Constellation 8065750957Recalls & reported problems

Alcon Laboratories, Inc. · Viscous Fluid Control Pak · Class 2

All models of Constellation (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Alcon Constellation viscous fluid control pak is a unit used in eye surgery. Many of the 134 reports describe the device failing to inject or extract oil during surgery, with tubing blocked or connectors popping open, and surgeons switching to a new product. There are no deaths and 4 injuries. There is 1 open recall, started in August 2025, because some pouches may have incomplete seals that keep the product sterile.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
120
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 120 adverse event reports that list this exact device (DI 00380657509577), received between May 7, 2021 and Jul 28, 2026 — classified as 4 injuries, 116 malfunctions.

The most frequently reported specific device problem is “Defective Device” (49 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 4 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 7 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 4 reportsSep 2024: 8 reportsOct 2024: 6 reportsNov 2024: 5 reportsDec 2024: 4 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 6 reportsAug 2025: 2 reportsSep 2025: 6 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 6 reportsFeb 2026: 2 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    Due to incomplete seals in the pouch which provide the sterile barrier.

    Z-0305-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Device49
  • Defective Component9
  • Suction Failure8
  • Fitting Problem7
  • Detachment of Device or Device Component5
  • Complete Blockage3
  • Disconnection3
  • Gas/Air Leak3

Patient problems

  • No Clinical Signs, Symptoms or Conditions96
  • Insufficient Information22
  • Bradycardia1
  • Endophthalmitis1
  • Intraocular Pressure Increased1
  • Retinal Tear1
  • Vitreous Hemorrhage1

Recent adverse event reports

The 25 most recent of 120 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2026Defective Device

    A physician reported that tubing was blocked preventing silicon oil injection during vitrectomy surgery. The surgery was successfully completed on same day by replacing it with new product. There was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50050625 ↗
  • MalfunctionJul 27, 2026Defective Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50016600 ↗
  • MalfunctionMar 6, 2026Defective Device

    A HEALTH CARE PROFESSIONAL REPORTED THAT THE INJECTION CANNULA WAS NOT CLAMPED CAUSING SILICONE LOSS DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH SAME PRODUCT. THERE WAS NO INFORMATION ABOUT ANY IMPACT ON PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24531676 ↗
  • MalfunctionMar 5, 2026Defective Component; Obstruction of Flow

    A CUSTOMER REPORTED THAT THE INFUSION CANNULA WAS DAMAGED AT THE TIP AND OBSTRUCTED AT AN UNKNOWN TIMING. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY COMPLETION DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. ADDITIONAL INFORMATION WAS REQUESTED AND NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 24516258 ↗
  • MalfunctionFeb 19, 2026Defective Device

    A NURSE REPORTED THAT VISCUS FLUID CONTROL DID NOT TO BE INJECTING PROPERLY AND EXTRACTION WAS NOT WORKING NORMALLY ON THE CONSOLE DURING SURGERY. THE PROCEDURE TYPE AND SURGERY DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24386548 ↗
  • MalfunctionFeb 5, 2026Defective Device

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THERE WAS AN ISSUE OCCURRED AT THE END OF THE SILICONE OIL TUBING, COULD NOT INJECT SILICONE OIL PROPERLY DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24269099 ↗
  • MalfunctionJan 21, 2026Defective Device

    A NURSE REPORTED THAT THE TUBING LEAKED OCCURRED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24141768 ↗
  • MalfunctionJan 21, 2026Defective Device

    A PHYSICIAN REPORTED THAT WHEN PREPARING TO INJECT SILICONE OIL DURING VITRECTOMY SURGERY, IT WAS FOUND THAT THE NEEDLE COVER OF THE TUBING WAS NOT TIGHT ENOUGH. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24132922 ↗
  • MalfunctionJan 14, 2026Defective Device

    A NURSE REPORTED THAT SILICON OIL TUBING RUPTURED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING WITH NEW PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064823 ↗
  • MalfunctionJan 14, 2026Defective Device

    A NURSE REPORTED THAT SILICON OIL TUBING RUPTURED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING WITH NEW PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064818 ↗
  • MalfunctionJan 13, 2026Defective Device

    A PHYSICIAN REPORTED THAT THE SYSTEM PROMPTED, THE TUBING WAS NOT CONNECTED WHEN TRYING TO EXTRACT OIL, THE OIL EXTRACTION OPERATION CANNOT BE COMPLETED DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052159 ↗
  • MalfunctionJan 9, 2026Defective Device

