SOLTIVE recallZ-1695-2026 · Olympus Corporation of the Americas · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
FDA root cause category: Process control
What the company is doing
On February 27, 2026 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Examine your inventory and identify any of the affected devices listed in the attachment. 2. Olympus requests that you acknowledge receipt of this letter through our recall web portal. 3. Olympus representatives will begin contacting customers to coordinate the service of affected unit(s). In the meantime, you may continue using your SOLTIVE Laser until the service is completed. Olympus will replace the 24V power supply module at no charge to you. Note: Any issues beyond the replacement of the 24V module in this corrective action will follow the standard SOLTIVE Laser repair process and your service contract, if applicable. 4. If you have further distributed the affected products, please forward this notice to other users who may be impacted.
Recall details
- Recall number
- Z-1695-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Mar 30, 2026
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 32 units
- Product code
- GEX
- FDA submissions
- K242191
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.
Product description
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Code information (models, lots, serial numbers)
Laser System: TFL-SLS; UDI: 00821925044135; Component Part: TFL-CSLU; UDI: 00821925044593; Serial #: MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164, MDUF230166, MDUF230167, MDUF230234, MDUF230283, MDUF230284, MDUF230331, MDUF230332, MDUF230361, MDUF230362, MDUF230363, MDUF230364, MDUF230365, MDUF230366, MDUF230411, MDUF230412, MDUF230413;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- SOLTIVE TFL-CSLUGyrus ACMI, LLC · Powered laser surgical instrument2 recalls0 reports
Product lines
Other devices under the same FDA submission
14 devices cleared under K242191 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- SOLTIVE TFL-FBX200BSGyrus ACMI, LLC · Powered Laser Surgical Instrument4 recalls84 reports
- SOLTIVE TFL-FBX200SGyrus ACMI, LLC · Powered Laser Surgical Instrument3 recalls93 reports
- SOLTIVE TFL-FBX550SGyrus ACMI, LLC · Powered laser surgical instrument3 recalls63 reports
- SOLTIVE TFL-FBX150BSGyrus ACMI, LLC · Powered Laser Surgical Instrument3 recalls18 reports
- SOLTIVE TFL-FBX150SGyrus ACMI, LLC · Powered Laser Surgical Instrument2 recalls21 reports
- SOLTIVE TFL-FBX365SGyrus ACMI, LLC · Powered Laser Surgical Instrument2 recalls11 reports
- Soltive TFL-CPLUGyrus ACMI, LLC · Powered laser surgical instrument2 recalls2 reports
- SOLTIVE TFL-FBX940SGyrus ACMI, LLC · Powered Laser Surgical Instrument2 recalls1 report
- SOLTIVE TFL-AFSWLGyrus ACMI, LLC · Powered laser surgical instrument83 reports
- SOLTIVE TFL-AFSWGyrus ACMI, LLC · Powered laser surgical instrument1 recall5 reports
- SOLTIVE TFL-FBX940RGyrus ACMI, LLC · Powered Laser Surgical Instrument1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98422).
- Z-1694-2026 · Open, ClassifiedBrand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Descriptio…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page