MEDLINE Medical Procedure Kits labeled as recallZ-2503-2026 · Medline Industries, LP · initiated May 4, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2503-2026
- Status
- Open, Classified
- Initiated by the company
- May 4, 2026
- Posted by the FDA
- Jun 17, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 44656 units
- Product code
- OHD
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Product description
MEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR PACK, etc. (see the recall documents for a full list of affected products)
Code information (models, lots, serial numbers)
Show the full code information (77k characters)
Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24GMJ742; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24FMC978; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24CMA556; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24BMI027; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24AMF247; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 24AMA807; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23IMF051; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23HMG141; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23EMF259; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23DMA689; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23CMF097; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 23CMC699; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22LME333; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22KMB390; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22HMC449; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22GMD678; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22FMG214; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22DME100; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22CME928; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22BMG698; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 22AMD289; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21KMD167; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21JMF926; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21IMD987; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21GME755; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21FBF900; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21CBF873; Medline Kit Number/SKU DYNJ905924B: UDI/DI 10193489773484 (each), 40193489773485 (case), Lot Number 21ABE015; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 26DME629; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25LMG110; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25LMB085; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25IMH993; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25EMC711; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25DMH514; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25CMD890; Medline Kit Number/SKU DYNJ905924D: UDI/DI 10198459256400 (each), 40198459256401 (case), Lot Number 25CMB602; Medline Kit Number/SKU DYNJ905924C: UDI/DI 10198459076527 (each), 40198459076528 (case), Lot Number 24KMH725; Medline Kit Number/SKU DYNJ905924C: UDI/DI 10198459076527 (each), 40198459076528 (case), Lot Number 24IMG083; Medline Kit Number/SKU DYNJ906047A: UDI/DI 10195327009014 (each), 40195327009015 (case), Lot Number 23ABJ617; Medline Kit Number/SKU DYNJ906047C: UDI/DI 10198459287473 (each), 40198459287474 (case), Lot Number 26DBN259; Medline Kit Number/SKU DYNJ906047C: UDI/DI 10198459287473 (each), 40198459287474 (case), Lot Number 26CBD812; Medline Kit Number/SKU DYNJ906047C: UDI/DI 10198459287473 (each), 40198459287474 (case), Lot Number 25GBK346; Medline Kit Number/SKU DYNJ906047C: UDI/DI 10198459287473 (each), 40198459287474 (case), Lot Number 25DBT775; Medline Kit Number/SKU DYNJ906047C: UDI/DI 10198459287473 (each), 40198459287474 (case), Lot Number 25DBA222; Medline Kit Number/SKU DYNJ901133A: UDI/DI 10888277949683 (each), 40888277949684 (case), Lot Number 22JBR884; Medline Kit Number/SKU DYNJ901133A: UDI/DI 10888277949683 (each), 40888277949684 (case), Lot Number 21KBP568; Medline Kit Number/SKU DYNJ901133A: UDI/DI 10888277949683 (each), 40888277949684 (case), Lot Number 21KBF079; Medline Kit Number/SKU DYNJ901133A: UDI/DI 10888277949683 (each), 40888277949684 (case), Lot Number 21JBV891; Medline Kit Number/SKU DYNJ901133A: UDI/DI 10888277949683 (each), 40888277949684 (case), Lot Number 21JBM245; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24LBS027; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24LBM384; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24KBU684; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24JBR707; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24JBI927; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24HBN678; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24GBM263; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24EBT854; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24DBJ743; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24DBA374; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24BBS868; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 24ABV100; Medline Kit Number/SKU DYNJ909989: UDI/DI 10195327531867 (each), 40195327531868 (case), Lot Number 23KBI149; Medline Kit Number/SKU DYNJ53820A: UDI/DI 10195327132187 (each), 40195327132188 (case), Lot Number 23JBM250; Medline Kit Number/SKU DYNJ53820A: UDI/DI 10195327132187 (each), 40195327132188 (case), Lot Number 23EBT822; Medline Kit Number/SKU DYNJ53820A: UDI/DI 10195327132187 (each), 40195327132188 (case), Lot Number 22IBL549; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23KBA066; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23JBF229; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23IBE972; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23HBR065; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23DBC310; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 23BBH824; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22LBC384; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22JBY519; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22HBG756; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22FBY290; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22FBX452; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22DBU965; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22DBD528; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 22BBY415; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21JBV162; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21GBW102; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21GBH884; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21EBO133; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21EBA344; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21CBV721; Medline Kit Number/SKU DYNJ68426: UDI/DI 10193489464672 (each), 40193489464673 (case), Lot Number 21BBH253; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 24ABN938; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23LBN389; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23IBN695; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23GBU793; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23FBK757; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23FBK840; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23EBD626; Medline Kit Number/SKU DYNJ906931C: UDI/DI 10193489921298 (each), 40193489921299 (case), Lot Number 23DBA618; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23LBN387; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23KBW414; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23JBB062; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23IBH652; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23GBH151; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23EBN137; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23DBM354; Medline Kit Number/SKU DYNJ906926B: UDI/DI 10193489921366 (each), 40193489921367 (case), Lot Number 23BBU275; Medline Kit Number/SKU DYNJ80875F: UDI/DI 10195327696917 (each), 40195327696918 (case), Lot Number 24DBN149; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 25DBE662; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 25BBS190; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 25ABQ281; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 25ABI056; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 24LBO793; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 24KBT823; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 24JBO829; Medline Kit Number/SKU DYNJ80875G: UDI/DI 10198459052613 (each), 40198459052614 (case), Lot Number 