Burlington Medical, Vest recallZ-1754-2026 · Burlington Medical, LLC · initiated Feb 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Burlington Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for attenuation degradation over time, decreasing the lifespan.
FDA root cause category: Process change control
What the company is doing
Burlington Medical notified consignees on about 02/11/2026 via "URGENT - FIELD SAFETY NOTICE (FSN)" letter. Consignees were instructed to ensure all users of the specific Burlington Medical radiation protection garments made aware of this Recall Notice, review inventory for affected units, arrange for the return of those affected units for proper disposal, and complete and return the Recall Response Form. Consignees were also instructed to notify appropriate individuals within your organization and to any organization where the potentially affected devices were transferred or sold.
Recall details
- Recall number
- Z-1754-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 11, 2026
- Posted by the FDA
- Apr 2, 2026
- Recalling firm
- Burlington Medical, LLC
- Quantity
- 15,438 units
- Product code
- IWO
- Distribution
- Worldwide distribution. Nationwide distribution in the US and international distribution to Canada, Australia, Bangladesh, Chile, Hong Kong, India, Mexico, Malaysia, Panama, Singapore, Thailand, Taiwan, Uruguay, and Vietnam.
Product description
Burlington Medical, Vest
Code information (models, lots, serial numbers)
Show the full code information (19k characters)
All serial numbers are affected that were manufactured with Xenolite 800 attenuation material from 01/16/2025 to 01/21/2026. 1. Model Number: C8PVEST; UDI-DI (Product Code): 00840331297652 (VESTC18-C8PVEST-2XM ), 00840331297652 (VESTC8-C8PVEST-LF ), 00840331297652 (VESTC8-C8PVEST-LM ), 00840331297652 (VESTC8-C8PVEST-MF ), 00840331297652 (VESTC8-C8PVEST-MM ), 00840331297652 (VESTC8-C8PVEST-SF ), 00840331297652 (VESTC8-C8PVEST-SM ), 00840331297652 (VESTC8-C8PVEST-XLM ), 00840331297652 (VESTT18-C8PVEST-LF ), 00840331297652 (VESTT18-C8PVEST-LM ), 00840331297652 (VESTT18-C8PVEST-MM ), 00840331297652 (VESTT18-C8PVEST-PLUS ), 00840331297652 (VESTT18-C8PVEST-SM ), 00840331297652 (VESTT28-C8PVEST-3XM ), 00840331297652 (VESTT28-C8PVEST-4XM ), 00840331297652 (VESTT28-C8PVEST-5XM ). 2. Model Number: C8VEST; UDI-DI (Product Code): 00840331297645 (VESTC18-C8VEST-2XM ), 00840331297645 (VESTC8-C8VEST-LF ), 00840331297645 (VESTC8-C8VEST-LM ), 00840331297645 (VESTC8-C8VEST-MF ), 00840331297645 (VESTC8-C8VEST-MM ), 00840331297645 (VESTC8-C8VEST-PLUS ), 00840331297645 (VESTC8-C8VEST-SF ), 00840331297645 (VESTC8-C8VEST-SM ), 00840331297645 (VESTC8-C8VEST-XLF ), 00840331297645 (VESTC8-C8VEST-XLM ), 00840331297645 (VESTC8-C8VEST-XSF ), 00840331297645 (VESTT18-C8VEST-LM ), 00840331297645 (VESTT18-C8VEST-MM ), 00840331297645 (VESTT18-C8VEST-SM ), 00840331297645 (VESTT18-C8VEST-XLM ), 00840331297645 (VESTT18-C8VEST-XSM ), 00840331297645 (VESTT28-C8VEST-2XM ), 00840331297645 (VESTT28-C8VEST-4XF ). 