THYROID SHIELD recallZ-1753-2026 · Burlington Medical, LLC · initiated Feb 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Burlington Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for attenuation degradation over time, decreasing the lifespan.
FDA root cause category: Process change control
What the company is doing
Burlington Medical notified consignees on about 02/11/2026 via "URGENT - FIELD SAFETY NOTICE (FSN)" letter. Consignees were instructed to ensure all users of the specific Burlington Medical radiation protection garments made aware of this Recall Notice, review inventory for affected units, arrange for the return of those affected units for proper disposal, and complete and return the Recall Response Form. Consignees were also instructed to notify appropriate individuals within your organization and to any organization where the potentially affected devices were transferred or sold.
Recall details
- Recall number
- Z-1753-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 11, 2026
- Posted by the FDA
- Apr 2, 2026
- Recalling firm
- Burlington Medical, LLC
- Quantity
- 41,962 units
- Product code
- IWO
- Distribution
- Worldwide distribution. Nationwide distribution in the US and international distribution to Canada, Australia, Bangladesh, Chile, Hong Kong, India, Mexico, Malaysia, Panama, Singapore, Thailand, Taiwan, Uruguay, and Vietnam.
Product description
Burlington Medical, Thyroid Shield.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
All serial numbers are affected that were manufactured with Xenolite 800 attenuation material from 01/16/2025 to 01/21/2026. 1. Model Number: R8COPEH; UDI-DI (Product Code): 00840331299052 (COPE-R8COPEH-LG). 2. Model Number: R8TSBBB; UDI-DI (Product Code): 00840331260311 (THYROIDSHIELDO8-R8TSBBB-SM). 3. Model Number: R8TSBBN; UDI-DI: 00840331260328; Product Codes: THYROIDSHIELDO8-R8TSBBN-LG, THYROIDSHIELDO8-R8TSBBN-MD, THYROIDSHIELDO8-R8TSBBN-SM. 4. Model Number: R8TSBBR; UDI-DI (Product Code): 00840331260359 (THYROIDSHIELDO8-R8TSBBR-LG), 00840331260366 (THYROIDSHIELDO8-R8TSBBR-MD), 00840331260373 (THYROIDSHIELDO8-R8TSBBR-SM). 5. Model Number: R8TSBBT; UDI-DI (Product Code): 00840331260397 (THYROIDSHIELDO8-R8TSBBT-MD), 00840331260403 (THYROIDSHIELDO8-R8TSBBT-SM). 6. Model Number: R8TSBHB; UDI-DI (Product Code): 00840331295948 (THYROIDSHIELDO8-R8TSBHB-MD), 00840331295955 (THYROIDSHIELDO8-R8TSBHB-SM). 7. Model Number: R8TSBHN; UDI-DI (Product Code): 00840331260427 (THYROIDSHIELDO8-R8TSBHN-MD ), 00840331260434 (THYROIDSHIELDO8-R8TSBHN-SM ). 8. Model Number: R8TSBHR; UDI-DI (Product Code): 00840331260465 (THYROIDSHIELDO8-R8TSBHR-SM ), 00840331260458 (THYROIDSHIELDO8-R8TSBHR-MD ), 00840331260441 (THYROIDSHIELDO8-R8TSBHR-LG ). 9. Model Number: R8TSBHT; UDI-DI (Product Code): 00840331260489 (THYROIDSHIELDO8-R8TSBHT-MD ), 00840331260496 (THYROIDSHIELDO8-R8TSBHT-SM ). 