GEM Premier 5000 recallZ-1543-2026 · Instrumentation Laboratory · initiated Jan 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notice was sent to consignees on 1/20/26 was sent to consignees. The notice instructs consignees to continue to use affected devices with the provided warmup instructions. Consignees are to inform all laboratory staff of this notification and forward the notification if product was further distributed. A copy of the notification is to be maintained in consignee records. Consignees may contact Werfen Technical Support at 1-800-678-0710 for a replacement if a GEM PAK experiences a Process Control Solution Not Detected error.
Recall details
- Recall number
- Z-1543-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 20, 2026
- Posted by the FDA
- Mar 10, 2026
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 1,989 units
- Product code
- CHL
- FDA submissions
- K203790
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.
Product description
GEM Premier 5000; Part No. 00055445010 & 00055445011.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part number: 00055445010; UDI: 08426950807810; Lots No.: 251023N, 251024J, 251114K, 251117A, 251117AC, 251117AE, 251117F, 251117I, 251117J, 251117Y, 251118A, 251119A, 251120U, 251121Q, 251121S, 251121T, 251121W, 251121X, 251124K, 251124O, 251124U, 251125I, 251126C, 251126D, 251202H, 251203L, 251203M, 251204R, 251204S, 251205K, 251205L, 251208J, 251208L, 251215D, 251215W, 251215X, 251216S, 251216V, 251217W, 251217X, 251218R, 251218Y, 251219B, 251219C, 251219E, 251219H, 251219I, 251219P, 251219Q, 251223B, 251223D, 251223F, 251223K, 251229AL, 251229D, 251229E, 251229F, 251229S, 251229U, 251230R, 251231B, 251231D, 251231E, 251231H, 251231M. Part number: 00055445011 UDI: 08426950807827 Kits distributed in US: 96 Lots No.: 251117L, 251118O, 251124K, 251125I, 251201X, 251202G, 251203M, 251204O, 251223K, 251229W, 251230A.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 450 TEST 00055445011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Other devices under the same FDA submission
20 devices cleared under K203790 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/COOX 75 TEST 00055407504Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 600 TEST 00055360010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 600 TEST 00055360011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 150 TEST 00055415010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 300 TEST 00055430011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/COOX 600 TEST 00055360008Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/COOX 75 TEST 00055407508Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 75 TEST 00055407510Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 75 TEST 00055407511Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/COOX 150 TEST 00055415004Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98349).
- Z-1524-2026 · Open, ClassifiedGEM Premier 5000 PAK, Part No. 00055360004.
- Z-1525-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055360008.
- Z-1526-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055360010.
- Z-1527-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055360011.
- Z-1528-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407504.
- Z-1529-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407508.
- Z-1530-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407510.
- Z-1531-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055407511.
- Z-1532-2026 · Open, ClassifiedGEM Premier 5000 PAK; Part No. 00055415004.
- Z-1533-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055415005.
- Z-1534-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055415008.
- Z-1535-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055415010.
- Z-1536-2026 · Open, ClassifiedGEM Premier 5000; Part No: 00055415011.
- Z-1537-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055430004.
- Z-1538-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055430008.
- Z-1539-2026 · Open, ClassifiedGEM Premier 5000; Model No. 00055430010.
- Z-1540-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055430011.
- Z-1541-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055445004.
- Z-1542-2026 · Open, ClassifiedGEM Premier 5000; Part No. 00055445008.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page