NAMIC recallZ-1724-2026 · Medline Industries, LP · initiated Feb 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1724-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 27, 2026
- Posted by the FDA
- Apr 7, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 79843 units
- Product code
- LRO
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Product description
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU # 60131209; 4. MTS LHK, Medline SKU # 60160075; 5. MTS LHK, Medline SKU # 60161954; 6. MTO LHK, Medline SKU # 60230148; 7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I; 8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686; 9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M; 10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K; 11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O; 12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P; 13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N; 14. CATH PACK-LF, Medline SKU # DYNJ0803367J; 15. E.P. TRAY, Medline SKU # DYNJ19343J; 16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T; 17. CATH LAB TRAY, Medline SKU # DYNJ24250G; 18. CATH LAB PACK-LF, Medline SKU # DYNJ38333D; 19. CATH LAB PACK-LF, Medline SKU # DYNJ38333F; 20. EP IMPLANT PACK-LF, Medline SKU # DYNJ38910D; 21. ADULT CATH LAB PACK, Medline SKU # DYNJ40380J; 22. CRANIOTOMY PACK-LF, Medline SKU # DYNJ43403D; 23. LAP CHOLE PACK, Medline SKU # DYNJ48487F; 24. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779G; 25. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779I; 26. HYBRID, Medline SKU # DYNJ57049N; 27. SPECIALS PK, Medline SKU # DYNJ57713C; 28. SPECIALS PK, Medline SKU # DYNJ57713D; 29. SPECIALS PK, Medline SKU # DYNJ57713F; 30. CARDIAC CATH LAB, Medline SKU # DYNJ57748A; 31. TAVR PACK, Medline SKU # DYNJ58771G; 32. CATH PACK GWV, Medline SKU # DYNJ59229F; 33. HEART CATH PACK, Medline SKU # DYNJ61294F; 34. CATH LAB PACK, Medline SKU # DYNJ61353C; 35. DEVICE IMPLANT PACK, Medline SKU # DYNJ64636A; 36. CUSTOM PROCEDURE KIT, Medline SKU # DYNJ65277A; 37. GENERAL VASCULAR PACK RFID-V2, Medline SKU # DYNJ66442F; 38. CARDIAC CATH PACK, Medline SKU # DYNJ69845C; 39. OMF MINOR PACK, Medline SKU # DYNJ83991C; 40. HEART CATH PACK, Medline SKU # DYNJ84133B; 41. SPECIAL PROCEDURES, Medline SKU # DYNJ85425B; 42. PRIMARY PACK, Medline SKU # DYNJT3093; 43. CARDIAC CATH PACK, Medline SKU # DYNJT3522A; 44. CARDIAC CATH PACK, Medline SKU # DYNJT3522B; 45. TOTAL KNEE PACK, Medline SKU # DYNJT4069; 46. CARDIAC CATH PACK, Medline SKU # DYNJT5685; 47. CATH PACK, Medline SKU # DYNJT6544; 48. PACK CARDIAC CATH, Medline SKU # DYNJT7591; 49. OBL CATH VEIN PACK, Medline SKU # DYNJT7863; 50. TUMESCENT SYRINGE KIT, Medline SKU # DYNJTUMSYR; 51. CSTM LH KIT, Medline SKU # VASC1054; 52. LHK, Medline SKU # VASC1130A; 53. MANIFOLD KIT, Medline SKU # VASC1137A; 54. LHK, Medline SKU # VASC1299A; 55. LHK, Medline SKU # VASC1399B; 56. LHK, Medline SKU # VASC1403; 57. IR 2 PORT, Medline SKU # VASC1404; 58. LHK, Medline SKU # VASC1405; 59. LHK, Medline SKU # VASC1455A; 60. TWO PORT MANIFOLD KIT, Medline SKU # VASC1496; 61. RHK, US, NAMIC MANIFOLD, Medline SKU # VASC1502; 62. LHK, Medline SKU # VASC1555; 63. 4 VALVE,PERC,MANIFOLD,KIT, Medline SKU # VASC1609; 64. LHK, Medline SKU # VASC1668; 65. LHK, Medline SKU # VASC1694; 66. LHK, Medline SKU # VASC1805; 67. TVR, Medline SKU # VASC1816.