10. Model Number: R8TSBMB; UDI-DI (Product Code): 00840331295979 (THYROIDSHIELDOM8-R8TSBMB-MD ), 00840331295986 (THYROIDSHIELDOM8-R8TSBMB-SM ). 11. Model Number: R8TSBMN; UDI-DI (Product Code): 00840331260519 (THYROIDSHIELDOM8-R8TSBMN-MD ). 12. Model Number: R8TSBMR; UDI-DI (Product Code): 00840331260533 (THYROIDSHIELDOM8-R8TSBMR-LG ), 00840331260540 (THYROIDSHIELDOM8-R8TSBMR-MD ), 00840331260557 (THYROIDSHIELDOM8-R8TSBMR-SM ). 13. Model Number: R8TSBMT; UDI-DI (Product Code): 00840331260571 (THYROIDSHIELDOM8-R8TSBMT-MD ). 14. Model Number: R8TSCBB; UDI-DI (Product Code): 00840331270525 (THYROIDSHIELD8-R8TSCBB-LG ), 00840331270532 (THYROIDSHIELD8-R8TSCBB-MD ), 00840331270549 (THYROIDSHIELD8-R8TSCBB-SM ). 15. Model Number: R8TSCBN; UDI-DI (Product Code): 00840331260595 (THYROIDSHIELD8-R8TSCBN-LG ), 00840331260601 (THYROIDSHIELD8-R8TSCBN-MD ), 00840331260618 (THYROIDSHIELD8-R8TSCBN-SM ). 16. Model Number: R8TSCBR; UDI-DI (Product Code): 00840331260625 (THYROIDSHIELD8-R8TSCBR-LG ), 00840331260632 (THYROIDSHIELD8-R8TSCBR-MD ), 00840331260649 (THYROIDSHIELD8-R8TSCBR-SM ). 17. Model Number: R8TSCBS; UDI-DI (Product Code): 00840331270594 (THYROIDSHIELD8-R8TSCBS-MD ), 00840331270587 (THYROIDSHIELD8-R8TSCBS-SM ). 18. Model Number: R8TSCBT; UDI-DI (Product Code): 00840331260656 (THYROIDSHIELD8-R8TSCBT-LG ), 00840331260663 (THYROIDSHIELD8-R8TSCBT-MD ), 00840331260670 (THYROIDSHIELD8-R8TSCBT-SM ). 19. Model Number: R8TSCHB; UDI-DI (Product Code): 00840331270495 (THYROIDSHIELD8-R8TSCHB-LG ), 00840331270501 (THYROIDSHIELD8-R8TSCHB-MD ), 00840331270518 (THYROIDSHIELD8-R8TSCHB-SM ). 20. Model Number: R8TSCHN; UDI-DI (Product Code): 00840331260687 (THYROIDSHIELD8-R8TSCHN-LG ), 00840331260694 (THYROIDSHIELD8-R8TSCHN-MD ), 00840331260700 (THYROIDSHIELD8-R8TSCHN-SM ). 21. Model Number: R8TSCHR; UDI-DI (Product Code): 00840331260717 (THYROIDSHIELD8-R8TSCHR-LG ), 00840331260724 (THYROIDSHIELD8-R8TSCHR-MD ), 00840331260731 (THYROIDSHIELD8-R8TSCHR-SM ). 22. Model Number: R8TSCHS; UDI-DI (Product Code): 00840331270563 (THYROIDSHIELD8-R8TSCHS-MD ), 00840331270556 (THYROIDSHIELD8-R8TSCHS-SM ). 23. Model Number: R8TSCHT; UDI-DI (Product Code): 00840331260748 (THYROIDSHIELD8-R8TSCHT-LG ), 00840331260755 (THYROIDSHIELD8-R8TSCHT-MD ), 00840331260762 (THYROIDSHIELD8-R8TSCHT-SM ). 24. Model Number: R8TSCMB; UDI-DI (Product Code): 00840331270464 (THYROIDSHIELDM8-R8TSCMB-LG ), 00840331270471 (THYROIDSHIELDM8-R8TSCMB-MD ), 00840331270488 (THYROIDSHIELDM8-R8TSCMB-SM ). 25. Model Number: R8TSCMN; UDI-DI (Product Code): 00840331260786 (THYROIDSHIELDM8-R8TSCMN-MD ), 00840331260793 (THYROIDSHIELDM8-R8TSCMN-SM ). 26. Model Number: R8TSCMR; UDI-DI (Product Code): 00840331260809 (THYROIDSHIELDM8-R8TSCMR-LG ), 00840331260816 (THYROIDSHIELDM8-R8TSCMR-MD ), 00840331260823 (THYROIDSHIELDM8-R8TSCMR-SM ). 