Skip to content

HIGH FLOW INSUFFLATION UNITRecalls & reported problems across all models

Olympus Medical Systems Corp. · 4 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name at least one HIGH FLOW INSUFFLATION UNIT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2,703
Deaths reported
6
Injuries reported
61
Recalls
4

We found 4 models of HIGH FLOW INSUFFLATION UNIT by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 2,703 adverse event reports in the FDA's MAUDE database, received between Sep 18, 2018 and Aug 11, 2026 — classified as 6 deaths, 61 injuries, 2,636 malfunctions.

The most frequently reported specific device problem is “Pressure Problem” (316 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 85 reportsDec 2023: 83 reportsJan 2024: 89 reportsFeb 2024: 56 reportsMar 2024: 74 reportsApr 2024: 56 reportsMay 2024: 68 reportsJun 2024: 91 reportsJul 2024: 81 reportsAug 2024: 85 reportsSep 2024: 130 reportsOct 2024: 156 reportsNov 2024: 86 reportsDec 2024: 33 reportsJan 2025: 20 reportsFeb 2025: 17 reportsMar 2025: 12 reportsApr 2025: 8 reportsMay 2025: 6 reportsJun 2025: 6 reportsJul 2025: 8 reportsAug 2025: 20 reportsSep 2025: 10 reportsOct 2025: 29 reportsNov 2025: 25 reportsDec 2025: 6 reportsJan 2026: 8 reportsFeb 2026: 13 reportsMar 2026: 11 reportsApr 2026: 3 reportsMay 2026: 6 reportsJun 2026: 5 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026156
Last 36 months. The most recent month may be incomplete.

HIGH FLOW INSUFFLATION UNIT recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 16, 2026Class IIOpen, Classified

    Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.

    Z-2776-2026 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗

  2. Jan 16, 2026Class IOpen, Classified

    Issue with software algorithm which may lead to overpressure events.

    Z-1354-2026 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Sep 25, 2023Class IIOpen, Classified

    Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".

    Z-1375-2024 · Root cause: Process control · Recall details · FDA record ↗

  4. Sep 22, 2023Class IOpen, Classified

    There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.

    Z-0075-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported HIGH FLOW INSUFFLATION UNIT problems

Device problems

  • Pressure Problem316
  • No Display/Image249
  • Failure to Power Up169
  • Display or Visual Feedback Problem168
  • Gas/Air Leak165
  • No Visual Prompts/Feedback128
  • Electrical /Electronic Property Problem121
  • Loss of Power113

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,614
  • No Consequences Or Impact To Patient33
  • Cardiac Arrest13
  • Bradycardia7
  • No Patient Involvement5
  • Abdominal Distention4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Insufficient Information4

HIGH FLOW INSUFFLATION UNIT models

4 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent HIGH FLOW INSUFFLATION UNIT adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Olympus Medical Systems Corp. products

All Olympus Medical Systems Corp. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page