Skip to content

CereLink recallZ-1037-2026 · Integra LifeSciences Corp. (NeuroSciences) · initiated Dec 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. (NeuroSciences) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

FDA root cause category: Process control

What the company is doing

Integra LifeSciences Corporation notified consignees on about 12/12/2025 via FedEx overnight mailing. Consignees were instructed to identify impacted part numbers and lot numbers on hand, remove them immediately from service and quarantine them, complete and return the Acknowledgement Form, and maintain the notice for records. Consignees were also instructed to forward the notification to those who utilize the product for awareness of the recall. Once the Acknowledgement Form is received, and noted that affected product are on hand, Customer Service will contact consignees and provide a Return Material Authorization (RMA) number and directions to return the affected product.

Recall details

Recall number
Z-1037-2026
Status
Open, Classified
Initiated by the company
Dec 12, 2025
Posted by the FDA
Jan 8, 2026
Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Quantity
5,116 units
Product code
GWM
FDA submissions
K173192
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).

Product description

Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Catalog Number: 826850. UDI-DI: 10381780520672. Lot Numbers: 7261958, 7266691, 7266692, 7288060, 7293837, 7293836, 7306040, 7306041, 7306042, 7306043, 7306044, 7306045, 7306046, 7306047, 7306050, 7306051, 7306052, 7307670, 7307671, 7307672, 7325482, 7307675, 7325481, 7325483, 7325484, 7307673, 7307674, 7325485, 7325486, 7336611, 7336612, 7336614, 7336623, 7336613, 7336615, 7336616, 7336618, 7336619, 7336620, 7336621, 7336622, 7306048, 7306049, 7336626, 7336627, 7336629, 7347166, 7336628, 7346295, 7346297, 7346299, 7346301, 7365434, 7365435, 7392019, 7396607, 7396608, 7396609, 7396610, 7392034, 7392067, 7405790, 7405791, 7405792, 7405793, 7406633, 7406634, 7406636, 7406637, 7406639, 7406640, 7406641, 7406642, 7406635, 7406638, 7408813, 7408815, 7408818, 7408814, 7408816, 7468726, 7468727, 7411801, 7411802, 7414641, 7414642, 7468731, 7468732, 7468745, 7468746, 7468741, 7468742, 7468743, 7468744, 7468771, 7468769, 7468770, 7468773, 7468774, 7468775, 7482179, 7482180, 7482181, 7482182, 7482183, 7482184, 7482185, 7482186, 7468772, 7482154, 7482155, 7482156, 7482157, 7468801, 7468804, 7468805, 7468806, 7482136, 7468802, 7468803, 7482132, 7482134, 7482135, 7482137, 7482159, 7482158, 7482160, 7482161, 7495593, 7495594, 7495596, 7495597, 7495598, 7495599, 7482212, 7495600, 7495601, 7495602, 7495603, 7495604, 7495605, 7495606, 7495607, 7495608, 7495635, 7495636, 7495637, 7495638, 7495639, 7495640, 7495641, 7495642, 7495643, 7495666, 7495644, 7495664, 7482213, 7495667, 7503597, 7503598, 7503599, 7503659, 7510140, 7510141, 7510142, 7503600, 7503601, 7517716, 7517708, 7503627, 7503628, 7503630, 7503631, 7503632, 7503633, 7503629, 7503664, 7503657, 7503662, 7503663, 7503602, 7503624, 7510156, 7517703, 7517704, 7517706, 7517707, 7517709, 7517710, 7517711, 7517712, 7517714, 7517715, 7408817, 7411797, 7411798, 7411799, 7411800, 7414643, 7414644, 7414645, 7414646, 7414647, 7414648, 7414649, 7417938, 7417940, 7417942, 7417944, 7417946, 7417948, 7417950, 7417952, 7417955, 7417957, 7417959, 7503625.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K173192 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98091).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page