27. Model Number: R8TSCMS; UDI-DI (Product Code): 00840331260830 (THYROIDSHIELDM8-R8TSCMS-MD ), 00840331260847 (THYROIDSHIELDM8-R8TSCMS-SM ). 28. Model Number: R8TSCMT; UDI-DI (Product Code): 00840331260854 (THYROIDSHIELDM8-R8TSCMT-LG ), 00840331260861 (THYROIDSHIELDM8-R8TSCMT-MD ), 00840331260878 (THYROIDSHIELDM8-R8TSCMT-SM ). 29. Model Number: R8TSOBB; UDI-DI (Product Code): 00840331271430 (THYROIDSHIELDO8-R8TSOBB-MD ). 30. Model Number: R8TSOBN; UDI-DI (Product Code): 00840331260892 (THYROIDSHIELDO8-R8TSOBN-MD ). 31. Model Number: R8TSOBR; UDI-DI (Product Code): 00840331260922 (THYROIDSHIELDO8-R8TSOBR-MD ), 00840331260939 (THYROIDSHIELDO8-R8TSOBR-SM ). 32. Model Number: R8TSOHB; UDI-DI (Product Code): 00840331260939 (THYROIDSHIELDO8-R8TSOHB-MD, THYROIDSHIELDO8-R8TSOHB-SM ). 33. Model Number: R8TSOHR; UDI-DI (Product Code): 00840331261011 (THYROIDSHIELDO8-R8TSOHR-MD ), 00840331261028 (THYROIDSHIELDO8-R8TSOHR-SM ). 34. Model Number: R8TSOHT; UDI-DI (Product Code): 00840331261035 (THYROIDSHIELDO8-R8TSOHT-LG ). 35. Model Number: R8TSOMN; UDI-DI (Product Code): 00840331261080 (THYROIDSHIELDOM8-R8TSOMN-SM ). 36. Model Number: R8TSOMR; UDI-DI (Product Code): 00840331261097 (THYROIDSHIELDOM8-R8TSOMR-LG ), 00840331261103 (THYROIDSHIELDOM8-R8TSOMR-MD ), 00840331261110 (THYROIDSHIELDOM8-R8TSOMR-SM ). 37. Model Number: R8TSOMT; UDI-DI (Product Code): 00840331261141 (THYROIDSHIELDOM8-R8TSOMT-SM ). 38. Model Number: R8TSSBB; UDI-DI (Product Code): 00840331270976 (THYROIDSHIELD8-R8TSSBB-LG ), 00840331270983 (THYROIDSHIELD8-R8TSSBB-MD ), 00840331270990 (THYROIDSHIELD8-R8TSSBB-SM ). 39. Model Number: R8TSSBN ; UDI-DI (Product Code): 00840331261165 (THYROIDSHIELD8-R8TSSBN-MD ), 00840331261172 (THYROIDSHIELD8-R8TSSBN-SM ), None (THYROIDSHIELDP8-R8TSSBN-PD ). 40. Model Number: R8TSSBR; UDI-DI (Product Code): 00840331261189 (THYROIDSHIELD8-R8TSSBR-LG ), 00840331261196 (THYROIDSHIELD8-R8TSSBR-MD ), 00840331261202 (THYROIDSHIELD8-R8TSSBR-SM ). 41. Model Number: R8TSSBS; UDI-DI (Product Code): 00840331271041 (THYROIDSHIELD8-R8TSSBS-MD ), 00840331271034 (THYROIDSHIELD8-R8TSSBS-SM ). 42. Model Number: R8TSSBT; UDI-DI (Product Code): 00840331261226 (THYROIDSHIELD8-R8TSSBT-MD ). 43. Model Number: R8TSSHB; UDI-DI (Product Code): 00840331270952 (THYROIDSHIELD8-R8TSSHB-MD ), 00840331270969 (THYROIDSHIELD8-R8TSSHB-SM ). 44. Model Number: R8TSSHN; UDI-DI (Product Code): 00840331261257 (THYROIDSHIELD8-R8TSSHN-MD ), 00840331261264 (THYROIDSHIELD8-R8TSSHN-SM ). 45. Model Number: R8TSSHR; UDI-DI (Product Code): 00840331261271 (THYROIDSHIELD8-R8TSSHR-LG ), 00840331261288 (THYROIDSHIELD8-R8TSSHR-MD ), 00840331261295 (THYROIDSHIELD8-R8TSSHR-SM ). 46. Model Number: R8TSSHS; UDI-DI (Product Code): 00840331271010 (THYROIDSHIELD8-R8TSSHS-MD ), 00840331271003 (THYROIDSHIELD8-R8TSSHS-SM ). 47. Model Number: R8TSSHT; UDI-DI (Product Code): 00840331261318 (THYROIDSHIELD8-R8TSSHT-MD ), 00840331261325 (THYROIDSHIELD8-R8TSSHT-SM ). 