Burlington Medical, Kilt recallZ-1755-2026 · Burlington Medical, LLC · initiated Feb 11, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Burlington Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for attenuation degradation over time, decreasing the lifespan.
FDA root cause category: Process change control
What the company is doing
Burlington Medical notified consignees on about 02/11/2026 via "URGENT - FIELD SAFETY NOTICE (FSN)" letter. Consignees were instructed to ensure all users of the specific Burlington Medical radiation protection garments made aware of this Recall Notice, review inventory for affected units, arrange for the return of those affected units for proper disposal, and complete and return the Recall Response Form. Consignees were also instructed to notify appropriate individuals within your organization and to any organization where the potentially affected devices were transferred or sold.
Recall details
- Recall number
- Z-1755-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 11, 2026
- Posted by the FDA
- Apr 2, 2026
- Recalling firm
- Burlington Medical, LLC
- Quantity
- 15,216 units
- Product code
- IWO
- Distribution
- Worldwide distribution. Nationwide distribution in the US and international distribution to Canada, Australia, Bangladesh, Chile, Hong Kong, India, Mexico, Malaysia, Panama, Singapore, Thailand, Taiwan, Uruguay, and Vietnam.
Product description
Burlington Medical, Kilt
Code information (models, lots, serial numbers)
Show the full code information (13k characters)
All serial numbers are affected that were manufactured with Xenolite 800 attenuation material from 01/16/2025 to 01/21/2026. 1. Model Number: C8K303; UDI-DI (Product Code): 00840331297638 (FKILTC18-C8K303-2XM, FKILTC8-C8K303-LF, FKILTC8-C8K303-LM, FKILTC8-C8K303-MF, FKILTC8-C8K303-MM, FKILTC8-C8K303-PLUS, FKILTC8-C8K303-SF, FKILTC8-C8K303-SM, FKILTC8-C8K303-XLF, FKILTC8-C8K303-XLM, FKILTC8-C8K303-XSF, FKILTT18-C8K303-LM, FKILTT18-C8K303-MM, FKILTT18-C8K303-PLUS, FKILTT18-C8K303-SM, FKILTT18-C8K303-XLM, FKILTT18-C8K303-XSM, FKILTT28-C8K303-2XM, FKILTT28-C8K303-4XM, FKILTT28-C8K303-5XM). 2. Model Number: R8K303; UDI-DI (Product Code): 00197513003618 (FKILTC18-R8K303-2XF ), 00197513003618 (FKILTC18-R8K303-3XF ), 00197513002932 (FKILTC8-R8K303-LF ), 00197513002925 (FKILTC8-R8K303-LM ), 00197513002918 (FKILTC8-R8K303-MF ), 00197513002901 (FKILTC8-R8K303-MM ), 00197513002888 (FKILTC8-R8K303-SF ), 00197513002871 (FKILTC8-R8K303-SM ), 00197513002864 (FKILTC8-R8K303-XLF ), 00197513000068 (FKILTT18-R8K303-LF ), 00197513000198 (FKILTT18-R8K303-LM ), 00197513000068 (FKILTT18-R8K303-MF ), 00197513000198 (FKILTT18-R8K303-MM ), 00197513000068 (FKILTT18-R8K303-PLUS ), 00197513000068 (FKILTT18-R8K303-SF ), 00197513000198 (FKILTT18-R8K303-SM ), 00197513000068 (FKILTT18-R8K303-XLF ), 00197513000198 (FKILTT18-R8K303-XLM ), 00197513000198 (FKILTT18-R8K303-XSM ), 00197513000068 (FKILTT28-R8K303-2XF ), 00197513000068 (FKILTT28-R8K303-3XF ), 00197513000198 (FKILTT28-R8K303-3XM ), 00197513000068 (FKILTT28-R8K303-4XF ), 00197513000068 (FKILTT38-R8K303-6XF ). 