HIGH FLOW INSUFFLATION UNIT recallZ-1354-2026 · Olympus Corporation of the Americas · initiated Jan 16, 2026
Class I Open, Classified All recalls posted in February 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue with software algorithm which may lead to overpressure events.
FDA root cause category: Software design
What the company is doing
Olympus notified consignees on about 01/16/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to cease usage of the affected product immediately. They are advised to use available alternatives including an existing UHI-4 if available or switching to an insufflation device from another manufacturer. If no alternative is available on site, contact Olympus to arrange a loaner or purchase a replacement. Additionally, consignees were instructed to examine inventory and quarantine any identified devices, immediately cease usage of any UHI, UHI-2, and/or UHI-3 in inventory, call Olympus Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return your device. Consignees were also requested to acknowledge receipt of this letter through their recall web portal as follows: a. Go to https://olympusamerica.com/recall, b. Enter the recall number "0489 and c. Complete the form as instructed. Please indicate in your response if you no longer have the device(s). If any consignees have further distributed this UHI product, they were instructed to identify affected customers and forward this notice to them immediately. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by email at [email protected].
Recall details
- Recall number
- Z-1354-2026
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Jan 16, 2026
- Posted by the FDA
- Feb 18, 2026
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 485 units
- Product code
- HIF
- FDA submissions
- K014166
- Distribution
- Nationwide distribution
Product description
Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.
Code information (models, lots, serial numbers)
Model Number: UHI-3; UDI-DI: 04953170140280, 04953170140297; All Serial Numbers.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall162 reports
- HIGH FLOW INSUFFLATION UNIT UHI-3Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC1 recall48 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K014166 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CYLINDER HOSE (PIN) FOR UHI-3 MAJ-1080Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC58 reports
- INSUFFLATION TUBE MAJ-590Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC2 reports
- SUCTION TUBE MAJ-591Olympus Medical Systems Corp. · INSUFFLATOR, LAPAROSCOPIC2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page