Polysorb recallZ-2281-2026 · Covidien, LP · initiated Apr 15, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
FDA root cause category: Process control
What the company is doing
On April 23, 2026 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: - Immediately identify and quarantine all unused impacted product listed above. - See Appendix A for guidance on identifying affected lots. - This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Return all unused impacted product listed above to Medtronic. Medtronic Actions: - Medtronic representatives will assist customers with the return of affected product, upon request. - Medtronic representatives will work with customers to ensure replacement products are available to avoid interruptions in patient care.
Recall details
- Recall number
- Z-2281-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2026
- Posted by the FDA
- Jun 1, 2026
- Recalling firm
- Covidien, LP
- Quantity
- 2448 units
- Product code
- OCW
- FDA submissions
- K934738
- Distribution
- Domestic: Nationwide distribution; International: Austria, Canada, Denmark, France, Italy;
Product description
Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
Code information (models, lots, serial numbers)
1. Product Number: 170053; UDI-DI: 10884521126732; UPN: 20884521126739; Lot number: J5L2332Y; 2. Product Number: 170071; UDI-DI: 10884521126787; UPN: 20884521126784; Lot number: J5H2924Y;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Polysorb 170053Covidien LP · Endoscopic tissue approximation device1 recall63 reports
- Polysorb 170071Covidien LP · Endoscopic tissue approximation device1 recall2 reports
Product lines
Other devices under the same FDA submission
24 devices cleared under K934738 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Endo Stitch 173016Covidien LP · Endoscopic tissue approximation device1 death1,333 reports
- Surgidac 173024Covidien LP · Endoscopic tissue approximation device70 reports
- Polysorb 170052Covidien LP · Endoscopic tissue approximation device57 reports
- Surgidac 173023Covidien LP · Endoscopic tissue approximation device55 reports
- Polysorb 170056Covidien LP · Endoscopic tissue approximation device34 reports
- Surgidac 173021Covidien LP · Endoscopic tissue approximation device27 reports
- Sofsilk 170004Covidien LP · Endoscopic tissue approximation device13 reports
- Surgidac 173026Covidien LP · Endoscopic tissue approximation device12 reports
- Surgidac 170043Covidien LP · Endoscopic tissue approximation device11 reports
- Polysorb 170051Covidien LP · Endoscopic tissue approximation device10 reports
- Surgidac 170040Covidien LP · Endoscopic tissue approximation device10 reports
- Sofsilk 170002Covidien LP · Endoscopic tissue approximation device9 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page