3. Model Number: R8FVEST; UDI-DI (Product Code): 00840331259520 (VESTC18-R8FVEST-2XF ), 00840331259537 (VESTC18-R8FVEST-2XM ), 00840331259551 (VESTC18-R8FVEST-3XM ), 00840331259568 (VESTC8-R8FVEST-LF ), 00840331259575 (VESTC8-R8FVEST-LM ), 00840331259582 (VESTC8-R8FVEST-MF ), 00840331259599 (VESTC8-R8FVEST-MM ), 00840331259605 (VESTC8-R8FVEST-PLUS ), 00840331259612 (VESTC8-R8FVEST-SF ), 00840331259629 (VESTC8-R8FVEST-SM ), 00840331259636 (VESTC8-R8FVEST-XLF ), 00840331259643 (VESTC8-R8FVEST-XLM ), 00840331259650 (VESTC8-R8FVEST-XSF ), 00840331269741 (VESTT18-R8FVEST-LF ), 00840331269758 (VESTT18-R8FVEST-LM ), 00840331269741 (VESTT18-R8FVEST-MF ), 00840331269758 (VESTT18-R8FVEST-MM ), 00840331269741 (VESTT18-R8FVEST-PLUS ), 00840331269741 (VESTT18-R8FVEST-SF ), 00840331269758 (VESTT18-R8FVEST-SM ), 00840331269741 (VESTT18-R8FVEST-XLF ), 00840331269758 (VESTT18-R8FVEST-XLM ), 00840331269758 (VESTT18-R8FVEST-XSM ), 00840331269741 (VESTT28-R8FVEST-2XF ), 00840331269758 (VESTT28-R8FVEST-2XM ), 00840331269741 (VESTT28-R8FVEST-3XF ), 00840331269758 (VESTT28-R8FVEST-3XM ), 00840331269741 (VESTT28-R8FVEST-4XF ), 00840331269758 (VESTT28-R8FVEST-4XM ), 00840331269741 (VESTT28-R8FVEST-5XF ), 00840331269758 (VESTT28-R8FVEST-5XM ), 00840331269741 (VESTT38-R8FVEST-6XF ), 00840331269758 (VESTT38-R8FVEST-6XM ), 00840331269741 (VESTT38-R8FVEST-8XF ), 00840331269758 (VESTT38-R8FVEST-8XM ), None (IVESTC8-R8FVEST-MFR ), None (IVESTC8-R8FVEST-MML ), None (IVESTC8-R8FVEST-LFL ), None (IVESTC8-R8FVEST-LFR ), None (IVESTC8-R8FVEST-LML ), None (IVESTC8-R8FVEST-MMR ), None (IVESTC8-R8FVEST-PLUSL ), None (IVESTC8-R8FVEST-SFR ), None (IVESTC8-R8FVEST-SFS ), None (IVESTC8-R8FVEST-SMR ), None (IVESTC8-R8FVEST-SMS ), None (IVESTC8-R8FVEST-XLML ), None (IVESTC8-R8FVEST-XSFS ), None (IVESTC8-R8FVEST-MMXL ), None (IVESTC8-R8FVEST-XLMXL ), None (IVESTC8-R8FVEST-LMR ), None (IVESTC8-R8FVEST-XSFR ), None (IVESTC8-R8FVEST-XSMR ). 4. Model Number: R8PVEST; UDI-DI (Product Code): None (IVESTC8-R8PVEST-LMR ), None (IVESTC8-R8PVEST-SMR ), 00840331259841 (VESTC18-R8PVEST-2XF ), 00840331259858 (VESTC18-R8PVEST-2XM ), 00840331259865 (VESTC18-R8PVEST-3XF ), 00840331259872 (VESTC18-R8PVEST-3XM ), 00840331259889 (VESTC8-R8PVEST-LF ), 00840331259896 (VESTC8-R8PVEST-LM ), 00840331259902 (VESTC8-R8PVEST-MF ), 00840331259919 (VESTC8-R8PVEST-MM ), 00840331259926 (VESTC8-R8PVEST-PLUS ), 00840331259933 (VESTC8-R8PVEST-SF ), 00840331259940 (VESTC8-R8PVEST-SM ), 00840331259957 (VESTC8-R8PVEST-XLF ), 00840331259964 (VESTC8-R8PVEST-XLM ), 00840331259971 (VESTC8-R8PVEST-XSF ), 00840331259988 (VESTC8-R8PVEST-XSM ), 00840331269680 (VESTT18-R8PVEST-LF ), 00840331269697 (VESTT18-R8PVEST-LM ), 00840331269680 (VESTT18-R8PVEST-MF ), 00840331269697 (VESTT18-R8PVEST-MM ), 00840331269680 (VESTT18-R8PVEST-PLUS ), 00840331269680 (VESTT18-R8PVEST-SF ), 00840331269697 (VESTT18-R8PVEST-SM ), 00840331269680 (VESTT18-R8PVEST-XLF ), 00840331269697 (VESTT18-R8PVEST-XLM ), 00840331269697 (VESTT18-R8PVEST-XSM ), 00840331269680 (VESTT28-R8PVEST-2XF ), 00840331269697 (VESTT28-R8PVEST-2XM ), 00840331269680 (VESTT28-R8PVEST-3XF ), 00840331269697 (VESTT28-R8PVEST-3XM ), 00840331269680 (VESTT28-R8PVEST-4XF ), 00840331269697 (VESTT28-R8PVEST-4XM ), 00840331269680 (VESTT28-R8PVEST-5XF ), 00840331269697 (VESTT28-R8PVEST-5XM ), 00840331269680 (VESTT38-R8PVEST-6XF ), 00840331269697 (VESTT38-R8PVEST-6XM ), 00840331269680 (VESTT38-R8PVEST-7XF ), 00840331269697 (VESTT38-R8PVEST-7XM ), 00840331269680 (VESTT38-R8PVEST-8XF ), 00840331269697 (VESTT38-R8PVEST-8XM ), None (IVESTC8-R8PVEST-LFR ), None (IVESTC8-R8PVEST-MFL ), None (IVESTC8-R8PVEST-MFR ), None (IVESTC8-R8PVEST-MML ), None (IVESTC8-R8PVEST-MMR ), None (IVESTC8-R8PVEST-SFR ), None (IVESTC8-R8PVEST-XSFR ), None (IVESTC8-R8PVEST-2XFR ), None (IVESTC8-R8PVEST-LML ), None (IVESTC8-R8PVEST-LMXL ), None (IVESTC8-R8PVEST-MFS ), None (IVESTC8-R8PVEST-SFL ), None (IVESTC8-R8PVEST-SFS ), None (IVESTC8-R8PVEST-XLFL ), None (IVESTC8-R8PVEST-XLMR ). 5. Model Number: R8QVEST; UDI-DI (Product Code): 00840331260076 (QVESTC18-R8QVEST-LF ), 00840331260083 (QVESTC18-R8QVEST-MF ), None (QVESTC18-R8QVEST-PLUS ), 00840331260106 (QVESTC18-R8QVEST-SF ), 00840331260113 (QVESTC18-R8QVEST-XLF ), 00840331260007 (QVESTC28-R8QVEST-2XF ), 00840331269802 (QVESTT18-R8QVEST-LF ), 00840331269802 (QVESTT18-R8QVEST-MF ), 00840331269802 (QVESTT18-R8QVEST-PLUS ), 00840331269802 (QVESTT18-R8QVEST-SF ), 00840331269802 (QVESTT18-R8QVEST-XLF ), 00840331269802 (QVESTT28-R8QVEST-2XF ), 00840331269802 (QVESTT28-R8QVEST-3XF ), 00840331269802 (QVESTT28-R8QVEST-4XF ), 00840331269802 (QVESTT28-R8QVEST-5XF ), 00840331269802 (QVESTT38-R8QVEST-8XF ), None (IQVESTC8-R8QVEST-SFR ), 00840331260120 (QVESTC18-R8QVEST-XSF ), None (IQVESTC8-R8QVEST-MFR). 6. Model Number: R8VEST; UDI-DI (Product Code): 00840331261752 (VESTC18-R8VEST-2XF ), 00840331261769 (VESTC18-R8VEST-2XM ), 00840331261776 (VESTC18-R8VEST-3XF ), 00840331261783 (VESTC18-R8VEST-3XM ), 00840331261790 (VESTC8-R8VEST-LF ), 00840331261806 (VESTC8-R8VEST-LM ), 00840331261813 (VESTC8-R8VEST-MF ), 00840331261820 (VESTC8-R8VEST-MM ), 00840331261837 (VESTC8-R8VEST-PLUS ), 00840331261844 (VESTC8-R8VEST-SF ), 00840331261851 (VESTC8-R8VEST-SM ), 00840331261868 (VESTC8-R8VEST-XLF ), 00840331261875 (VESTC8-R8VEST-XLM ), 00840331261882 (VESTC8-R8VEST-XSF ), 00840331261899 (VESTC8-R8VEST-XSM ), 00840331261790 (VESTS8-R8VEST-LF ), 00840331261806 (VESTS8-R8VEST-LM ), 00840331261813 (VESTS8-R8VEST-MF ), 00840331261820 (VESTS8-R8VEST-MM ), 00840331261837 (VESTS8-R8VEST-PLUS ), 00840331261844 (VESTS8-R8VEST-SF ), 00840331261851 (VESTS8-R8VEST-SM ), 00840331261868 (VESTS8-R8VEST-XLF ), 00840331261875 (VESTS8-R8VEST-XLM ), 00840331269628 (VESTT18-R8VEST-LF ), 00840331269635 (VESTT18-R8VEST-LM ), 00840331269628 (VESTT18-R8VEST-MF ), 00840331269635 (VESTT18-R8VEST-MM ), 00840331269628 (VESTT18-R8VEST-PLUS ), 00840331269628 (VESTT18-R8VEST-SF ), 00840331269635 (VESTT18-R8VEST-SM ), 00840331269628 (VESTT18-R8VEST-XLF ), 00840331269635 (VESTT18-R8VEST-XLM ), 00840331269628 (VESTT18-R8VEST-XSF ), 00840331269635 (VESTT18-R8VEST-XSM ), 00840331269628 (VESTT28-R8VEST-2XF ), 00840331269635 (VESTT28-R8VEST-2XM ), 00840331269628 (VESTT28-R8VEST-3XF ), 00840331269635 (VESTT28-R8VEST-3XM ), 00840331269628 (VESTT28-R8VEST-4XF ), 00840331269635 (VESTT28-R8VEST-4XM ), 00840331269628 (VESTT28-R8VEST-5XF ), 00840331269635 (VESTT28-R8VEST-5XM ), 00840331269628 (VESTT38-R8VEST-6XF ), 00840331269635 (VESTT38-R8VEST-6XM ), 00840331269628 (VESTT38-R8VEST-7XF ), 00840331269635 (VESTT38-R8VEST-7XM ), 00840331269628 (VESTT38-R8VEST-8XF ), 00840331269635 (VESTT38-R8VEST-8XM ), None (IVESTC8-R8VEST-LFR ), None (IVESTC8-R8VEST-LMR ), None (IVESTC8-R8VEST-MMR ), None (IVESTC8-R8VEST-SMS ), None (IVESTC8-R8VEST-MFR ), None (IVESTC8-R8VEST-SFR), None (IVESTC8-R8VEST-LML ), None (IVESTC8-R8VEST-SMR ). 7. Model Number: R8ZVEST; UDI-DI (Product Code): 00840331262063 (ZVESTC18-R8ZVEST-LM), 00840331262087 (ZVESTC18-R8ZVEST-MM), 00840331262087 (ZVESTC18-R8ZVEST-SF), 00840331262148 (ZVESTC18-R8ZVEST-XSF), 00840331269840 (ZVESTT18-R8ZVEST-LM), 00840331269833 (ZVESTT18-R8ZVEST-PLUS), 00840331269840 (ZVESTT28-R8ZVEST-2XM), 00840331269833 (ZVESTT28-R8ZVEST-3XF), None (IZVESTC8-R8ZVEST-LFR). 8. Model Number: S8FVEST; UDI-DI (Product Code): 00840331263275 (VESTC18-S8FVEST-2XF ), 00840331263305 (VESTC18-S8FVEST-3XM ), 00840331263312 (VESTC8-S8FVEST-LF ), 00840331263329 (VESTC8-S8FVEST-LM ), 00840331263336 (VESTC8-S8FVEST-MF ), 00840331263343 (VESTC8-S8FVEST-MM ), 00840331263350 (VESTC8-S8FVEST-PLUS ), 00840331263367 (VESTC8-S8FVEST-SF ), 00840331263374 (VESTC8-S8FVEST-SM ), 00840331263381 (VESTC8-S8FVEST-XLF ), 00840331263398 (VESTC8-S8FVEST-XLM ), 00840331269765 (VESTT18-S8FVEST-LF ), 00840331269772 (VESTT18-S8FVEST-LM ), 00840331269765 (VESTT18-S8FVEST-MF ), 00840331269772 (VESTT18-S8FVEST-MM ), 00840331269765 (VESTT18-S8FVEST-PLUS ), 00840331269765 (VESTT18-S8FVEST-SF ), 00840331269772 (VESTT18-S8FVEST-SM ), 00840331269765 (VESTT18-S8FVEST-XLF ), 00840331269772 (VESTT18-S8FVEST-XLM ), 00840331269772 (VESTT28-S8FVEST-2XM ), 00840331269765 (VESTT28-S8FVEST-3XF ), 00840331269772 (VESTT28-S8FVEST-3XM ), 00840331269765 (VESTT28-S8FVEST-4XF ), 00840331269772 (VESTT28-S8FVEST-4XM ), 00840331269765 (VESTT28-S8FVEST-5XF ), 00840331269772 (VESTT28-S8FVEST-5XM ), 00840331269765 (VESTT38-S8FVEST-6XF ), 00840331269772 (VESTT38-S8FVEST-6XM ), 00840331269765 (VESTT38-S8FVEST-7XF ), 00840331269765 (VESTT38-S8FVEST-8XF ). 