48. Model Number: R8TSSMB; UDI-DI (Product Code): 00840331270921 (THYROIDSHIELDM8-R8TSSMB-MD ), 00840331270938 (THYROIDSHIELDM8-R8TSSMB-SM ). 49. Model Number: R8TSSMN; UDI-DI (Product Code): 00840331261349 (THYROIDSHIELDM8-R8TSSMN-MD ), 00840331261356 (THYROIDSHIELDM8-R8TSSMN-SM ). 50. Model Number: R8TSSMR; UDI-DI (Product Code): 00840331261363 (THYROIDSHIELDM8-R8TSSMR-LG ), 00840331261370 (THYROIDSHIELDM8-R8TSSMR-MD ), 00840331261387 (THYROIDSHIELDM8-R8TSSMR-SM ). 51. Model Number: R8TSSMS; UDI-DI (Product Code): (THYROIDSHIELDM8-R8TSSMS-MD ), (THYROIDSHIELDM8-R8TSSMS-SM ). 52. Model Number: R8TSSMT; UDI-DI (Product Code): 00840331261400 (THYROIDSHIELDM8-R8TSSMT-MD ), 00840331261417 (THYROIDSHIELDM8-R8TSSMT-SM ). 53. Model Number: R8TSVBN; UDI-DI (Product Code): 00840331261431 (THYROIDSHIELD8-R8TSVBN-MD ). 54. Model Number: R8TSVBR; UDI-DI (Product Code): 00840331261462 (THYROIDSHIELD8-R8TSVBR-MD ), 00840331261479 (THYROIDSHIELD8-R8TSVBR-SM ). 55. Model Number: R8TSVHR; UDI-DI (Product Code): 00840331261547 (THYROIDSHIELD8-R8TSVHR-LG ), 00840331261554 (THYROIDSHIELD8-R8TSVHR-MD ), 00840331261561 (THYROIDSHIELD8-R8TSVHR-SM ). 56. Model Number: R8TSVHT; UDI-DI (Product Code): 00840331261561 (THYROIDSHIELD8-R8TSVHT-MD ), 00840331261622 (THYROIDSHIELD8-R8TSVHT-SM ). 57. Model Number: R8TSVMN; UDI-DI (Product Code): 00840331261646 (THYROIDSHIELDM8-R8TSVMN-MD ), 00840331261653 (THYROIDSHIELDM8-R8TSVMN-SM ). 58. Model Number: R8TSVMR; UDI-DI (Product Code): 00840331261660 (THYROIDSHIELDM8-R8TSVMR-LG ), 00840331261677 (THYROIDSHIELDM8-R8TSVMR-MD ), 00840331261684 (THYROIDSHIELDM8-R8TSVMR-SM ). 59. Model Number: R8TSVMT; UDI-DI (Product Code): 00840331261738 (THYROIDSHIELDM8-R8TSVMT-MD ), 00840331261745 (THYROIDSHIELDM8-R8TSVMT-SM ). 60. Model Number: S8TSBHR; UDI-DI (Product Code): 00840331296006 (THYROIDSHIELDO8-S8TSBHR-MD ). 61. Model Number: S8TSBHT; UDI-DI (Product Code): 00840331264197 (THYROIDSHIELDO8-S8TSBHT-LG ). 62. Model Number: S8TSBMB; UDI-DI (Product Code): 00840331264265 (THYROIDSHIELDOM8-S8TSBMB-MD ). 63. Model Number: S8TSBMR; UDI-DI (Product Code): 00840331264265 (THYROIDSHIELDOM8-S8TSBMR-MD ), 00840331264272 (THYROIDSHIELDOM8-S8TSBMR-SM ). 64. Model Number: S8TSCBB; UDI-DI (Product Code): 00840331270693 (THYROIDSHIELD8-S8TSCBB-SM ). 65. Model Number: S8TSCBN; UDI-DI (Product Code): 00840331264326 (THYROIDSHIELD8-S8TSCBN-MD ). 66. Model Number: S8TSCBR; UDI-DI (Product Code): 00840331264340 (THYROIDSHIELD8-S8TSCBR-LG ), 00840331264357 (THYROIDSHIELD8-S8TSCBR-MD ), 00840331264364 (THYROIDSHIELD8-S8TSCBR-SM ). 67. Model Number: S8TSCBT; UDI-DI (Product Code): 00840331264388 (THYROIDSHIELD8-S8TSCBT-MD ), 00840331264395 (THYROIDSHIELD8-S8TSCBT-SM ). 68. Model Number: S8TSCHN; UDI-DI (Product Code): 00840331264401 (THYROIDSHIELD8-S8TSCHN-LG ), 00840331264418 (THYROIDSHIELD8-S8TSCHN-MD ). 