Medline medical convenience kits, containing NAMIC… recallZ-1726-2026 · Medline Industries, LP · initiated Feb 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-1726-2026
Status
Open, Classified
Initiated by the company
Feb 27, 2026
Posted by the FDA
Apr 7, 2026
Recalling firm
Medline Industries, LP
Quantity
2154 units
Product code
OJG
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Product description

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ57032D; 2. NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C; 3. NEURO INERVENTIONAL RADIOLOGY, Medline SKU # VASC1650.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Medline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBN999; Medline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBO039; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24JBC741; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBM270; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBE756; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24GBC735; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24FBF002; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24DBC238; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24ABY360; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 23LBH735; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABP305; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABI725; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25KBD893; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBM531; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBA055; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25EBC929; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBL096; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBD577; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBI493; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25ABI074; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 24LBG466; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26BBG430; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25LBC439.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98598).