Fuse ULTRA Foot Plating System Instructions for Use recallZ-0230-2026 · CPM Medical Consultants, LLC. · initiated Sep 17, 2025
Class II Open, Classified All recalls posted in October 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CPM Medical Consultants, LLC. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.
FDA root cause category: Labeling design
What the company is doing
On 9/17/2025, correction notices were mailed to customers and distributors who were informed of the following: - The only sterilization requirement now shown on the new Rev D IFU is 4 mins exposure time at 132-Degrees C (270-Degrees F) and 20-30mins drying time. - All affected versions of the IFU should be immediately discarded. - You can access the e-IFU at: https://fusemedical.com/wp-content/uploads/2025/09/Fuse-ULTRA-Foot-Plating-System-IFU-Rev-D-9.2025.pdf - Please pass on this Medical Device Correction to all users of the products and all other people who need to be aware within your organization. - If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form, and updated IFU. - Complete and return the response form via email to [email protected] If you have any questions, contact the firm at [email protected]
Recall details
- Recall number
- Z-0230-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 17, 2025
- Posted by the FDA
- Oct 21, 2025
- Recalling firm
- CPM Medical Consultants, LLC.
- Quantity
- 3317
- Product code
- HRS
- FDA submissions
- K141784
- Distribution
- US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.
Product description
Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS200326, OS328026, OS328028C, OS216150, OS218150, OS292110, OS292120, OS292160, OS292200, OS200516, OS200518, OS200520, OS200522, OS200616, OS200618, OS200620, OS200622, OS421300, OS421302, OS421304, OS421306, OS421308, OS421208, OS421210, OS421212, OS421000L, OS421000R, OS421002L, OS421002R, OS421004L, OS421004R, OS421006L, OS421006R, OS42260LL, OS42260LR, OS42260ML, OS42260MR, OS42260SL, OS42260SR, OS42260XSL, OS42260XSR, OS422800, OS422803, OS422804, OS422805, OS422807, OS422825, OS422400, OS422403, OS422404, OS422405, OS422406, OS422407, OS421406, OS421408, OS421414, OS422702, OS422703, OS422712, OS422714, OS422716, OS421512, OS421514, OS421516, OS421112, OS421116, OS421120, OS421124, OS421130, OS42150L, OS42150M, OS42150S, OS42150XS, OS326920-T, OS326922-T, OS326924-T, OS326925, OS326925-T, OS326926-T, OS326928, OS326928-T, OS326930, OS326930-T, OS326932, OS326932-T, OS326934, OS326934-T, OS326936, OS326936-T, OS326938, OS326938-T, OS326940, OS326940-T, OS326945, OS326945-T, OS326950, OS326950-T, OS326955, OS326955-T, OS326960, OS326960-T, OS327920, OS327922, OS327924, OS327926, OS327928, OS327930, OS327932, OS327934, OS327936, OS327938, OS327940, OS327945, OS327950, OS327955, OS327960, OS421710, OS421712, OS421714, OS421716, OS421718, OS421720, OS421722, OS421724, OS421726, OS421728, OS421730, OS421735, OS421740, OS421745, OS421750, OS422508, OS422510, OS422512, OS422514, OS422516, OS422518, OS422520, OS422522, OS422524, OS422526, OS422528, OS422530, OS421814, OS421816, OS421818, OS421820, OS421822, OS421824, OS421826, OS421828, OS421830, OS421835, OS421840, OS421845, OS421850, OS422608, OS422610, OS422612, OS422614, OS422616, OS422618, OS422620, OS422622, OS422624, OS422626, OS422628, OS422630, R327924, R327928, R327932, R327936, R327940, R327945, R327950, R421000L, R421000R, R421710, R421712, R421714, R421716, R421718, R421720, R421722, R421724, R421726, R421728, R421730, R421735, R421740, R421814, R421816, R421818, R421820, R421822, R421824, R421826, R421828, R421830, R421835, R421840, R421845, R421850, R42260ML, R42260MR, R42260SL, R42260SR, R42260XSL, R42260XSR
