Why it was recalled

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

FDA root cause category: Process control

What the company is doing

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

Recall details

Recall number
Z-2418-2026
Status
Open, Classified
Initiated by the company
Apr 27, 2026
Posted by the FDA
Jun 12, 2026
Recalling firm
Medline Industries, LP
Quantity
49,654 kits total
Product code
OEZ
Distribution
US Nationwide distribution. OUS distribution pending.

Product description

Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M

Code information (models, lots, serial numbers)

DYNJ61460C UDI-DI 10198459422133 lot 25KBA098 CDS840150Y UDI-DI 10198459303029 lots 25JBW220 26BBD028 DYNJ56297C UDI-DI 10198459513954 lot 25LMJ874 DYNJT3454 UDI-DI 10198459325663 lot 25KBG761 DYNJ33822L UDI-DI 10198459385582 lot 26AMA072 DYNJ900962O UDI-DI 10198459452581 lot 26CBC302 DYNJ54238L UDI-DI 10198459443114 lot 25LMH549 DYNJT6766 UDI-DI 10198459580888 lot 25KBM870 DYNJ909794C UDI-DI 10198459528002 lot 25LBF702 CVI5323A UDI-DI 10198459582035 lot 25KBM758 CDS983040M UDI-DI 10198459579813 lots 25KBO682 26BBB926

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98951).

Sources and updates