    A NURSE REPORTED THAT CATHETER LACKS A SILICONE SOFT TIP, MAKING IT UNSUITABLE FOR OIL EXTRACTION DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24018309 ↗
  • MalfunctionDec 30, 2025Defective Component; Defective Component

    A NURSE REPORTED THAT LEAKAGE OF GAS DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23931146 ↗
  • MalfunctionDec 29, 2025Defective Device; Defective Device

    A PHYSICIAN REPORTED THAT TUBING HAD AIR LEAK DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905152 ↗
  • MalfunctionDec 15, 2025Defective Device; Defective Device

    A HEALTH CARE PROFESSIONAL REPORTED THAT WHEN INJECTING SILICONE OIL DURING VITRECTOMY AND SILICONE OIL REPLACEMENT SURGERY, NEEDLE WAS CONNECTED AND INJECTION BEGAN, BUT AFTER APPROXIMATELY 5 MINUTES, THE NEEDLE WAS RELEASED FROM THE SYRINGE. IT WAS THEN TIGHTENED AND RELEASED AGAIN. NEEDLE WAS THEN RECONNECTED AND AGAIN DISCONNECTED. IT WAS OBSERVED THAT WHEN THE INJECTION BEGAN, THE NEEDLE ROTATED AND THEN RELEASED. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23800583 ↗
  • InjuryNov 25, 2025Defective Component; Defective Component

    A PHARMACIST REPORTED THAT DURING THE POSTERIOR APPROACH VITRECTOMY WITH RETINOPEXY AND POSTERIOR VITRECTOMY WITH SILICONE OR GAS INSERTION ALONG WITH CHORIO-RETINAL LESION ABLATION SURGERY IN THE RIGHT EYE OF AN ADULT PATIENT, THE FINAL INJECTION OF THE SILICONE CAUSED A SUDDEN INCREASE IN INTRAOCULAR PRESSURE, WITH HIGH-IMPACT EXPULSION OF THE SYSTEM'S SYRINGE INTO THE EYE, GENERATING TRANSIENT BRADYCARDIA, EXTENSIVE RETINAL RUPTURE, AND PRERETINAL MACULAR HEMORRHAGE. THE PATIENT HAD A HISTORY OF CONGENITAL CATARACT WITH LENS EXTRACTION SURGERY PLUS LENS IMPLANTATION. THIS HAD COMPROMISED THE PATIENT'S VISUAL INTEGRITY, WITH PROBABLE IRREVERSIBLE SEQUELAE. THE PROCEDURE WAS COMPLETED ON SAME DAY BUT THE PATIENT HAD TO UNDERGO MEDICAL INTERVENTION (TREATMENT WITH ANTIBIOTIC EYE DROP, CORTICOSTEROID AND NON STEROIDAL ANTI INFLAMMATORY EYE DROPS) AS WELL AS HOSPITALIZATION. SURGICAL INTERVENTION WAS ALSO REQUIRED DURING THE SAME PROCEDURE, AND A NEW INTERVENTION WILL BE NECESSARY. THE FINAL DIAGNOSIS WAS RETINAL DETACHMENT WITHOUT MACULAR INVOLVEMENT. THE PATIENT WAS UNDER OBSERVATION WITH SUGGESTION TO MAINTAIN SUPINE (FACE UP POSITION).

    Patient problems: Bradycardia; Intraocular Pressure Increased; Retinal Tear; Vitreous Hemorrhage

    MAUDE report 23642805 ↗
  • MalfunctionNov 11, 2025Defective Device; Defective Device

    A NURSE REPORTED THAT INFUSION TIP WAS FRACTURED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23529173 ↗
  • MalfunctionNov 7, 2025Defective Device; Defective Device

    THE PHYSICIAN REPORTED THAT PRODUCT COULD NOT BE CONNECTED TO THE CONNECTION PART DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494541 ↗
  • MalfunctionOct 27, 2025Defective Device; Defective Device

    A CUSTOMER REPORTED THAT DURING SURGERY, THE PATIENT WAS UNDER AIR AND SURGEON WAS TRYING TO PERFORM OIL INJECTION. THE PROCEDURE TYPE WAS UNKNOWN. AS TREADLE WAS DEPRESSED, THE OPHTHALMIC SYSTEM WOULD CUT OFF VISCOUS FLUIDICS CONTROL PRESSURE AND DISPLAYED SYSTEM MESSAGE. CHANGING SURGEON SETTINGS OR CASSETTE DID NOT GET RID OF THE SYSTEM MESSAGE. THE SURGERY WAS COMPLETED ON SAME DAY WITH AN ALTERNATIVE SYSTEM AND PROCEED WITH FINISHING THE OIL INJECTION. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. THIS REPORT PERTAINING TO TWO OF THE TWO REPORTS FROM THE SAME FACILITY.