69. Model Number: S8TSCHR; UDI-DI (Product Code): 00840331264432 (THYROIDSHIELD8-S8TSCHR-LG ), 00840331264449 (THYROIDSHIELD8-S8TSCHR-MD ), 00840331264456 (THYROIDSHIELD8-S8TSCHR-SM ). 70. Model Number: S8TSCHT; UDI-DI (Product Code): 00840331264470 (THYROIDSHIELD8-S8TSCHT-MD ). 71. Model Number: S8TSCMB; UDI-DI (Product Code): 00840331270617 (THYROIDSHIELDM8-S8TSCMB-LG ). 72. Model Number: S8TSCMN; UDI-DI (Product Code): 00840331264517 (THYROIDSHIELDM8-S8TSCMN-SM ). 73. Model Number: S8TSCMR; UDI-DI (Product Code): 00840331264524 (THYROIDSHIELDM8-S8TSCMR-LG ), 00840331264531 (THYROIDSHIELDM8-S8TSCMR-MD ), 00840331264548 (THYROIDSHIELDM8-S8TSCMR-SM ). 74. Model Number: S8TSCMS; UDI-DI (Product Code): 00840331264555 (THYROIDSHIELDM8-S8TSCMS-MD ), 00840331264562 (THYROIDSHIELDM8-S8TSCMS-SM ). 75. Model Number: S8TSCMT; UDI-DI (Product Code): 00840331264579 (THYROIDSHIELDM8-S8TSCMT-LG ), 00840331264586 (THYROIDSHIELDM8-S8TSCMT-MD ), 00840331264593 (THYROIDSHIELDM8-S8TSCMT-SM ). 76. Model Number: S8TSOBR; UDI-DI (Product Code): 00840331264807 (THYROIDSHIELDO8-S8TSOBR-LG ), 00840331264784 (THYROIDSHIELDO8-S8TSOBR-SM ). 77. Model Number: S8TSSBN; UDI-DI (Product Code): 00840331264883 (THYROIDSHIELD8-S8TSSBN-MD ). 78. Model Number: S8TSSBR; UDI-DI (Product Code): 00840331264906 (THYROIDSHIELD8-S8TSSBR-LG ), 00840331264913 (THYROIDSHIELD8-S8TSSBR-MD ), 00840331264920 (THYROIDSHIELD8-S8TSSBR-SM ). 79. Model Number: S8TSSBT; UDI-DI (Product Code): 00840331264944 (THYROIDSHIELD8-S8TSSBT-MD ). 80. Model Number: S8TSSHR; UDI-DI (Product Code): 00840331264999 (THYROIDSHIELD8-S8TSSHR-LG ), 00840331265002 (THYROIDSHIELD8-S8TSSHR-MD ), 00840331265019 (THYROIDSHIELD8-S8TSSHR-SM ). 81. Model Number: S8TSSHS; UDI-DI (Product Code): 00840331271164 (THYROIDSHIELD8-S8TSSHS-MD ). 82. Model Number: S8TSSHT; UDI-DI (Product Code): 00840331265033 (THYROIDSHIELD8-S8TSSHT-MD ). 83. Model Number: S8TSSMN; UDI-DI (Product Code): 00840331265064 (THYROIDSHIELDM8-S8TSSMN-MD ). 84. Model Number: S8TSSMR; UDI-DI (Product Code): 00840331265088 (THYROIDSHIELDM8-S8TSSMR-LG ), 00840331265095 (THYROIDSHIELDM8-S8TSSMR-MD ), 00840331265101 (THYROIDSHIELDM8-S8TSSMR-SM ). 85. Model Number: S8TSSMS; UDI-DI (Product Code): (THYROIDSHIELDM8-S8TSSMS-MD ). 86. Model Number: S8TSSMT; UDI-DI (Product Code): 00840331265118 (THYROIDSHIELDM8-S8TSSMT-LG ), 00840331265132 (THYROIDSHIELDM8-S8TSSMT-SM ). 87. Model Number: S8TSVBB; UDI-DI (Product Code): 00840331271799 (THYROIDSHIELD8-S8TSVBB-MD ). 88. Model Number: S8TSVBR; UDI-DI (Product Code): 00840331265187 (THYROIDSHIELD8-S8TSVBR-MD ). 89. Model Number: S8TSVHR; UDI-DI (Product Code): 00840331265279 (THYROIDSHIELD8-S8TSVHR-MD ), 00840331265286 (THYROIDSHIELD8-S8TSVHR-SM ). 90. Model Number: S8TSVMR; UDI-DI (Product Code): 00840331265392 (THYROIDSHIELDM8-S8TSVMR-MD ), 00840331265408 (THYROIDSHIELDM8-S8TSVMR-SM ). 