Skip to content

Polysorb 170051Recalls & reported problems

Covidien LP · Absorbable Single Stitch Reload · Class 2

All models of Polysorb (291)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 20884521126753), received between Sep 28, 2017 and Mar 13, 2024 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

2 other recalls cover devices cleared under the same FDA submission (K934738); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Human-Device Interface Problem5
  • Break3
  • Device Dislodged or Dislocated3
  • Entrapment of Device2
  • Detachment Of Device Component1
  • Difficult to Remove1
  • Use of Device Problem1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Known Impact Or Consequence To Patient4
  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Unintended Radiation Exposure1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC TRANSABDOMINAL PRE-PERITONEAL (TAPP) HERNIOPLASTY, THE NEEDLE BROKE WITHIN THE SURGICAL FIELD DURING THE INTERVENTION. THE NEEDLE WAS REMOVED WITH A HOLDER TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18893763 ↗
  • InjuryOct 7, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, IN CLOSURE OF VAGINAL CUFF, A PORTION OF THE NEEDLE FELL INTO PATIENT'S CAVITY. THE PIECE WAS NOT RETRIEVED AS THE SURGEON WAS UNABLE TO LOCATE. TO RESOLVE THE ISSUE AND COMPLETE THE CASE, A NEW RELOAD WAS USED.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; Unintended Radiation Exposure

    MAUDE report 15565940 ↗
  • MalfunctionFeb 3, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC ROUX-EN-Y BYPASS, WHEN COMPLETING THE GASTROJEJUNAL ANASTOMOSIS, TRAUMA TO THE TISSUE WAS NOTED DUE TO THE NEEDLE NOT PASSING THROUGH THE JAWS SMOOTHLY. UPON CLOSER INSPECTION, WHEN THE NEEDLE WAS IN THE LEFT JAW OF THE DEVICE, THE NEEDLE APPEARED TO BE LOOSE. NEW HANDLE AND RELOAD WERE USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11272979 ↗
  • InjuryOct 2, 2020Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRECTOMY, A NEEDLE FRAGMENT PARTIALLY BROKE OFF WHILE THE SURGEON WAS SUTURING THE STOMACH WALL. THE REMAINING TIP OF APPROX. 2MM WAS LEFT ON THE DEVICE. THE REMAINING FRAGMENT WAS UNABLE TO BE FOUND. THE SURGEON PROCEEDED WITH THE CASE THEREAFTER.

