Elekta Medical Linear Accelerator recallZ-1164-2026 · Elekta, Inc. · initiated Dec 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a manufacturing issue and electrical grounding of systems.
FDA root cause category: Process control
What the company is doing
On 12/14/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" Letter to customer informing them that Elekta has identified a safety issue relating to the modulator discharge rod, where the supplier has been fitting the crimp in the crimp tool back to front. The issue was discovered during production. Note - The discharge rod is used by Field Service Engineers (FSE) when completing service activities on the machine around the HT and RF (High Tension (high voltage) and Radio Frequency) assembly. Instructions to customer: No actions are required to be taken by the Clinical user. Instructions for Field Service Engineers: Possible impact on the Field Service Engineer:- The FSE uses the HT discharge rod to fully discharge the voltage capacitor. Once the FSE has completed the procedure using the discharge rod, the FSE will start to work on the system around the HT and RF assembly. If the discharge rod is not fully connected to the ground or not connected at all, the system will still be fully charged and the FSE could receive an electric shock. Elekta will release an Important Field Safety Modification (IFSM) to identify potentially faulty discharge rod crimping and provide further instructions. The IFSM will include procedures for verifying the reliability of the discharge rod crimping and replacing affected components if necessary. local Elekta service representative will contact you to arrange for this activity to be completed.
Recall details
- Recall number
- Z-1164-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 14, 2025
- Posted by the FDA
- Jan 23, 2026
- Recalling firm
- Elekta, Inc.
- Quantity
- 5391
- Product code
- IYE
- FDA submissions
- K210500
- Distribution
- Worldwide
Product description
Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No
Code information (models, lots, serial numbers)
Lot Code: 05060191071550 (NFB) Elekta Synergy 05060191071505 (FB) 05060191071543 (NFB) Elekta Infinity 05060191071512 (FB) 05060191071574 (NFB) Versa HD 05060191071529 (FB) 05060191071581 Harmony 05060191071604 Harmony Pro 07340201503461 Elekta Evo
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- Elekta Medical Linear Accelerator 1552093Elekta Solutions AB · Accelerator, linear, medical5 recalls6 reports
- Elekta Medical Linear Accelerator 1552092Elekta Solutions AB · Accelerator, linear, medical5 recalls0 reports
- Elekta Medical Linear Accelerator 1550358Elekta Solutions AB · Accelerator, linear, medical4 recalls2 reports
- Elekta Medical Linear Accelerator 1552091Elekta Solutions AB · Accelerator, linear, medical4 recalls1 report
- Elekta Medical Linear Accelerator 1550359Elekta Solutions AB · Accelerator, linear, medical4 recalls0 reports
- Elekta Medical Linear Accelerator 1550357Elekta Solutions AB · Accelerator, linear, medical3 recalls1 report
- Elekta Medical Linear Accelerator 1553032Elekta Solutions AB · Accelerator, linear, medical2 recalls0 reports
- Elekta Medical Linear Accelerator 1555143Elekta Solutions AB · Accelerator, linear, medical2 recalls0 reports
- EMLA (Elekta EVO) 1571644Elekta Solutions AB · Accelerator, linear, medical1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page