Skip to content

Brand Name: Paradise Ultrasound Renal Denervation System recallZ-1654-2026 · ReCor Medical Inc. · initiated Jan 29, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ReCor Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.

FDA root cause category: Under Investigation by firm

What the company is doing

On 01/28/2026, the firm informed customers verbally and sent via mail an "MEDICAL DEVICE RECALL" Letter that nonconforming products (cracking in the hub adjacent to the guidewire port) that may lead to fluid leaks and could trigger a pressure error. Customers are instructed to: 1. Do not use the affected catheters for any patient procedures. 2. Identify the catheters from the affected lot and return to Recor Clinical Specialist through RMA process. 3. Review and complete the Acknowledgement and Receipt Form and provide to the Recor Clinical Specialist. For questions or further assistance - contact Director of Regulatory & Quality Compliance at [email protected] or telephone at 650-542-7700

Recall details

Recall number
Z-1654-2026
Status
Open, Classified
Initiated by the company
Jan 29, 2026
Posted by the FDA
Mar 27, 2026
Recalling firm
ReCor Medical Inc.
Quantity
1
Product code
QYI
FDA submissions
P220023
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, KS, MO, NC, OH, TX, and VA.

Product description

Brand Name: Paradise Ultrasound Renal Denervation System Product Name: Paradise Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation

Code information (models, lots, serial numbers)

Lot Code: Model No PRDS-068-02; UDI-DI 00810008950043; Lot Number: M4907

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under P220023 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98365).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page