3. Model Number: R8KILT; UDI-DI (Product Code): None (IKILTC8-R8KILT-LMR ), None (IKILTC8-R8KILT-SMR ), 00840331259681 (KILTC18-R8KILT-2XF ), 00840331259698 (KILTC18-R8KILT-2XM ), 00840331259704 (KILTC18-R8KILT-3XF ), 00840331259711 (KILTC18-R8KILT-3XM ), 00840331259728 (KILTC8-R8KILT-LF ), 00840331259735 (KILTC8-R8KILT-LM ), 00840331259735 (KILTC8-R8KILT-MF ), 00840331259759 (KILTC8-R8KILT-MM ), 00840331259766 (KILTC8-R8KILT-PLUS ), 00840331259773 (KILTC8-R8KILT-SF ), 00840331259780 (KILTC8-R8KILT-SM ), 00840331259797 (KILTC8-R8KILT-XLF ), 00840331259803 (KILTC8-R8KILT-XLM ), 00840331259810 (KILTC8-R8KILT-XSF ), 00840331259827 (KILTC8-R8KILT-XSM ), 00840331259728 (KILTS8-R8KILT-LF ), 00840331259735 (KILTS8-R8KILT-LM ), 00840331259742 (KILTS8-R8KILT-MF ), 00840331259759 (KILTS8-R8KILT-MM ), 00840331259766 (KILTS8-R8KILT-PLUS ), 00840331259773 (KILTS8-R8KILT-SF ), 00840331259780 (KILTS8-R8KILT-SM ), 00840331259797 (KILTS8-R8KILT-XLF ), 00840331259803 (KILTS8-R8KILT-XLM ), 00840331269925 (KILTT18-R8KILT-LF ), 00840331269932 (KILTT18-R8KILT-LM ), 00840331269925 (KILTT18-R8KILT-MF ), 00840331269932 (KILTT18-R8KILT-MM ), 00840331269925 (KILTT18-R8KILT-PLUS ), 00840331269925 (KILTT18-R8KILT-SF ), 00840331269932 (KILTT18-R8KILT-SM ), 00840331269925 (KILTT18-R8KILT-XLF ), 00840331269932 (KILTT18-R8KILT-XLM ), 00840331269925 (KILTT18-R8KILT-XSF ), 00840331269932 (KILTT18-R8KILT-XSM ), 00840331269925 (KILTT28-R8KILT-2XF ), 00840331269932 (KILTT28-R8KILT-2XM ), 00840331269925 (KILTT28-R8KILT-3XF ), 00840331269932 (KILTT28-R8KILT-3XM ), 00840331269925 (KILTT28-R8KILT-4XF ), 00840331269932 (KILTT28-R8KILT-4XM ), 00840331269925 (KILTT28-R8KILT-5XF ), 00840331269932 (KILTT28-R8KILT-5XM ), 00840331269925 (KILTT38-R8KILT-6XF ), 00840331269932 (KILTT38-R8KILT-6XM ), 00840331269925 (KILTT38-R8KILT-7XF ), 00840331269932 (KILTT38-R8KILT-7XM ), 00840331269925 (KILTT38-R8KILT-8XF ), 00840331269932 (KILTT38-R8KILT-8XM ), None (IKILTC8-R8KILT-LFL ), None (IKILTC8-R8KILT-LFR ), None (IKILTC8-R8KILT-MFL ), None (IKILTC8-R8KILT-MFR ), None (IKILTC8-R8KILT-MFXL ), None (IKILTC8-R8KILT-MML ), None (IKILTC8-R8KILT-MMR ), None (IKILTC8-R8KILT-SFL ), None (IKILTC8-R8KILT-SML ), None (IKILTC8-R8KILT-XSFR ), None (IKILTC8-R8KILT-2XFL ), None (IKILTC8-R8KILT-2XFR ), None (IKILTC8-R8KILT-3XFR ), None (IKILTC8-R8KILT-LFS ), None (IKILTC8-R8KILT-LML ), None (IKILTC8-R8KILT-LMXL ), None (IKILTC8-R8KILT-MFS ), None (IKILTC8-R8KILT-SFS ), None (IKILTC8-R8KILT-XLMR ), None (IKILTC8-R8KILT-2XMXL ), None (IKILTC8-R8KILT-PLUSL ), None (IKILTC8-R8KILT-SFR ), None (IKILTC8-R8KILT-XLML ), None (IKILTC8-R8KILT-XSFS ), None (IKILTC8-R8KILT-MMXL ), None (IKILTC8-R8KILT-XLMXL ), None (IKILTC8-R8KILT-2XMR ), None (IKILTC8-R8KILT-XSML ), None (IKILTC8-R8KILT-XSMR ), None (IKILTC8-R8KILT-XSMXL ), None (IFKILTC8-R8K303-LFR), None (IFKILTC8-R8K303-MMR). 