Revolution Apex SelectRecalls & reported problems
GE Hangwei Medical Systems Company Ltd. · Revolution Apex Select - Designated System Medical Device · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
9 other recalls cover devices cleared under the same FDA submission (K213715); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 26, 2026Open, Classified
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Z-2155-2026 · Root cause: Software design · FDA record ↗
- Feb 18, 2025Open, Classified
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Z-1297-2025 · Root cause: Process change control · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K213715. One submission often covers many models, so check the product and lot details in each record.
- Z-2154-2026 · Mar 26, 2026 · Open, ClassifiedRevolution Apex, X-ray Computed Tomography, Model Number 5590000-20
- Z-2153-2026 · Mar 26, 2026 · Open, ClassifiedRevolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
- Z-2152-2026 · Mar 26, 2026 · Open, ClassifiedRevolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
- Z-2150-2026 · Mar 26, 2026 · Open, ClassifiedRevolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
- Z-2151-2026 · Mar 26, 2026 · Open, ClassifiedRevolution Apex Elite, X-ray Computed Tomography, Model Number 5590000-23
- Z-1298-2025 · Feb 18, 2025 · Open, ClassifiedGE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
- Z-1303-2025 · Feb 18, 2025 · Open, ClassifiedGE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
- Z-1300-2025 · Feb 18, 2025 · Open, ClassifiedGE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
- Z-1301-2025 · Feb 18, 2025 · Open, ClassifiedGE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.