    Patient problems: Insufficient Information

    MAUDE report 23389167 ↗
  • MalfunctionOct 23, 2025Defective Device; Defective Device

    A NURSE REPORTED THAT THE OIL INJECTION NEEDLE COULD NOT BE PROPERLY CONNECTED TO THE SILICONE OIL SYRINGE-IT CANNOT BE TIGHTENED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23365735 ↗
  • MalfunctionSep 19, 2025Defective Device; Defective Device

    A PHYSICIAN REPORTED THAT PRODUCT COULD NOT BE REMOVED WHEN CONNECTED DURING CATARACT AND VITRECTOMY COMBINATION SURGERY. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095161 ↗
  • MalfunctionSep 17, 2025Defective Device; Defective Device

    A PHYSICIAN REPORTED THAT THE ASPIRATION COULD NOT BE PERFORMED WHEN TRIED TO REMOVE OIL DURING CATARACT AND VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE INVOLVED EYE AND THE PATIENT IDENTIFIER ARE NOT AVAILABLE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23070326 ↗
  • MalfunctionSep 10, 2025Defective Device; Defective Device

    A PHYSICIAN REPORTED THAT A PRODUCT COULD NOT BE USED BECAUSE OF ITS DEFECTIVENESS DURING CATARACT/VITRECTOMY COMBINATION SURGERY. THE SURGERY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23013760 ↗
  • InjurySep 4, 2025Component Missing; Defective Device; Component Missing; Defective Device

    A NURSE REPORTED THAT APPROXIMATELY 50 PERCENT OF THE PAKS OF THIS LOT WERE MISSING THE RUBBER LIP ON THE BLUE TIP USED FOR OIL REMOVAL. THE NURSE ALSO REPORTED THAT TWO PATIENTS EXPERIENCED IRRITATION UP TO AND INCLUDING ENDOPHTHALMITIS FOLLOWING OIL INSERTION. THE TWO PATIENTS¿ OUTCOME AND INFORMATION RELATED TO THE DEFECT WITH THE OTHER PAKS OF THIS LOT HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED ON 29-AUG-2025 STATED THAT THIS ISSUE OCCURRED ABOUT 10 TIMES WITH THIS LOT. WITH THE OTHER OCCURRENCES THERE WAS NO PATIENT IMPACT; THEY DISCARDED THE PAK AND USED ANOTHER ONE WHEN THEY NOTICED THAT THE RUBBER LIP WAS MISSING. THIS REPORT IS FOR THE INCIDENTS INVOLVING THE TWO PATIENTS WHO EXPERIENCED IRRITATION UP TO AND INCLUDING ENDOPHTHALMITIS. IF FURTHER DIFFERENTIATING INFORMATION IS LATER RECEIVED, SEPARATE REPORTS WILL BE SUBMITTED, AS APPROPRIATE.

    Patient problems: Endophthalmitis

    MAUDE report 22964540 ↗
  • MalfunctionSep 3, 2025Defective Device; Defective Device

    A NURSE REPORTED THAT TUBING LEAKED AIR DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22959949 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Constellation 8065750957

Has the Constellation 8065750957 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0305-2026 (Class II), began on Aug 5, 2025: Due to incomplete seals in the pouch which provide the sterile barrier. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Constellation 8065750957?

The FDA's MAUDE database lists 120 adverse event reports for this device, received between May 7, 2021 and Jul 28, 2026. In those reports, the most common specific device problems are “Defective Device” (49), “Defective Component” (9) and “Suction Failure” (8); the most common patient problems are “Bradycardia” (1), “Endophthalmitis” (1) and “Intraocular Pressure Increased” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Constellation 8065750957?

None of the reports are classified as deaths, 4 as injuries and 116 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Constellation 8065750957, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Constellation, model 8065750957). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →