NAMIC recallZ-1713-2026 · Medline Industries, LP · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1713-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Apr 7, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 192690 units
- Product code
- DXT
- FDA submissions
- K113198
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Product description
NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
1. Medline SKU 70083007, UDI/DI each 10193489065855, UDI/DI case 30193489065859, Lot Numbers: 0000077029, 0000081537, 0000084335, 0000087347, 0000089563, 0000113792, 0000115753, 0000122956, 0000127785, 0000131986, 0000158748, 0000175535, 0000187926; 2. Medline SKU 70084007, UDI/DI each 10193489065893, UDI/DI case 30193489065897, Lot numbers: 0000078186, 0000081539, 0000084274, 0000086756, 0000089554, 0000094842, 0000096918, 0000105613, 0000109479, 0000111244, 0000114562, 0000115421, 0000117317, 0000118251, 0000120265, 0000120769, 0000121714, 0000123466, 0000124589, 0000131984, 0000131989, 0000138071, 0000140322, 0000142850, 0000147002, 0000152069, 0000162694, 0000173507, 0000174662, 0000174773, 0000176555, 0000178806, 0000189895; 3. Medline SKU 70084017, UDI/DI each 10193489065916, UDI/DI case 20193489065913, Lot Numbers: 0000084509, 0000086823, 0000095687, 0000165551, 0000165868, 0000166300; 4. Medline SKU 80083007, UDI/DI each 10193489066753, UDI/DI case 30193489066757, Lot Numbers: 0000097388, 0000127072, 0000135012, 0000137004; 5. Medline SKU 80084007, UDI/DI each 10193489066791, UDI/DI case 30193489066795, Lot Numbers: 0000080553, 0000080919, 0000082969, 0000085041, 0000088044, 0000094514, 0000099870, 0000101882, 0000102372, 0000105155, 0000107328, 0000109442, 0000112810, 0000112966, 0000114317, 0000115744, 0000117572, 0000118722, 0000118962, 0000119950, 0000120931, 0000121305, 0000121393, 0000122583, 0000122606, 0000123102, 0000126198, 0000126566, 0000127132, 0000130798, 0000131678, 0000132838, 0000137176, 0000139543, 0000140507, 0000142057, 0000142279, 0000143110, 0000143264, 0000143714, 0000148872, 0000150059, 0000150488, 0000151054, 0000151295, 0000151300, 0000151768, 0000152754, 0000153438, 0000154148, 0000156028, 0000158045, 0000158122, 0000159113, 0000159652, 0000160479, 0000160753, 0000161269, 0000162083, 0000162090, 0000162336, 0000162895, 0000164540, 0000164541, 0000166504, 0000166983, 0000167995, 0000168000, 0000169147, 0000169560, 0000169793, 0000170823, 0000171778, 0000173402, 0000174087, 0000174697, 0000174747, 0000176011, 0000176240, 0000176850, 0000177923, 0000179193, 0000179414, 0000180511, 0000182186, 0000182779, 0000182782, 0000185643, 0000186234, 0000186874, 0000187362, 0000187363, 0000188837, 0000188842, 0000188845; 6. Medline SKU 80084017, UDI/DI each 10193489066814, UDI/DI case 20193489066811, Lot Numbers: 0000101535, 0000103862, 0000105095, 0000106954, 0000108237, 0000122887, 0000126590, 0000127159, 0000128360, 0000128852, 0000132717, 0000134047, 0000135616, 0000140813, 0000144899, 0000148530, 0000149322, 0000152430, 0000161279, 0000164538, 0000169107, 0000173193, 0000173256, 0000182141, 0000183619, 0000187345, 0000188399.
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- NAMIC 70083007Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
- NAMIC 70084007Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
- NAMIC 70084017Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
- NAMIC 80083007Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
- NAMIC 80084007Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
- NAMIC 80084017Medline Industries, Inc. · Injector and syringe, angiographic1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98598).
- Z-1712-2026 · Open, ClassifiedNAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,T…
- Z-1714-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline S…
- Z-1715-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medl…
- Z-1716-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 0…
- Z-1718-2026 · Open, ClassifiedNAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG…
- Z-1719-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU …
- Z-1720-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: SPECIAL PROCEDURE TRAY, Medline SK…
- Z-1721-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. HEAD AND NECK CDS-LF, Medline S…
- Z-1722-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC,…
- Z-1723-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU …
- Z-1724-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU…
- Z-1725-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: GLOVE PACK 7.0 (15PK), Medline SKU…
- Z-1726-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ5…
- Z-1727-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. DELIVERY PACK, Medline SKU # DY…
- Z-1728-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYN…
- Z-1729-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
- Z-1730-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. SHOULDER ARTHROSCOPY PACK, Medl…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page