9. Model Number: S8PVEST; UDI-DI (Product Code): 00840331263596 (VESTC18-S8PVEST-2XF ), 00840331263602 (VESTC18-S8PVEST-2XM ), 00840331263633 (VESTC8-S8PVEST-LF ), 00840331263640 (VESTC8-S8PVEST-LM ), 00840331263657 (VESTC8-S8PVEST-MF ), 00840331263664 (VESTC8-S8PVEST-MM ), 00840331263671 (VESTC8-S8PVEST-PLUS ), 00840331263688 (VESTC8-S8PVEST-SF ), 00840331263695 (VESTC8-S8PVEST-SM ), 00840331263701 (VESTC8-S8PVEST-XLF ), 00840331263718 (VESTC8-S8PVEST-XLM ), 00840331269703 (VESTT18-S8PVEST-LF ), 00840331269710 (VESTT18-S8PVEST-LM ), 00840331269703 (VESTT18-S8PVEST-MF ), 00840331269710 (VESTT18-S8PVEST-MM ), 00840331269703 (VESTT18-S8PVEST-PLUS ), 00840331269703 (VESTT18-S8PVEST-SF ), 00840331269710 (VESTT18-S8PVEST-SM ), 00840331269703 (VESTT18-S8PVEST-XLF ), 00840331269710 (VESTT18-S8PVEST-XLM ), 00840331269710 (VESTT18-S8PVEST-XSM ), 00840331269703 (VESTT28-S8PVEST-2XF ), 00840331269710 (VESTT28-S8PVEST-2XM ), 00840331269703 (VESTT28-S8PVEST-3XF ), 00840331269710 (VESTT28-S8PVEST-3XM ), 00840331269703 (VESTT28-S8PVEST-4XF ), 00840331269710 (VESTT28-S8PVEST-4XM ), 00840331269703 (VESTT28-S8PVEST-5XF ), 00840331269710 (VESTT28-S8PVEST-5XM ), 00840331269703 (VESTT38-S8PVEST-6XF ), 00840331269710 (VESTT38-S8PVEST-6XM ), 00840331269703 (VESTT38-S8PVEST-7XF ), 00840331269710 (VESTT38-S8PVEST-7XM ), 00840331269703 (VESTT38-S8PVEST-8XF ), 00840331269710 (VESTT38-S8PVEST-8XM ). 10. Model Number: S8QVEST; UDI-DI (Product Code): 00840331263794 (QVESTC18-S8QVEST-LF ), 00840331263800 (QVESTC18-S8QVEST-MF ), 00840331263817 (QVESTC18-S8QVEST-PLUS ), 00840331263824 (QVESTC18-S8QVEST-SF ), 00840331263831 (QVESTC18-S8QVEST-XLF ), 00840331269819 (QVESTT18-S8QVEST-LF ), 00840331269819 (QVESTT18-S8QVEST-MF ), 00840331269819 (QVESTT18-S8QVEST-PLUS ), 00840331269819 (QVESTT18-S8QVEST-XLF ), 00840331269819 (QVESTT28-S8QVEST-2XF ), 00840331269819 (QVESTT28-S8QVEST-3XF ), 00840331269819 (QVESTT28-S8QVEST-4XF ), 00840331269819 (QVESTT28-S8QVEST-5XF ), 00840331269819 (QVESTT38-S8QVEST-6XF ), 00840331269819 (QVESTT38-S8QVEST-7XF ). 11. Model Number: S8VEST; UDI-DI (Product Code): 00840331265477 (VESTC18-S8VEST-2XF ), 00840331265484 (VESTC18-S8VEST-2XM ), 00840331265491 (VESTC18-S8VEST-3XF ), 00840331265514 (VESTC8-S8VEST-LF ), 00840331265521 (VESTC8-S8VEST-LM ), 00840331265538 (VESTC8-S8VEST-MF ), 00840331265545 (VESTC8-S8VEST-MM ), 00840331265552 (VESTC8-S8VEST-PLUS ), 00840331265569 (VESTC8-S8VEST-SF ), 00840331265576 (VESTC8-S8VEST-SM ), 00840331265583 (VESTC8-S8VEST-XLF ), 00840331265590 (VESTC8-S8VEST-XLM ), 00840331265606 (VESTC8-S8VEST-XSF ), 00840331265613 (VESTC8-S8VEST-XSM ), 00840331265514 (VESTS8-S8VEST-LF ), 00840331265521 (VESTS8-S8VEST-LM ), 00840331265538 (VESTS8-S8VEST-MF ), 00840331265545 (VESTS8-S8VEST-MM ), 00840331265569 (VESTS8-S8VEST-SF ), 00840331265576 (VESTS8-S8VEST-SM ), 00840331265583 (VESTS8-S8VEST-XLF ), 00840331265590 (VESTS8-S8VEST-XLM ), 00840331269642 (VESTT18-S8VEST-LF ), 00840331269659 (VESTT18-S8VEST-LM ), 00840331269642 (VESTT18-S8VEST-MF ), 00840331269659 (VESTT18-S8VEST-MM ), 00840331269642 (VESTT18-S8VEST-PLUS ), 00840331269642 (VESTT18-S8VEST-SF ), 00840331269659 (VESTT18-S8VEST-SM ), 00840331269642 (VESTT18-S8VEST-XLF ), 00840331269659 (VESTT18-S8VEST-XLM ), 00840331269642 (VESTT18-S8VEST-XSF ), 00840331269659 (VESTT18-S8VEST-XSM ), 00840331269642 (VESTT28-S8VEST-2XF ), 00840331269659 (VESTT28-S8VEST-2XM ), 00840331269642 (VESTT28-S8VEST-3XF ), 00840331269659 (VESTT28-S8VEST-3XM ), 00840331269642 (VESTT28-S8VEST-4XF ), 00840331269659 (VESTT28-S8VEST-4XM ), 00840331269642 (VESTT28-S8VEST-5XF ), 00840331269659 (VESTT28-S8VEST-5XM ), 00840331269642 (VESTT38-S8VEST-6XF ), 00840331269659 (VESTT38-S8VEST-6XM ), 00840331269642 (VESTT38-S8VEST-7XF ), 00840331269659 (VESTT38-S8VEST-7XM ), 00840331269642 (VESTT38-S8VEST-8XF ), 00840331269659 (VESTT38-S8VEST-8XM ), None (IVESTC8-S8VEST-SMS ), None (IVESTC8-S8VEST-LMR ), None (IVESTC8-S8VEST-MMR ). 12. Model Number: S8ZVEST; UDI-DI (Product Code): 00840331265781 (ZVESTC18-S8ZVEST-LM ), 00840331265804 (ZVESTC18-S8ZVEST-MM ), 00840331269864 (ZVESTT18-S8ZVEST-LF ), 00840331269864 (ZVESTT18-S8ZVEST-PLUS ). 13. Model Number: U8FVEST; UDI-DI (Product Code): 00840331267044 (VESTC8-U8FVEST-LF ), 00840331267051 (VESTC8-U8FVEST-LM ), 00840331267075 (VESTC8-U8FVEST-MM ), 00840331267099 (VESTC8-U8FVEST-SF ), 00840331267105 (VESTC8-U8FVEST-SM ), 00840331267112 (VESTC8-U8FVEST-XLF ), 00840331267136 (VESTC8-U8FVEST-XSF ), 00840331269796 (VESTT18-U8FVEST-LM ), 00840331269789 (VESTT18-U8FVEST-PLUS ), 00840331269789 (VESTT18-U8FVEST-XLF ), 00840331269796 (VESTT28-U8FVEST-2XM ), 00840331269789 (VESTT28-U8FVEST-3XF ), 00840331269796 (VESTT28-U8FVEST-3XM ), 00840331269789 (VESTT38-U8FVEST-6XF ), None (IVESTC8-U8FVEST-LML ), None (IVESTC8-U8FVEST-LMR ), None (IVESTC8-U8FVEST-MFR ), None (IVESTC8-U8FVEST-MMR ), None (IVESTC8-U8FVEST-SFR ), None (IVESTC8-U8FVEST-XLML ), None (IVESTC8-U8FVEST-XLMR ). 14. Model Number: U8PVEST; UDI-DI (Product Code): 00840331267013 (VESTC18-U8PVEST-2XM ), 00840331267365 (VESTC8-U8PVEST-LF ), 00840331267372 (VESTC8-U8PVEST-LM ), 00840331267389 (VESTC8-U8PVEST-MF ), 00840331267396 (VESTC8-U8PVEST-MM ), 00840331267402 (VESTC8-U8PVEST-PLUS ), 00840331267419 (VESTC8-U8PVEST-SF ), 00840331267433 (VESTC8-U8PVEST-XLF ), 00840331267440 (VESTC8-U8PVEST-XLM ), 00840331267464 (VESTC8-U8PVEST-XSM ), 00840331269727 (VESTT18-U8PVEST-LF ), 00840331269734 (VESTT18-U8PVEST-LM ), 00840331269727 (VESTT18-U8PVEST-MF ), 00840331269734 (VESTT18-U8PVEST-MM ), 00840331269727 (VESTT18-U8PVEST-PLUS ), 00840331269727 (VESTT18-U8PVEST-SF ), 00840331269734 (VESTT18-U8PVEST-SM ), 00840331269727 (VESTT18-U8PVEST-XLF ), 00840331269734 (VESTT18-U8PVEST-XLM ), 00840331269734 (VESTT18-U8PVEST-XSM ), 00840331269727 (VESTT28-U8PVEST-2XF ), 00840331269734 (VESTT28-U8PVEST-2XM ), 00840331269727 (VESTT28-U8PVEST-3XF ), 00840331269734 (VESTT28-U8PVEST-3XM ), 00840331269734 (VESTT28-U8PVEST-4XF ), 00840331269727 (VESTT28-U8PVEST-4XF ), 00840331269734 (VESTT28-U8PVEST-4XM ), 00840331269727 (VESTT28-U8PVEST-5XF ), 00840331269734 (VESTT28-U8PVEST-5XM ), 00840331269727 (VESTT38-U8PVEST-6XF ), None (IVESTC8-U8PVEST-2XMR ), None (IVESTC8-U8PVEST-LMR ), None (IVESTC8-U8PVEST-MFR ), None (IVESTC8-U8PVEST-MMR ), None (IVESTC8-U8PVEST-SFR ), None (IVESTC8-U8PVEST-SMR ). 15. Model Number: U8QVEST; UDI-DI (Product Code): 00840331267556 (QVESTC18-U8QVEST-LF ), 00840331267563 (QVESTC18-U8QVEST-MF ), 00840331267594 (QVESTC18-U8QVEST-XLF ), 00840331267525 (QVESTC28-U8QVEST-3XF ), 00840331269826 (QVESTT18-U8QVEST-LF ), 00840331269826 (QVESTT18-U8QVEST-PLUS ), 00840331269826 (QVESTT28-U8QVEST-3XF ), 00840331296136 (QVESTT38-U8QVEST-8XF ), None (IQVESTC8-U8QVEST-LFR ). 16. Model Number: U8VEST; UDI-DI (Product Code): 00840331269215 (VESTC18-U8VEST-2XF ), 00840331269222 (VESTC18-U8VEST-2XM ), 00840331269253 (VESTC8-U8VEST-LF ), 00840331269260 (VESTC8-U8VEST-LM ), 00840331269277 (VESTC8-U8VEST-MF ), 00840331269284 (VESTC8-U8VEST-MM ), 00840331269291 (VESTC8-U8VEST-PLUS ), 00840331269307 (VESTC8-U8VEST-SF ), 00840331269314 (VESTC8-U8VEST-SM ), 00840331269321 (VESTC8-U8VEST-XLF ), 00840331269338 (VESTC8-U8VEST-XLM ), 00840331269352 (VESTC8-U8VEST-XSM ), 00840331269260 (VESTS8-U8VEST-LM ), 00840331269277 (VESTS8-U8VEST-MF ), 00840331269666 (VESTT18-U8VEST-LF ), 00840331269673 (VESTT18-U8VEST-LM ), 00840331269666 (VESTT18-U8VEST-MF ), 00840331269673 (VESTT18-U8VEST-MM ), 00840331269666 (VESTT18-U8VEST-PLUS ), 00840331269666 (VESTT18-U8VEST-SF ), 00840331269673 (VESTT18-U8VEST-SM ), 00840331269666 (VESTT18-U8VEST-XLF ), 00840331269673 (VESTT18-U8VEST-XLM ), 00840331269673 (VESTT18-U8VEST-XSM ), 00840331269666 (VESTT28-U8VEST-2XF ), 00840331269673 (VESTT28-U8VEST-2XM ), 00840331269666 (VESTT28-U8VEST-3XF ), 00840331269673 (VESTT28-U8VEST-3XM ), 00840331269666 (VESTT28-U8VEST-4XF ), 00840331269673 (VESTT28-U8VEST-4XM ), 00840331269666 (VESTT28-U8VEST-5XF ), 00840331269673 (VESTT28-U8VEST-5XM ), 00840331269666 (VESTT38-U8VEST-6XF ), 00840331269673 (VESTT38-U8VEST-6XM ), 00840331269666 (VESTT38-U8VEST-7XF ), 00840331269666 (VESTT38-U8VEST-7XM ), 00840331269666 (VESTT38-U8VEST-8XF ), 00840331269673 (VESTT38-U8VEST-8XM ), None (IVESTC8-U8VEST-2XMR ), None (IVESTC8-U8VEST-LMR ), None (IVESTC8-U8VEST-MMR ), None (IVESTC8-U8VEST-SFR ), None (IVESTC8-U8VEST-SMR ), None (IVESTC8-U8VEST-XLMR ), None (IVESTC8-U8VEST-XSFR ). 17. Model Number: U8ZVEST; UDI-DI (Product Code): 00840331269536 (ZVESTC18-U8ZVEST-MF ), 00840331269543 (ZVESTC18-U8ZVEST-MM ), 00840331269901 (ZVESTT28-U8ZVEST-5XM ).
Devices named in this recall