Code information (models, lots, serial numbers)
Show the full code information (34k characters)
Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26BBI625; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 26ABB718; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25JBJ140; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25HBF702; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25GBC747; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25DBR341; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25CBK686; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 25ABR102; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24IBK765; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24GBS605; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24EBO156; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24BBG055; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 24ABH036; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23LBI829; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23KBN009; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23FBO996; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23FBA259; Medline SKU # 60010508, UDI/DI each 10193489040340, UDI/DI case 40193489040341, Lot Number: 23EBD151; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24IBE239; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24HBR112; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24GBY032; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24BBO470; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 24ABB436; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23LBA225; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23GBK257; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23FBG433; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23FBC618; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23EBX036; Medline SKU # 60100701, UDI/DI each 10193489045246, UDI/DI case 40193489045247, Lot Number: 23DBD220; Medline SKU # 60131209, UDI/DI each 10193489046779, UDI/DI case 40193489046770, Lot Number: 24HBJ211; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24EBQ349; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24CBD613; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 24ABL640; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23KBB857; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23IBC156; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23FBJ161; Medline SKU # 60160075, UDI/DI each 10193489048872, UDI/DI case 40193489048873, Lot Number: 23FBC907; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 26ABD974; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25JBM213; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25FBK284; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 25FBK967; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24KBN942; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24JBC074; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24HBQ995; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24GBP001; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24CBF239; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 24ABJ769; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23LBP098; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23HBB863; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23GBQ587; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23FBG492; Medline SKU # 60161954, UDI/DI each 10193489049091, UDI/DI case 40193489049092, Lot Number: 23CBR482; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25CBJ809; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25BBS609; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 25BBL037; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24KBI829; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24JBX172; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24IBO369; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24EBI926; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24DBV041; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 24ABB294; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23KBT232; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23KBA553; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23IBD723; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23HBV742; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23HBU186; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23GBA723; Medline SKU # 60230148, UDI/DI each 10193489052985, UDI/DI case 40193489052986, Lot Number: 23DBF014; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 26BLA621; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25LLA491; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25JLA720; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25JLA241; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25HLB071; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25GLA678; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ELA960; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ELA670; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25DLA414; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25BLA539; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 25ALA504; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24JLA106; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24JLA071; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24ILA631; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24HLA635; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24HLA514; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24GLA105; Medline SKU # 00-500510I, UDI/DI each 10195327695361, UDI/DI case 40195327695362, Lot Number: 24FLA889; Medline SKU # BIOB686, UDI/DI each 10193489684025, UDI/DI case 40193489684026, Lot Number: 25IMF801; Medline SKU # BIOB686, UDI/DI each 10193489684025, UDI/DI case 40193489684026, Lot Number: 24FMH284; Medline SKU # DYNJ0376145M, UDI/DI each 10889942941643, UDI/DI case 40889942941644, Lot Number: 24ILA248; Medline SKU # DYNJ0392706K, UDI/DI each 10195327588212, UDI/DI case 40195327588213, Lot Number: 25LBP577; Medline SKU # DYNJ0392706K, UDI/DI each 10195327588212, UDI/DI case 40195327588213, Lot Number: 25IBR953; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25FBF110; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25EBI930; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 25DBE958; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 24JBN523; Medline SKU # DYNJ0451426O, UDI/DI each 10195327698690, UDI/DI case 40195327698691, Lot Number: 24IBC823; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26BBK812; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26ABI992; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 26ABJ025; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25KBD666; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25JBR728; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25JBR851; Medline SKU # DYNJ0451426P, UDI/DI each 10198459386336, UDI/DI case 40198459386337, Lot Number: 25GBS862; Medline SKU # DYNJ07953N, UDI/DI each 10198459123412, UDI/DI case 40198459123413, Lot Number: 25IBK486; Medline SKU # DYNJ07953N, UDI/DI each 10198459123412, UDI/DI case 40198459123413, Lot Number: 25IBF054; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25KBI063; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25IBK151; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25EBD486; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25DBL885; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 25CBL230; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24LBG685; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24KBO528; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24JBH940; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24HBQ365; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24HBJ343; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24GBE178; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24FBN018; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24EBR279; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24CBP560; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 24ABP475; Medline SKU # DYNJ0803367J, UDI/DI each 10195327542276, UDI/DI case 40195327542277, Lot Number: 23LBS358; Medline SKU # DYNJ19343J, UDI/DI each 10888277827981, UDI/DI case 40888277827982, Lot Number: 24KBL564; Medline SKU # DYNJ20898T, UDI/DI each 10198459477003, UDI/DI case 40198459477004, Lot Number: 25KMH265; Medline SKU # DYNJ20898T, UDI/DI each 10198459477003, UDI/DI case 40198459477004, Lot Number: 25JML139; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC371; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 26AMC433; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25KBH544; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25JBG417; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBQ931; Medline SKU # DYNJ24250G, UDI/DI each 10198459416590, UDI/DI case 40198459416591, Lot Number: 25GBI702; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 25AMG765; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24KMJ018; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24JME459; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24HMB740; Medline SKU # DYNJ38333D, UDI/DI each 10198459032981, UDI/DI case 40198459032982, Lot Number: 24GMH056; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25KME742; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25HME818; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25EMB706; Medline SKU # DYNJ38333F, UDI/DI each 10198459258633, UDI/DI case 40198459258634, Lot Number: 25CME621; Medline SKU # DYNJ38910D, UDI/DI each 10198459319815, UDI/DI case 40198459319816, Lot Number: 25GBJ993; Medline SKU # DYNJ38910D, UDI/DI each 10198459319815, UDI/DI case 40198459319816, Lot Number: 25EBD960; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 26ABR937; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25KBI764; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25JBR971; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBM264; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBI796; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25IBD184; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBV148; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBR538; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25FBB566; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25EBN741; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25EBF462; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25DBK395; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBU744; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBJ971; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25CBH568; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 25ABK344; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24KBG845; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24IBM450; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24IBH804; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24HBL014; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24GBU989; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24EBQ673; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24EBO710; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24DBQ947; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24CBH211; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24BBE538; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24ABU829; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 24ABA618; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23KBO467; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23IBS704; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23IBM775; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23HBX517; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23GBO122; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23FBW049; Medline SKU # DYNJ40380J, UDI/DI each 10193489201697, UDI/DI case 40193489201698, Lot Number: 23EBG231; Medline SKU # DYNJ43403D, UDI/DI each 10195327179236, UDI/DI case 40195327179237, Lot Number: 23FLA639; Medline SKU # DYNJ48487F, UDI/DI each 10198459572937, UDI/DI case 40198459572938, Lot Number: 26AMA021; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25DMH709; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25CMI786; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMF563; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 25BMC031; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24KMC197; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24IMI472; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24HMA312; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24GMA788; Medline SKU # DYNJ50779G, UDI/DI each 10195327686499, UDI/DI case 40195327686490, Lot Number: 24EMG321; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25LMC853; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25KMB250; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25IME495; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25HME177; Medline SKU # DYNJ50779I, UDI/DI each 10198459423758, UDI/DI case 40198459423759, Lot Number: 25GMF741; Medline SKU # DYNJ57049N, UDI/DI each 10198459519994, UDI/DI case 40198459519995, Lot Number: 25IMH467; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24GBV109; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24GBK784; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24EBR407; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24DBR592; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24CBJ904; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 24ABS845; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23KBR978; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23JBX436; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23IBJ923; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23HBV971; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23GBV284; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23FBK372; Medline SKU # DYNJ57713C, UDI/DI each 10195327357498, UDI/DI case 40195327357499, Lot Number: 23DBO863; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25DBT278; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25CBD829; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25BBU169; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 