4. Model Number: R8S302; UDI-DI (Product Code): 00197513004950 (SKIRTWBC8-R8S302-LF ), 00197513004943 (SKIRTWBC8-R8S302-LM ), 00197513004936 (SKIRTWBC8-R8S302-MF ), 00197513004929 (SKIRTWBC8-R8S302-MM ), 00197513004912 (SKIRTWBC8-R8S302-PLUS ), 00197513004905 (SKIRTWBC8-R8S302-SF ), 00197513004899 (SKIRTWBC8-R8S302-SM ), 00197513004875 (SKIRTWBC8-R8S302-XLM ), 00197513000327 (SKIRTWBT18-R8S302-LF ), 00197513000457 (SKIRTWBT18-R8S302-LM ), 00197513000327 (SKIRTWBT18-R8S302-MF ), 00197513000457 (SKIRTWBT18-R8S302-MM ), 00197513000327 (SKIRTWBT18-R8S302-PLUS ), 00197513000457 (SKIRTWBT18-R8S302-SM ), 00197513000327 (SKIRTWBT18-R8S302-XLF ), 00197513000457 (SKIRTWBT18-R8S302-XLM ), 00197513000457 (SKIRTWBT28-R8S302-2XM ), 00197513000327 (SKIRTWBT28-R8S302-3XF ), 00197513000457 (SKIRTWBT28-R8S302-3XM ), 00197513000457 (SKIRTWBT28-R8S302-4XM ), 00197513000327 (SKIRTWBT28-R8S302-5XF ), 00197513000457 (SKIRTWBT28-R8S302-5XM ), 00197513000327 (SKIRTWBT38-R8S302-6XF ), 00197513000457 (SKIRTWBT38-R8S302-6XM ), 00197513000327 (SKIRTWBT38-R8S302-8XF ), 00197513000457 (SKIRTWBT38-R8S302-8XM ), None (ISKIRTWBC8-R8S302-MFL ), None (ISKIRTWBC8-R8S302-MFR ), None (ISKIRTWBC8-R8S302-MML ), None (ISKIRTWBC8-R8S302-MMR ), None (ISKIRTWBC8-R8S302-MMXL ), None (ISKIRTWBC8-R8S302-PLUSL ), None (ISKIRTWBC8-R8S302-SFL ), None (ISKIRTWBC8-R8S302-SML ). 5. Model Number: R8SKIRT; UDI-DI (Product Code): 00840331260182 (SKIRTC8-R8SKIRT-LM ), 00840331260205 (SKIRTC8-R8SKIRT-MM ), 00840331260250 (SKIRTC8-R8SKIRT-XLM ), 00840331269987 (SKIRTT18-R8SKIRT-LF ), 00840331269994 (SKIRTT18-R8SKIRT-LM ), 00840331269987 (SKIRTT18-R8SKIRT-MF ), 00840331269994 (SKIRTT18-R8SKIRT-MM ), 00840331269987 (SKIRTT18-R8SKIRT-PLUS ), 00840331269994 (SKIRTT18-R8SKIRT-SM ), 00840331269987 (SKIRTT18-R8SKIRT-XLF ), 00840331269994 (SKIRTT18-R8SKIRT-XLM ), 00840331269994 (SKIRTT18-R8SKIRT-XSM ), 00840331269987 (SKIRTT28-R8SKIRT-2XF ), 00840331269994 (SKIRTT28-R8SKIRT-2XM ), 00840331269994 (SKIRTT28-R8SKIRT-3XM ), 00840331269994 (SKIRTT28-R8SKIRT-4XM ), 00840331270013 (SKIRTT38-R8SKIRT-6XM ). 6. Model Number: S8K303; UDI-DI (Product Code): 00197513002802 (FKILTC8-S8K303-LM ), 00197513002796 (FKILTC8-S8K303-MF ), 00197513002789 (FKILTC8-S8K303-MM ), 00197513002734 (FKILTC8-S8K303-XLM ), 00197513002727 (FKILTC8-S8K303-XSF ), 00197513000051 (FKILTT18-S8K303-MF ), 00197513000181 (FKILTT18-S8K303-MM ), 00197513000051 (FKILTT18-S8K303-PLUS ), 00197513000181 (FKILTT18-S8K303-SM ), 00197513000181 (FKILTT18-S8K303-XSM ), 00197513000181 (FKILTT28-S8K303-2XM ), 00197513000051 (FKILTT28-S8K303-3XF ), 00197513000181 (FKILTT28-S8K303-3XM ), 00197513000181 (FKILTT28-S8K303-4XM ). 