157 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- FVEST R8FVEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST R8FVEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST R8PVEST-XSMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST R8QVEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST R8VEST-XSMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- ZVEST R8ZVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- ZVEST R8ZVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- ZVEST R8ZVEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST S8FVEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- PVEST S8PVEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST S8QVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST S8QVEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST S8QVEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST S8QVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- QVEST S8QVEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- VEST S8VEST-XSMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- ZVEST S8ZVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- ZVEST S8ZVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST U8FVEST-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST U8FVEST-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST U8FVEST-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST U8FVEST-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- FVEST U8FVEST-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
Product lines
- VESTBurlington Medical, LLC · 48 models · 0 reports · 1 recall
- PVESTBurlington Medical, LLC · 42 models · 0 reports · 1 recall
- FVESTBurlington Medical, LLC · 38 models · 0 reports · 1 recall
- QVESTBurlington Medical, LLC · 18 models · 0 reports · 1 recall
- ZVESTBurlington Medical, LLC · 11 models · 0 reports · 1 recall
Other products in the same recall event
The FDA groups related products into one recall event (event 98415).
- Z-1753-2026 · Open, ClassifiedBurlington Medical, Thyroid Shield.
- Z-1755-2026 · Open, ClassifiedBurlington Medical, Kilt
- Z-1756-2026 · Open, ClassifiedBurlington Medical, Frontal Aprons
- Z-1757-2026 · Open, ClassifiedBurlington Medical, Wrap Aprons.
- Z-1758-2026 · Open, ClassifiedBurlington Medical, BAT (Breast, Axilla and Thyroid) Coverage.
- Z-1759-2026 · Open, ClassifiedBurlington Medical, Protective Sleeves.
- Z-1760-2026 · Open, ClassifiedBurlington Medical, Caps.
- Z-1761-2026 · Open, ClassifiedBurlington Medical, Leg Wraps.
- Z-1762-2026 · Open, ClassifiedBurlington Medical, Half Aprons.
- Z-1763-2026 · Open, ClassifiedBurlington Medical, Demi Half Aprons.
- Z-1764-2026 · Open, ClassifiedBurlington Medical, Blockers.
- Z-1765-2026 · Open, ClassifiedBurlington Medical, Kilt Blockers.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page