25ABB803; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24KBS852; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24IBP601; Medline SKU # DYNJ57713D, UDI/DI each 10198459106583, UDI/DI case 40198459106584, Lot Number: 24IBA400; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 26ABU685; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 26ABS702; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25KBM524; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25JBD873; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25IBR638; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25IBB474; Medline SKU # DYNJ57713F, UDI/DI each 10198459380761, UDI/DI case 40198459380762, Lot Number: 25FBQ397; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24JME277; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24FMA227; Medline SKU # DYNJ57748A, UDI/DI each 10888277746015, UDI/DI case 40888277746016, Lot Number: 24EMG444; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 25ALB024; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24KLA767; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24JLA065; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24ILA329; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24HLA502; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24FLA645; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24ELA005; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24DLA522; Medline SKU # DYNJ58771G, UDI/DI each 10195327663308, UDI/DI case 40195327663309, Lot Number: 24DLA423; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 26BLA718; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25LLA775; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA548; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA434; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25ILA036; Medline SKU # DYNJ59229F, UDI/DI each 10198459483936, UDI/DI case 40198459483937, Lot Number: 25HLA821; Medline SKU # DYNJ61294F, UDI/DI each 10195327691158, UDI/DI case 40195327691159, Lot Number: 25HBP380; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25LMG882; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25KMB331; Medline SKU # DYNJ61353C, UDI/DI each 10198459503054, UDI/DI case 40198459503055, Lot Number: 25JMK949; Medline SKU # DYNJ64636A, UDI/DI each 10195327374617, UDI/DI case 40195327374618, Lot Number: 25FBE213; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 26ABC312; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25LBG126; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25IBS553; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25IBN181; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25GBT426; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25FBP518; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 25BBR414; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24KBO515; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24JBT061; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24GBY011; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24EBS347; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24EBS501; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24CBS901; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24BBD609; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 24BBD609A; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23KBJ915; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23JBH747; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23HBF936; Medline SKU # DYNJ65277A, UDI/DI each 10193489409536, UDI/DI case 40193489409537, Lot Number: 23GBA827; Medline SKU # DYNJ66442F, UDI/DI each 10198459334061, UDI/DI case 40198459334062, Lot Number: 25DBO052A; Medline SKU # DYNJ69845C, UDI/DI each 10198459657467, UDI/DI case 40198459657468, Lot Number: 26ABV680; Medline SKU # DYNJ83991C, UDI/DI each 10198459629945, UDI/DI case 40198459629946, Lot Number: 26AMF581; Medline SKU # DYNJ84133B, UDI/DI each 10198459501777, UDI/DI case 40198459501778, Lot Number: 25KBB504; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 26BMH078; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25LMH057; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25JMD619; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25IMH487; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25IMD325; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25FMJ567; Medline SKU # DYNJ85425B, UDI/DI each 10198459305566, UDI/DI case 40198459305567, Lot Number: 25FMH927; Medline SKU # DYNJT3093, UDI/DI each 10198459305801, UDI/DI case 40198459305802, Lot Number: 25EBB099; Medline SKU # DYNJT3522A, UDI/DI each 10198459540479, UDI/DI case 40198459540470, Lot Number: 25JMI340; Medline SKU # DYNJT3522B, UDI/DI each 10198459602825, UDI/DI case 40198459602826, Lot Number: 25LMJ747; Medline SKU # DYNJT3522B, UDI/DI each 10198459602825, UDI/DI case 40198459602826, Lot Number: 25KME591; Medline SKU # DYNJT4069, UDI/DI each 10198459394751, UDI/DI case 40198459394752, Lot Number: 25JMI514; Medline SKU # DYNJT4069, UDI/DI each 10198459394751, UDI/DI case 40198459394752, Lot Number: 25GMB597; Medline SKU # DYNJT5685, UDI/DI each 10198459514098, UDI/DI case 40198459514099, Lot Number: 25JLA132; Medline SKU # DYNJT6544, UDI/DI each 10198459558689, UDI/DI case 40198459558680, Lot Number: 25KMF858; Medline SKU # DYNJT7591, UDI/DI each 10198459635823, UDI/DI case 40198459635824, Lot Number: 26AMH405; Medline SKU # DYNJT7863, UDI/DI each 10198459654831, UDI/DI case 40198459654832, Lot Number: 26ABT695; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 26BBF460; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 26ABT525; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBP400; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBK861; Medline SKU # DYNJTUMSYR, UDI/DI each 10884389882054, UDI/DI case 40884389882055, Lot Number: 25LBK862; Medline SKU # VASC1054, UDI/DI each 10889942379217, UDI/DI case 40889942379218, Lot Number: 25GBM922; Medline SKU # VASC1054, UDI/DI each 10889942379217, UDI/DI case 40889942379218, Lot Number: 25FBV097; Medline SKU # VASC1130A, UDI/DI each 10198459040092, UDI/DI case 40198459040093, Lot Number: 25HBE924; Medline SKU # VASC1137A, UDI/DI each 10195327636692, UDI/DI case 40195327636693, Lot Number: 25HBU244; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 25FBQ566; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 25FBH292; Medline SKU # VASC1299A, UDI/DI each 10195327076450, UDI/DI case 40195327076451, Lot Number: 24HBE146; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 26ABT499; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25LBD803; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25IBN891; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25GBM607; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25FBI750; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25EBQ507; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25DBT209; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 25BBS353; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24KBQ595; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24HBX234; Medline