91. Model Number: U8TSBBR; UDI-DI (Product Code): 00840331267853 (THYROIDSHIELDO8-U8TSBBR-SM ). 92. Model Number: U8TSBMN; UDI-DI (Product Code): 00840331267990 (THYROIDSHIELDOM8-U8TSBMN-MD ). 93. Model Number: U8TSBMR; UDI-DI (Product Code): 00840331268010 (THYROIDSHIELDOM8-U8TSBMR-LG ), 00840331268027 (THYROIDSHIELDOM8-U8TSBMR-MD ). 94. Model Number: U8TSCBN; UDI-DI (Product Code): 00840331268089 (THYROIDSHIELD8-U8TSCBN-MD ). 95. Model Number: U8TSCBR; UDI-DI (Product Code): 00840331268102 (THYROIDSHIELD8-U8TSCBR-LG ), 00840331268119 (THYROIDSHIELD8-U8TSCBR-MD ), 00840331268126 (THYROIDSHIELD8-U8TSCBR-SM ). 96. Model Number: U8TSCBT; UDI-DI (Product Code): 00840331268140 (THYROIDSHIELD8-U8TSCBT-MD ). 97. Model Number: U8TSCHR; UDI-DI (Product Code): 00840331268201 (THYROIDSHIELD8-U8TSCHR-MD ), 00840331268218 (THYROIDSHIELD8-U8TSCHR-SM ). 98. Model Number: U8TSCMR; UDI-DI (Product Code): 00840331268294 (THYROIDSHIELDM8-U8TSCMR-MD ), 00840331268300 (THYROIDSHIELDM8-U8TSCMR-SM ). 99. Model Number: U8TSCMT; UDI-DI (Product Code): 00840331268324 (THYROIDSHIELDM8-U8TSCMT-MD ), 00840331268331 (THYROIDSHIELDM8-U8TSCMT-SM ). 100. Model Number: U8TSOMR; UDI-DI (Product Code): 00840331268560 (THYROIDSHIELDOM8-U8TSOMR-MD ). 101. Model Number: U8TSSBN; UDI-DI (Product Code): 00840331268621 (THYROIDSHIELD8-U8TSSBN-MD ). 102. Model Number: U8TSSBR; UDI-DI (Product Code): 00840331268652 (THYROIDSHIELD8-U8TSSBR-MD ). 103. Model Number: U8TSSHN; UDI-DI (Product Code): 00840331268713 (THYROIDSHIELD8-U8TSSHN-MD ). 104. Model Number: U8TSSHR; UDI-DI (Product Code): 00840331268737 (THYROIDSHIELD8-U8TSSHR-LG ), 00840331268744 (THYROIDSHIELD8-U8TSSHR-MD ). 105. Model Number: U8TSSMR; UDI-DI (Product Code): 00840331268829 (THYROIDSHIELDM8-U8TSSMR-LG ), 00840331268836 (THYROIDSHIELDM8-U8TSSMR-MD ), 00840331268843 (THYROIDSHIELDM8-U8TSSMR-SM ). 106. Model Number: U8TSSMS; UDI-DI (Product Code): None (THYROIDSHIELDM8-U8TSSMS-SM ). 107. Model Number: U8TSSMT; UDI-DI (Product Code): 00840331268867 (THYROIDSHIELDM8-U8TSSMT-MD ), 00840331268874 (THYROIDSHIELDM8-U8TSSMT-SM ). 108. Model Number: U8TSVMR; UDI-DI (Product Code): 00840331269130 (THYROIDSHIELDM8-U8TSVMR-MD ), 00840331269147 (THYROIDSHIELDM8-U8TSVMR-SM ).
Devices named in this recall
196 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- THYROID SHIELD R8TSBBB-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBN-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBBT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBHT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBMN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBMR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSBMT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBN-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBT-