Revolution recallZ-2155-2026 · GE Medical Systems, LLC · initiated Mar 26, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
FDA root cause category: Software design
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to [email protected] Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare
Recall details
- Recall number
- Z-2155-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 26, 2026
- Posted by the FDA
- May 8, 2026
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 5 units
- Product code
- JAK
- FDA submissions
- K213715
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27
Code information (models, lots, serial numbers)
UDI-DI: 00195278469700; Serial Numbers: REV242400001CN REV242400004CN REV242300006CN REV242400002CN REV242300004CN
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Revolution Apex SelectGE Hangwei Medical Systems Company Ltd. · System, x-ray, tomography, computed2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98738).
- Z-2149-2026 · Open, ClassifiedRevolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300
- Z-2150-2026 · Open, ClassifiedRevolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
- Z-2151-2026 · Open, ClassifiedRevolution Apex Elite, X-ray Computed Tomography, Model Number 5590000-23
- Z-2152-2026 · Open, ClassifiedRevolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
- Z-2153-2026 · Open, ClassifiedRevolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
- Z-2154-2026 · Open, ClassifiedRevolution Apex, X-ray Computed Tomography, Model Number 5590000-20
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.