 8065751014, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Alcon, model 8065751014). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page