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCBT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHN-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHT-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCHT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMS-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMS-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMT-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSCMT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOBN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOHR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOHR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOHT-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOMN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOMR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSOMT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSBT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSHT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSSMT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVBN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVHR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVHR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVHR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVHT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMN-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD R8TSVMT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSBHT-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSBMR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSBMR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSCBN-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSCBR-LGBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSCBR-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSCBR-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- THYROID SHIELD S8TSCBT-MDBurlington Medical, LLC · Apron, Protective1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98415).
- Z-1754-2026 · Open, ClassifiedBurlington Medical, Vest
- Z-1755-2026 · Open, ClassifiedBurlington Medical, Kilt
- Z-1756-2026 · Open, ClassifiedBurlington Medical, Frontal Aprons
- Z-1757-2026 · Open, ClassifiedBurlington Medical, Wrap Aprons.
- Z-1758-2026 · Open, ClassifiedBurlington Medical, BAT (Breast, Axilla and Thyroid) Coverage.
- Z-1759-2026 · Open, ClassifiedBurlington Medical, Protective Sleeves.
- Z-1760-2026 · Open, ClassifiedBurlington Medical, Caps.
- Z-1761-2026 · Open, ClassifiedBurlington Medical, Leg Wraps.
- Z-1762-2026 · Open, ClassifiedBurlington Medical, Half Aprons.
- Z-1763-2026 · Open, ClassifiedBurlington Medical, Demi Half Aprons.
- Z-1764-2026 · Open, ClassifiedBurlington Medical, Blockers.
- Z-1765-2026 · Open, ClassifiedBurlington Medical, Kilt Blockers.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page