7. Model Number: S8KILT; UDI-DI (Product Code): 00840331263435 (KILTC18-S8KILT-2XF ), 00840331263442 (KILTC18-S8KILT-2XM ), 00840331263459 (KILTC18-S8KILT-3XF ), 00840331263466 (KILTC18-S8KILT-3XM ), 00840331263473 (KILTC8-S8KILT-LF ), 00840331263480 (KILTC8-S8KILT-LM ), 00840331263497 (KILTC8-S8KILT-MF ), 00840331263503 (KILTC8-S8KILT-MM ), 00840331263510 (KILTC8-S8KILT-PLUS ), 00840331263527 (KILTC8-S8KILT-SF ), 00840331263534 (KILTC8-S8KILT-SM ), 00840331263541 (KILTC8-S8KILT-XLF ), 00840331263558 (KILTC8-S8KILT-XLM ), 00840331263565 (KILTC8-S8KILT-XSF ), 00840331263565 (KILTC8-S8KILT-XSM ), 00840331263473 (KILTS8-S8KILT-LF ), 00840331263480 (KILTS8-S8KILT-LM ), 00840331263497 (KILTS8-S8KILT-MF ), 00840331263503 (KILTS8-S8KILT-MM ), 00840331263527 (KILTS8-S8KILT-SF ), 00840331263534 (KILTS8-S8KILT-SM ), 00840331263541 (KILTS8-S8KILT-XLF ), 00840331263558 (KILTS8-S8KILT-XLM ), 00840331269949 (KILTT18-S8KILT-LF ), 00840331269956 (KILTT18-S8KILT-LM ), 00840331269949 (KILTT18-S8KILT-MF ), 00840331269956 (KILTT18-S8KILT-MM ), 00840331269949 (KILTT18-S8KILT-PLUS ), 00840331269949 (KILTT18-S8KILT-SF ), 00840331269956 (KILTT18-S8KILT-SM ), 00840331269949 (KILTT18-S8KILT-XLF ), 00840331269956 (KILTT18-S8KILT-XLM ), 00840331269949 (KILTT18-S8KILT-XSF ), 00840331269956 (KILTT18-S8KILT-XSM ), 00840331269949 (KILTT28-S8KILT-2XF ), 00840331269956 (KILTT28-S8KILT-2XM ), 00840331269949 (KILTT28-S8KILT-3XF ), 00840331269956 (KILTT28-S8KILT-3XM ), 00840331269949 (KILTT28-S8KILT-4XF ), 00840331269956 (KILTT28-S8KILT-4XM ), 00840331269949 (KILTT28-S8KILT-5XF ), 00840331269956 (KILTT28-S8KILT-5XM ), 00840331269949 (KILTT38-S8KILT-6XF ), 00840331269956 (KILTT38-S8KILT-6XM ), 00840331269949 (KILTT38-S8KILT-7XF ), 00840331269956 (KILTT38-S8KILT-7XM ), 00840331269949 (KILTT38-S8KILT-8XF ), 00840331269956 (KILTT38-S8KILT-8XM ), None (IKILTC8-S8KILT-LMR ), None (IKILTC8-S8KILT-MMR ). 8. Model Number: S8S302; UDI-DI (Product Code): 00197513004837 (SKIRTWBC8-S8S302-LF ), 00197513004820 (SKIRTWBC8-S8S302-LM ), 00197513004813 (SKIRTWBC8-S8S302-MF), 00197513004806 (SKIRTWBC8-S8S302-MM ), 00197513004790 (SKIRTWBC8-S8S302-PLUS), 00197513000310 (SKIRTWBT18-S8S302-LF ), 00197513000440 (SKIRTWBT18-S8S302-LM), 00197513000310 (SKIRTWBT18-S8S302-MF ), 00197513000440 (SKIRTWBT18-S8S302-MM), 00197513000310 (SKIRTWBT18-S8S302-PLUS ), 00197513000440 (SKIRTWBT18-S8S302-SM ), 00197513000310 (SKIRTWBT18-S8S302-XLF ), 00197513000440 (SKIRTWBT18-S8S302-XLM ), 00197513000440 (SKIRTWBT18-S8S302-XSM ), 00197513000310 (SKIRTWBT28-S8S302-3XF ), 00197513000440 (SKIRTWBT28-S8S302-3XM), 00197513000440 (SKIRTWBT28-S8S302-4XM ), 00197513000440 (SKIRTWBT28-S8S302-5XM ). 9. Model Number: S8SKIRT; UDI-DI (Product Code): 00840331270006 (SKIRTT18-S8SKIRT-LF ), 00840331270013 (SKIRTT18-S8SKIRT-LM ), 00840331270006 (SKIRTT18-S8SKIRT-MF ), 00840331270006 (SKIRTT18-S8SKIRT-PLUS ), 00840331270013 (SKIRTT18-S8SKIRT-SM ), 00840331270006 (SKIRTT18-S8SKIRT-XLF ), 00840331270006 (SKIRTT28-S8SKIRT-3XF ), 00840331270013 (SKIRTT28-S8SKIRT-4XM ). 