Medline convenience kits recallZ-2443-2026 · Medline Industries, LP · initiated Apr 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
FDA root cause category: Process control
What the company is doing
Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Recall details
- Recall number
- Z-2443-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 27, 2026
- Posted by the FDA
- Jun 12, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 49,654 kits total
- Product code
- OES
- Distribution
- US Nationwide distribution. OUS distribution pending.
Product description
Medline convenience kits: PORT A CATH DYNJ906818A
Code information (models, lots, serial numbers)
DYNJ906818A 10195327027261 25LBC291
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ906818AMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98951).
- Z-2414-2026 · Open, ClassifiedMedline EBSI Kit EBSI1534
- Z-2416-2026 · Open, ClassifiedBLOCK TRAY DYNJRA2119A INDWELLING NERVE BLOCK DYNJRA1011A
- Z-2417-2026 · Open, ClassifiedMINOR CATH LAB PACK DYNJ84775A OR VASCULAR ANGIO PACK DYNJ47269D PORT-A-CATH DYNJ63276G
- Z-2418-2026 · Open, ClassifiedMedline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PED…
- Z-2419-2026 · Open, ClassifiedMedline convenience kits: INPATIENT CENTRAL LN DRSG KIT DYNDC3525 PICC/CVC/MIDLINE DRSG CHG KIT DT13191A PORT A CATH CDS982047S VEIN ABLATION PACK DYNJ62889C
- Z-2420-2026 · Open, ClassifiedMedline convenience kits: BREAST-HERNIA-PORT CDS-LF CDS984853L
- Z-2421-2026 · Open, ClassifiedMedline convenience kits: DISCECTOMY PACK DYNJ86476B EAR DYNJ81460C DYNJ900004J DYNJ901779J DYNJ902900F EAR PACK DYNJ81419D ENT PACK DYNJ40498B HEAD & NECK CDS DYNJ901246J JE…
- Z-2422-2026 · Open, ClassifiedMedline convenience kits: ABD PACK DYNJ81673B BARIATRIC DYNJ902515K BARIATRIC + DYNJ904119O BARIATRIC CDS CDS983883J BWNBORO LAP CHOLE PACK-LF DYNJ32836B CCSB KARIN JWIRE LUM…
- Z-2423-2026 · Open, ClassifiedMedline convenience kits: BREAST IMAGING DYNJT5893 CHOG PLASTICS SYNJ10306A ENDO BREAST AUGMENTATION PACK DYNJ68023 FACIAL PLASTIC DYNJ909926C JVL - MODULE PLASTIC PK DYNJ8408…
- Z-2424-2026 · Open, ClassifiedMedline convenience kits: BASIC DYNJ905869G BASIC PACK DYNJ68484B BIB GENERAL LAPAROSCOPY PA DYNJ58305 CHRISTUS CH PEDIATRIC BASIC DYNJ61167G CVC INSERTION BUNDLE CVI3940A DA…
- Z-2425-2026 · Open, ClassifiedMedline convenience kit: IV START KIT W/HUBGUARD DYNDV2662
- Z-2426-2026 · Open, ClassifiedMedline convenience kit: KIT: DRESSING CHANGE PED DYNDC3359
- Z-2428-2026 · Open, ClassifiedMedline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7315 NEURO PACK-LF DY…
- Z-2430-2026 · Open, ClassifiedMedline convenience kit: SMSC-PELVISCOPY PACK-LF DYNJ0161619M
- Z-2431-2026 · Open, ClassifiedMedline convenience kits: DAVINCI DYNJ905836D DAVINCI-LF DYNJ905293I GENERAL ROBOTICS PACK DYNJT8308 GYN LAP CDS860087Q GYN LAPAROSCOPY DYNJ910084C GYN LAPAROSCOPY CDS CDS982…
- Z-2432-2026 · Open, ClassifiedMedline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C
- Z-2435-2026 · Open, ClassifiedMedline convenience kits: BASIC EYE DYNJ905154I TUBAL LIGATION CDS CDS760053K
- Z-2436-2026 · Open, ClassifiedMedline convenience kits: ACD DYNJ904999J ACDF DYNJ907336B DYNJ910821A ACDF PACK DYNJ81098C ACL DYNJ911996 ACL ACCESSORY PACK DYNJ86584B ACL- LF DYNJ905281B ACL PACK DYNJ672…
- Z-2437-2026 · Open, ClassifiedCenturion convenience kits: CENTRAL LINE DRESSING SV43 FCP DYNDC2356B DIALYSIS CATHETER MAINTENANCE KIT DT17570 DIALYSIS DRESSING CHANGE KIT DT21370B DRESSING CHANGE TRAY - GOL…
- Z-2438-2026 · Open, ClassifiedMedline convenience kits: CMH CENTRAL LINE DRESSING CHANGE KIT EBSI1705A CVC/PICC/MIDLINE DRESSING CHANGE KIT EBSI1057A DIALYSIS DRESSING CHANGE EBSI1498 EBSI CVC DRESSING CHAN…
- Z-2439-2026 · Open, ClassifiedMedline convenience kits: 20CM CVC INSERTION KIT ECVC6805B 2L 8.5FR 16CM CVC INSERTION KIT ECVC8055A VANTEX PI CVC 7F, 3L, 20CM BUNDLE ECVC6915A ECVC6985B
- Z-2440-2026 · Open, ClassifiedMedline convenience kits: DRIVELINE MANAGEMENT KIT EBSI1688 PORT DRESSING CHANGE KIT EBSI1328A
- Z-2441-2026 · Open, ClassifiedMedline convenience kits: ACL PACK DYNJ64161C ARTHROSCOPY PACK DYNJ47103D DYNJ67197G DYNJ85197A DYNJT4007 BASIC ACCESS PACK-LF DYNJ36698B BASIC BACK TRAY DYNJ89037 BASIC PEDI …
- Z-2442-2026 · Open, ClassifiedMedline convenience kits: FEMORAL BLOCK TRAY DYNJRA1739B
- Z-2444-2026 · Open, ClassifiedMedline convenience kits: BIB AV FISTULA PACK DYNJ58304B CARDIAC MINOR SURGICOUNT PACK DYNJ56573B ENDOVASCULAR PACK DYNJ29135N EP DEVICE DYNJ65916B FULL LINE ANEURYSM CDS CDS9…
- Z-2445-2026 · Open, ClassifiedMedline convenience kits: ADD-A-CATH ARTERIAL INSERTION ARTNC185 PED SCHOOL AGE PICC DRSG TRAY DYNDC3291
- Z-2446-2026 · Open, ClassifiedMedline convenience kits: C SECTION CDS985505G C SECTION PACK DYNJ68260D C-SECTION DYNJ905094B C-SECTION RECOVERY #2 CDS820091P
- Z-2447-2026 · Open, ClassifiedMedline convenience kits: HYSTERECTOMY PACK-LF DYNJ20485L
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page