IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens… recallZ-0163-2026 · IMPRIS Imaging Inc · initiated Sep 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from IMPRIS Imaging Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
FDA root cause category: Under Investigation by firm
What the company is doing
Imris issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 9/30/2025 via email. The notice explained the issue, the risk to health, and provided the following information: "If the error message, "Magnet Supervision: The average heater power is above the alarm threshold. Please contact Siemens Service." appears, DO NOT REBOOT THE SYSTEM. Please call your local Siemens Healthineers service organization. Grant access to the system so that the inspection can be carried out as directed by Siemens Healthineers service organization." "Please immediately instruct your personnel accordingly, and please contact our customer support team via our service line at 1-866-475-0525 or e-mail at [email protected] if you have any questions or concerns. Please keep this information until the corrective measures have been finalized." This notice included a copy of the Siemens Healthineers Urgent Medical Device Correction notice.
Recall details
- Recall number
- Z-0163-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2025
- Posted by the FDA
- Oct 28, 2025
- Recalling firm
- IMPRIS Imaging Inc
- Product code
- LNH
- FDA submissions
- K083137, K101813, K133692, K212367
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Product description
IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
Code information (models, lots, serial numbers)
IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IMRIS iMRI 3T V 120823-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall6 reports
Other devices under the same FDA submission
7 devices cleared under K083137, K101813, K133692, K212367 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- iMRI 3T S OR-DR, 116118-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall5 reports
- iMRI 3T S OR-DR-OR 116053-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall2 reports
- iMRI 3T S OR-DR-OR 116117-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
- iMRI 3T S OR-DR 116119-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
- Neuro III-SV iMRI SystemIMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
- Neuro III-SV iMRI SystemIMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
- InVision™ 3T Operating Suite 120823-000IMRIS Imaging, Inc. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97750).
- Z-0161-2026 · Open, ClassifiedIMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio
- Z-0162-2026 · Open, ClassifiedIMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
- Z-0164-2026 · Open, ClassifiedIMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page