    Patient problems: Foreign Body In Patient

    MAUDE report 10618624 ↗
  • MalfunctionDec 20, 2019Entrapment of Device; Human-Device Interface Problem; Entrapment of Device; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BYPASS, THE SURGEON EXPERIENCED RESISTANCE WHEN GOING THROUGH THE TISSUE AS THE SURGEON TOGGLED THE NEEDLE BETWEEN THE JAWS. THE NEEDLE WOULD GET STUCK AS IF THERE WAS A BARB ON THE NEEDLE SHAFT THAT WAS CATCHING ON THE TISSUE. DOCTOR MENTIONED THAT 3 OF 4 INSTRUMENTS FELT GRITTY UPON NEEDLE EXCHANGE BETWEEN THE JAWS WHEN TOGGLING. TISSUE THICKNESS APPEARED NORMAL. THE FOURTH AND FINAL LOAD FELT SMOOTH AND WENT THROUGH WITH OUT RESISTANCE AND THE SURGEON CONTINUED TO USE THE HANDLE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9504977 ↗
  • MalfunctionDec 20, 2019Entrapment of Device; Human-Device Interface Problem; Entrapment of Device; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BYPASS, THE SURGEON EXPERIENCED RESISTANCE WHEN GOING THROUGH THE TISSUE AS THE SURGEON TOGGLED THE NEEDLE BETWEEN THE JAWS. THE NEEDLE WOULD GET STUCK AS IF THERE WAS A BARB ON THE NEEDLE SHAFT THAT WAS CAUGHT ON THE TISSUE. DOCTOR MENTIONED THAT 3 OF 4 INSTRUMENTS FELT GRITTY UPON NEEDLE EXCHANGE BETWEEN THE JAWS WHEN TOGGLING. TISSUE THICKNESS APPEARED NORMAL. THE FOURTH AND FINAL LOAD FELT SMOOTH AND WENT THROUGH WITH OUT RESISTANCE AND THE SURGEON CONTINUED TO USE THE HANDLE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9503705 ↗
  • MalfunctionJul 25, 2018Detachment of Device or Device Component; Device Dislodged or Dislocated; Human-Device Interface Problem; Detachment of Device or Device Component; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, WHEN THE DEVICE WAS BEING APPLIED TO THE JEJUNUM DURING THE SEWING OF THE GASTROJEJUNOSTOMY ON A LAPAROSCOPIC ROUX EN Y GASTRIC BYPASS, THE DEVICE WAS DIFFICULT TO UNLOAD AT ONE POINT. THE NEEDLE DISENGAGED COMPLETELY FROM THE DEVICE BUT WAS STILL IN THE SMALL BOWEL WITH THE STRAND OF SUTURE ATTACHED, SO IT WAS VISIBLE THE ENTIRE TIME. A NEW RELOAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7719179 ↗
  • MalfunctionJul 23, 2018Detachment of Device or Device Component; Device Dislodged or Dislocated; Human-Device Interface Problem; Detachment of Device or Device Component; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, WHEN THE DEVICE WAS BEING APPLIED TO THE JEJUNUM DURING THE SEWING OF THE GASTROJEJUNOSTOMY ON A LAPAROSCOPIC ROUX EN Y GASTRIC BYPASS, THE DEVICE WAS DIFFICULT TO UNLOAD AT ONE POINT. THE NEEDLE DISENGAGED COMPLETELY FROM THE DEVICE BUT WAS STILL IN THE SMALL BOWEL WITH THE STRAND OF SUTURE ATTACHED, SO IT WAS VISIBLE THE ENTIRE TIME. A NEW RELOAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7712521 ↗
  • MalfunctionJul 10, 2018Difficult to Remove; Detachment of Device or Device Component; Device Dislodged or Dislocated; Human-Device Interface Problem; Detachment of Device or Device Component; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, WHEN THE DEVICE WAS BEING APPLIED TO THE JEJUNUM DURING THE SEWING OF THE GASTROJEJUNOSTOMY ON A LAPAROSCOPIC ROUX EN Y GASTRIC BYPASS, THE DEVICE WAS DIFFICULT TO UNLOAD AT ONE POINT. THE NEEDLE DISENGAGED COMPLETELY FROM THE DEVICE BUT WAS STILL IN THE SMALL BOWEL WITH THE STRAND OF SUTURE ATTACHED, SO IT WAS VISIBLE THE ENTIRE TIME. A NEW RELOAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7674237 ↗
  • MalfunctionSep 28, 2017Detachment Of Device Component; Detachment Of Device Component

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BYPASS, THE NEEDLE FELL OUT OF THE JAWS AND WAS RETRIEVED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6901613 ↗

FDA clearance and classification

Cleared through 510(k) K934738 on Mar 16, 1994 (decision: Substantially Equivalent). Submitted by United States Surgical, A Division of Tyco Healthc as “Autosuture Endoscopic Suturing Device”. The FDA review took 166 days. FDA record ↗

The FDA regulates product code OCW (endoscopic tissue approximation device) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K934738. One submission often covers many models, so check the product and lot details in each record.

  • Z-2281-2026 · Apr 15, 2026 · Open, Classified
    Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
  • Z-2543-2017 · May 3, 2017 · Terminated
    Sutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH POLYSORB 2/0 48" U/D DLU

Common questions about the Polysorb 170051

Has the Polysorb 170051 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20884521126753), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Polysorb 170051?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Sep 28, 2017 and Mar 13, 2024. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (5), “Human-Device Interface Problem” (5) and “Break” (3); the most common patient problems are “Foreign Body In Patient” (2), “Device Embedded In Tissue or Plaque” (1) and “Unintended Radiation Exposure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Polysorb 170051?

None of the reports are classified as deaths, 2 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Polysorb 170051 FDA approved?

It was cleared through 510(k) K934738 on Mar 16, 1994. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Polysorb 170051, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand Polysorb, model 170051). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Endoscopic tissue approximation device devices

All Endoscopic tissue approximation device devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page