10. Model Number: U8K303; UDI-DI (Product Code): 00197513002680 (FKILTC8-U8K303-LM), 00197513002673 (FKILTC8-U8K303-MF ), 00197513002666 (FKILTC8-U8K303-MM ), 00197513002642 (FKILTC8-U8K303-SF ), 00197513002635 (FKILTC8-U8K303-SM ), 00197513000174 (FKILTT18-U8K303-LM ), 00197513000044 (FKILTT18-U8K303-MF ), 00197513000174 (FKILTT18-U8K303-MM ), 00197513000044 (FKILTT18-U8K303-PLUS ), 00197513000174 (FKILTT18-U8K303-SM ), 00197513000174 (FKILTT28-U8K303-2XM ), 00197513000044 (FKILTT28-U8K303-4XF ), 00197513000174 (FKILTT28-U8K303-4XM ), 00197513000044 (FKILTT28-U8K303-5XF ), 00197513000174 (FKILTT28-U8K303-5XM ), None (IFKILTC8-U8K303-2XMR ), None (IFKILTC8-U8K303-LMR ), None (IFKILTC8-U8K303-MMR ), None (IFKILTC8-U8K303-SFR ), None (IFKILTC8-U8K303-SMR ), None (IFKILTC8-U8K303-XLMR ). 11. Model Number: U8KILT; UDI-DI (Product Code): 00840331267167 (KILTC18-U8KILT-2XF ), 00840331267174 (KILTC18-U8KILT-2XM ), 00840331267181 (KILTC18-U8KILT-3XF ), 00840331267204 (KILTC8-U8KILT-LF ), 00840331267211 (KILTC8-U8KILT-LM ), 00840331267228 (KILTC8-U8KILT-MF ), 00840331267235 (KILTC8-U8KILT-MM ), 00840331267242 (KILTC8-U8KILT-PLUS ), 00840331267259 (KILTC8-U8KILT-SF ), 00840331267266 (KILTC8-U8KILT-SM ), 00840331267273 (KILTC8-U8KILT-XLF ), 00840331267280 (KILTC8-U8KILT-XLM ), 00840331267297 (KILTC8-U8KILT-XSF ), 00840331267303 (KILTC8-U8KILT-XSM ), 00840331267204 (KILTS8-U8KILT-LF ), 00840331267211 (KILTS8-U8KILT-LM ), 00840331267228 (KILTS8-U8KILT-MF ), 00840331269963 (KILTT18-U8KILT-LF ), 00840331269970 (KILTT18-U8KILT-LM ), 00840331269963 (KILTT18-U8KILT-MF ), 00840331269970 (KILTT18-U8KILT-MM ), 00840331269963 (KILTT18-U8KILT-PLUS ), 00840331269963 (KILTT18-U8KILT-SF ), 00840331269970 (KILTT18-U8KILT-SM ), 00840331269963 (KILTT18-U8KILT-XLF ), 00840331269970 (KILTT18-U8KILT-XLM ), 00840331269970 (KILTT18-U8KILT-XSM ), 00840331269963 (KILTT28-U8KILT-2XF ), 00840331269970 (KILTT28-U8KILT-2XM ), 00840331269963 (KILTT28-U8KILT-3XF ), 00840331269970 (KILTT28-U8KILT-3XM ), 00840331269963 (KILTT28-U8KILT-4XF ), 00840331269970 (KILTT28-U8KILT-4XM ), 00840331269963 (KILTT28-U8KILT-5XF ), 00840331269970 (KILTT28-U8KILT-5XM ), 00840331269963 (KILTT38-U8KILT-6XF ), 00840331269970 (KILTT38-U8KILT-6XM ), 00840331269963 (KILTT38-U8KILT-7XF ), 00840331269970 (KILTT38-U8KILT-7XM ), 00840331269963 (KILTT38-U8KILT-8XF ), 00840331269970 (KILTT38-U8KILT-8XM ), None (IKILTC8-U8KILT-2XMR ), None (IKILTC8-U8KILT-LFR ), None (IKILTC8-U8KILT-LMR ), None (IKILTC8-U8KILT-LMXL ), None (IKILTC8-U8KILT-MFR ), None (IKILTC8-U8KILT-MMR ), None (IKILTC8-U8KILT-SFR ), None (IKILTC8-U8KILT-SMR ), None (IKILTC8-U8KILT-XLMR ), None (IKILTC8-U8KILT-XLMXL ), None (IKILTC8-U8KILT-XSFR ). 12. Model Number: U8S302; UDI-DI (Product Code): 00197513004714 (SKIRTWBC8-U8S302-LF ), 00197513004691 (SKIRTWBC8-U8S302-MF ), 00197513004684 (SKIRTWBC8-U8S302-MM ), 00197513004677 (SKIRTWBC8-U8S302-PLUS ), 00197513004660 (SKIRTWBC8-U8S302-SF ), 00197513004622 (SKIRTWBC8-U8S302-XSF ), 00197513000303 (SKIRTWBT18-U8S302-PLUS ).