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
All versions of IFU DC-0495 prior to Rev D. All Lots Up to Date 2025/09/10. UDI-DI: B565OS200020L, B565OS200025L, B565OS200120L, B565OS200125, B565OS200125L, B565OS200135, B565OS200324, B565OS200325, B565OS200326, B565OS200400, B565OS200401, B565OS200516, B565OS200518, B565OS200520, B565OS200522, B565OS200616, B565OS200618, B565OS200620, B565OS200622, B565OS216150, B565OS218150, B565OS292110, B565OS292120, B565OS292160, B565OS292200, B565OS326920T, B565OS326922T, B565OS326924T, B565OS326925, B565OS326925T, B565OS326926T, B565OS326928, B565OS326928T, B565OS326930, B565OS326930T, B565OS326932, B565OS326932T, B565OS326934, B565OS326934T, B565OS326936, B565OS326936T, B565OS326938, B565OS326938T, B565OS326940, B565OS326940T, B565OS326945, B565OS326945T, B565OS326950, B565OS326950T, B565OS326955, B565OS326955T, B565OS326960, B565OS326960T, B565OS327920, B565OS327922,B565OS327924, B565OS327926, B565OS327928, B565OS327930, B565OS327932, B565OS327934, B565OS327936, B565OS327938, B565OS327940, B565OS327945, B565OS327950, B565OS327955, B565OS327960, B565OS328017, B565OS328026, B565OS328028C, B565OS421000L, B565OS421000R, B565OS421002L, B565OS421002R, B565OS421004L, B565OS421004R, B565OS421006L, B565OS421006R, B565OS421112, B565OS421116, B565OS421120, B565OS421124, B565OS421130, B565OS421208, B565OS421210, B565OS421212, B565OS421300, B565OS421302, B565OS421304, B565OS421306, B565OS421308, B565OS421406, B565OS421408, B565OS421414, B565OS42150L, B565OS42150M, B565OS42150S, B565OS42150XS, B565OS421512, B565OS421514, B565OS421516, B565OS421710, B565OS421712, B565OS421714, B565OS421716, B565OS421718, B565OS421720, B565OS421722, B565OS421724, B565OS421726, B565OS421728, B565OS421730, B565OS421735, B565OS421740, B565OS421745, B565OS421750, B565OS421814, B565OS421816, B565OS421818, B565OS421820, B565OS421822, B565OS421824, B565OS421826, B565OS421828, B565OS421830, B565OS421835, B565OS421840, B565OS421845, B565OS421850, B565OS422400, B565OS422403, B565OS422404, B565OS422405, B565OS422406, B565OS422407, B565OS422508, B565OS422510, B565OS422512, B565OS422514, B565OS422516, B565OS422518, B565OS422520, B565OS422522, B565OS422524, B565OS422526, B565OS422528, B565OS422530, B565OS422608, B565OS42260LL, B565OS42260LR, B565OS42260ML, B565OS42260MR, B565OS42260SL, B565OS42260SR, B565OS42260XSL, B565OS42260XSR, B565OS422610, B565OS422612, B565OS422614, B565OS422616, B565OS422618, B565OS422620, B565OS422622, B565OS422624, B565OS422626, B565OS422628, B565OS422630, B565OS422702, B565OS422703, B565OS422712, B565OS422714, B565OS422716, B565OS42280, B565OS422803, B565OS422804, B565OS422805, B565OS422807, B565OS422825, B565R327924, B565R327928, B565R327932, B565R327936, B565R327940, B565R327945, B565R327950, B565R421000L, B565R421000R, B565R421710, B565R421712, B565R421714, B565R421716, B565R421718, B565R421720, B565R421722, B565R421724, B565R421726, B565R421728, B565R421730, B565R421735, B565R421740, B565R421814, B565R421816, B565R421818, B565R421820, B565R421822, B565R421824, B565R421826, B565R421828, B565R421830, B565R421835, B565R421840, B565R421845, B565R421850, B565R42260M, B565R42260MR, B565R42260SL, B565R42260SR, B565R42260XSL, B565R42260XSR
Other devices under the same FDA submission
5 devices cleared under K141784 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS200125Ortho Solutions Holdings Limited · Bit, Drill4 reports
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS292200Ortho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS200025LOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS422702-NSOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS422716-NSOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