SKU # VASC1399B, UDI/DI each 10198459001802, UDI/DI case 40198459001803, Lot Number: 24EBT407; Medline SKU # VASC1403, UDI/DI each 10195327366285, UDI/DI case 40195327366286, Lot Number: 24BBU522; Medline SKU # VASC1403, UDI/DI each 10195327366285, UDI/DI case 40195327366286, Lot Number: 23KBJ722; Medline SKU # VASC1404, UDI/DI each 10195327366568, UDI/DI case 40195327366569, Lot Number: 25DBE744; Medline SKU # VASC1404, UDI/DI each 10195327366568, UDI/DI case 40195327366569, Lot Number: 25FBH475; Medline SKU # VASC1405, UDI/DI each 10195327367015, UDI/DI case 40195327367016, Lot Number: 25DBJ392; Medline SKU # VASC1455A, UDI/DI each 10198459025396, UDI/DI case 40198459025397, Lot Number: 25IBD178; Medline SKU # VASC1496, UDI/DI each 10195327615666, UDI/DI case 40195327615667, Lot Number: 25FBH317; Medline SKU # VASC1502, UDI/DI each 10195327667719, UDI/DI case 40195327667710, Lot Number: 24GBJ600; Medline SKU # VASC1502, UDI/DI each 10195327667719, UDI/DI case 40195327667710, Lot Number: 24FBL559; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25HBP433; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25HBG954; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25GBH483; Medline SKU # VASC1555, UDI/DI each 10198459058288, UDI/DI case 40198459058289, Lot Number: 25FBT540; Medline SKU # VASC1609, UDI/DI each 10198459144653, UDI/DI case 40198459144654, Lot Number: 25IBQ779; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25JBK828; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25IBG375; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25EBG231; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25EBE966; Medline SKU # VASC1668, UDI/DI each 10198459228155, UDI/DI case 40198459228156, Lot Number: 25BBE722; Medline SKU # VASC1694, UDI/DI each 10198459252501, UDI/DI case 40198459252502, Lot Number: 25HBN801; Medline SKU # VASC1805, UDI/DI each 10198459399022, UDI/DI case 40198459399023, Lot Number: 25HMI038; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 26BMF266; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 25JBQ974; Medline SKU # VASC1816, UDI/DI each 10198459403224, UDI/DI case 40198459403225, Lot Number: 25HBN802.
Devices named in this recall
53 devices in the FDA device registry whose identifier appears in the recall's code information.
- NAMIC 60010508Medline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ38333DMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ38333FMedline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNJ0803367JMedline Industries, Inc. · General surgery tray3 recalls4 reports
- NAMIC 60161954Medline Industries, Inc. · Cardiovascular procedure kit3 recalls1 report
- Medline Industries, Inc. 60160075Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYNJ43403DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ50779GMedline Industries, Inc. · General surgery tray3 recalls0 reports
- NAMIC 60100701Medline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- NAMIC 60230148Medline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ65277AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. BIOB686Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. VASC1404Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. VASC1405Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ24250GMedline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ50779IMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJTUMSYRMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. VASC1299AMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. VASC1455AMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. DYNJT3093Medline Industries, Inc. · General surgery tray1 recall1 report
- NAMIC 60131209Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ40380JMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57713CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1403Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64636AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ0392706KMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1496Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1137AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58771GMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1502Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61294FMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. 00-500510IMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ0451426OMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1399BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1130AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1555Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57713DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ07953NMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1609Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1668Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1694Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ85425BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ38910DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66442FMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ0451426PMedline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJT4069Medline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. VASC1805Medline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. VASC1816Medline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ20898TMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57748AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ19343JMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. VASC1054Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ0376145MMedline Industries, Inc. · General surgery tray1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98598).
- Z-1712-2026 · Open, ClassifiedNAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,T…
- Z-1713-2026 · Open, ClassifiedNAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3.…
- Z-1714-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline S…
- Z-1715-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medl…
- Z-1716-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 0…
- Z-1718-2026 · Open, ClassifiedNAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG…
- Z-1719-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU …
- Z-1720-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: SPECIAL PROCEDURE TRAY, Medline SK…
- Z-1721-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. HEAD AND NECK CDS-LF, Medline S…
- Z-1722-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC,…
- Z-1723-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU …
- Z-1725-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: GLOVE PACK 7.0 (15PK), Medline SKU…
- Z-1726-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ5…
- Z-1727-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. DELIVERY PACK, Medline SKU # DY…
- Z-1728-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYN…
- Z-1729-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
- Z-1730-2026 · Open, ClassifiedMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. SHOULDER ARTHROSCOPY PACK, Medl…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page