Skip to content

Medline convenience kit recallZ-2428-2026 · Medline Industries, LP · initiated Apr 27, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

FDA root cause category: Process control

What the company is doing

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

Recall details

Recall number
Z-2428-2026
Status
Open, Classified
Initiated by the company
Apr 27, 2026
Posted by the FDA
Jun 12, 2026
Recalling firm
Medline Industries, LP
Quantity
49,654 kits total
Product code
OJG
Distribution
US Nationwide distribution. OUS distribution pending.

Product description

Medline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7315 NEURO PACK-LF DYNJ25097I NEURO VP SHUNT/VAGAL NERVE-RF DYNJ21925S PK NEURO SHUNT BUMC DYNJT6428 RR-NEURO DBS HEAD PACK-LF DYNJ0785736M VP SHUNT SYNJ10307B

Code information (models, lots, serial numbers)

DYNJ904900N UDI-DI 10198459504273 Lot 25LBB457 DYNJ67267B UDI-DI 10198459084331 Lot 25LBQ732 DYNJ9814630O UDI-DI 10195327587291 Lot 25LBD665 CDS983383L UDI-DI 10198459224577 Lot 25LBC092 DYNJ59397Q UDI-DI 10198459495427 Lot 25LMG049 DYNJT7315 UDI-DI 10198459620225 Lot 25LBO095 DYNJ25097I UDI-DI 10198459449550 Lot 25KBC747 DYNJ21925S UDI-DI 10195327105020 Lots 25KBD835 25LBR371 DYNJT6428 UDI-DI 10198459553660 Lot 26AMA602 DYNJ0785736M UDI-DI 10198459487149 Lot 25LMI831 SYNJ10307B UDI-DI 10195327007249 Lot 25LBE309

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98951).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page