Skip to content

GEM Premier 5000 recallZ-1539-2026 · Instrumentation Laboratory · initiated Jan 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION notice was sent to consignees on 1/20/26 was sent to consignees. The notice instructs consignees to continue to use affected devices with the provided warmup instructions. Consignees are to inform all laboratory staff of this notification and forward the notification if product was further distributed. A copy of the notification is to be maintained in consignee records. Consignees may contact Werfen Technical Support at 1-800-678-0710 for a replacement if a GEM PAK experiences a Process Control Solution Not Detected error.

Recall details

Recall number
Z-1539-2026
Status
Open, Classified
Initiated by the company
Jan 20, 2026
Posted by the FDA
Mar 10, 2026
Recalling firm
Instrumentation Laboratory
Quantity
1,607 units
Product code
CHL
FDA submissions
K203790
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.

Product description

GEM Premier 5000; Model No. 00055430010.

Code information (models, lots, serial numbers)

Part number: 00055430010; UDI: 08426950807759; Lots No.: 251021K, 251027AA, 251110AH, 251114K, 251114O, 251114P, 251114Y, 251117AA, 251117AC, 251117AG, 251117AM, 251117AN, 251117P, 251117T, 251117V, 251117Y, 251118A, 251118B, 251118R, 251118X, 251119A, 251120R, 251120U, 251121Q, 251124O, 251124Q, 251125I, 251126S, 251126T, 251126U, 251201W, 251201X, 251202G, 251202H, 251203L, 251204R, 251205L, 251208N, 251211W, 251215D, 251215Z, 251216O, 251216S, 251217L, 251217V, 251218AB, 251218B, 251218X, 251218Y, 251219A, 251219O, 251222AE, 251222E, 251222F, 251222K, 251222M, 251222P, 251222Z, 251223A, 251223B, 251223C, 251223I, 251223M, 251223O, 251229D, 251229E, 251229I, 251229S, 251231B, 251231G.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

21 devices cleared under K203790 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98349).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page