Devices named in this recall
105 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- K303 U8K303*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8K303-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 S8K303-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 R8K303-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- K303 U8S302-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 U8S302-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 S8S302-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- S302 R8S302-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT-XSMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- SKIRT R8SKIRT-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- SKIRT R8SKIRT-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- SKIRT R8SKIRT-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-3XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-2XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-2XMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-3XFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-LFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-LMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-MFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-MMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-PLUSBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-SFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-SMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-XLFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-XLMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-XSFBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT-XSMBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT R8KILT*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT S8KILT*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT*FBurlington Medical, LLC · Apron, Protective1 recall0 reports
- KILT U8KILT*MBurlington Medical, LLC · Apron, Protective1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98415).
- Z-1753-2026 · Open, ClassifiedBurlington Medical, Thyroid Shield.
- Z-1754-2026 · Open, ClassifiedBurlington Medical, Vest
- Z-1756-2026 · Open, ClassifiedBurlington Medical, Frontal Aprons
- Z-1757-2026 · Open, ClassifiedBurlington Medical, Wrap Aprons.
- Z-1758-2026 · Open, ClassifiedBurlington Medical, BAT (Breast, Axilla and Thyroid) Coverage.
- Z-1759-2026 · Open, ClassifiedBurlington Medical, Protective Sleeves.
- Z-1760-2026 · Open, ClassifiedBurlington Medical, Caps.
- Z-1761-2026 · Open, ClassifiedBurlington Medical, Leg Wraps.
- Z-1762-2026 · Open, ClassifiedBurlington Medical, Half Aprons.
- Z-1763-2026 · Open, ClassifiedBurlington Medical, Demi Half Aprons.
- Z-1764-2026 · Open, ClassifiedBurlington Medical, Blockers.
- Z-1765-2026 · Open, ClassifiedBurlington Medical, Kilt Blockers.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page