24GBE201; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22JDB540; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22IDA093; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22HDA991; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22FDA312; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22CDC365; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 22ADC069; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21LDA011; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21JDB653; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21IDB153; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21GDB523; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21FDB411; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21DDB687; Medline Kit Number/SKU CDS984860F: UDI/DI 10193489261974 (each), 40193489261975 (case), Lot Number 21ADB205; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 25EMJ345; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 25DMH136; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24KDA185; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24JDA302; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24IDA871; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24IDA284; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24GDA064; Medline Kit Number/SKU CDS984860I: UDI/DI 10195327689131 (each), 40195327689132 (case), Lot Number 24EDB542; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 25BMC364; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 24KMI740; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 24IMB560; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 24FMD487; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 24CMH415; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 24BBO497; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23JBW362; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23IBP888; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23GBQ447; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23EBE671; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23CBU838; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 23BBL424; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 22LBQ484; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 22KBG442; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 22IBN003; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 22HBT961; Medline Kit Number/SKU DYNJ44329I: UDI/DI 10195327126100 (each), 40195327126101 (case), Lot Number 22EBM455; Medline Kit Number/SKU DYNJ44329J: UDI/DI 10198459288869 (each), 40198459288860 (case), Lot Number 26AME227; Medline Kit Number/SKU DYNJ44329J: UDI/DI 10198459288869 (each), 40198459288860 (case), Lot Number 25JMB468; Medline Kit Number/SKU DYNJ44329J: UDI/DI 10198459288869 (each), 40198459288860 (case), Lot Number 25EMH042; Medline Kit Number/SKU DYNJ44329J: UDI/DI 10198459288869 (each), 40198459288860 (case), Lot Number 25CME017; Medline Kit Number/SKU DYNJ909637: UDI/DI 10195327418915 (each), 40195327418916 (case), Lot Number 23FBI194; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 26CMH317; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 25LMI435; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 25JMB123; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 24CBS174; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 23KBX746; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 23KBB409; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 23GBT959; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 23FBS151; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 23CBG791; Medline Kit Number/SKU DYNJ69307A: UDI/DI 10195327221171 (each), 40195327221172 (case), Lot Number 22JBU653; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 23FBS350; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 23DBE139; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 23BBM490; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 23ABB459; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22JBK461; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22HBP270; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22FBS212; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22FBE919; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22OBE689; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 22CBP286; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21KBZ712; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21HBT618; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21HBT619; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21HBT621; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21EBK927; Medline Kit Number/SKU CDS980810Q: UDI/DI 10193489302660 (each), 40193489302661 (case), Lot Number 21DBH511; Medline Kit Number/SKU DYNJ54963A: UDI/DI 10193489744781 (each), 40193489744782 (case), Lot Number 24ABM286; Medline Kit Number/SKU DYNJ54963A: UDI/DI 10193489744781 (each), 40193489744782 (case), Lot Number 23FBQ139; Medline Kit Number/SKU DYNJ37256D: UDI/DI 10889942450138 (each), 40889942450139 (case), Lot Number 21BBG902; Medline Kit Number/SKU DYNJ37256D: UDI/DI 10889942450138 (each), 40889942450139 (case), Lot Number 21ABV665; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 23KBR848; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 23JBP395; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 23FBK998; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 23EBS722; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 23ABO821; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 22HBH568; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 22FBO458; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 22EBO914; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 22DBL479; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 22BBW080; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21KBJ252; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21IBH546; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21FBR226; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21FBB348; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21DBP955; Medline Kit Number/SKU DYNJ37256F: UDI/DI 10193489849691 (each), 40193489849692 (case), Lot Number 21CBL966; Medline Kit Number/SKU DYNJ37256G: UDI/DI 10195327595296 (each), 40195327595297 (case), Lot Number 24CBB688; Medline Kit Number/SKU DYNJ37256G: UDI/DI 10195327595296 (each), 40195327595297 (case), Lot Number 24ABY586; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 23DLA270; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 23ALA937; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22KLB006; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22JLA119; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22HLA193; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22FLA949; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22ELB150; Medline Kit Number/SKU DYNJ61437: UDI/DI 10193489565652 (each), 40193489565653 (case), Lot Number 22ALA976; Medline Kit Number/SKU DYNJ61437A: UDI/DI 10195327385293 (each), 40195327385294 (case), Lot Number 23ILA957; Medline Kit Number/SKU DYNJ61437A: UDI/DI 10195327385293 (each), 40195327385294 (case), Lot Number 23FLA285; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24JLB138; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24JLA079; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24HLB016; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24ELA847; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24DLB021; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24DLA505; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24CLA850; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 24BLA953; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 23LLA455; Medline Kit Number/SKU DYNJ61437B: UDI/DI 10195327548322 (each), 40195327548323 (case), Lot Number 23LLA335; Medline Kit Number/SKU DYNJ61437C: UDI/DI 10198459205798 (each), 40198459205799 (case), Lot Number 26ALA616; Medline Kit Number/SKU DYNJ61437C: UDI/DI 10198459205798 (each), 40198459205799 (case), Lot Number 25HLA983; Medline Kit Number/SKU DYNJ61437C: UDI/DI 10198459205798 (each), 40198459205799 (case), Lot Number 25FLA087; Medline Kit Number/SKU DYNJ61437C: UDI/DI 10198459205798 (each), 40198459205799 (case), Lot Number 25CLA784; Medline Kit Number/SKU DYNJ61437C: UDI/DI 10198459205798 (each), 40198459205799 (case), Lot Number 25BLA892; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 22CBD854; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 22ABF902; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21IBE965; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21HBI211; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21GBA971; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21FBG272; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21DBG163; Medline Kit Number/SKU DYNJ900357O: UDI/DI 10193489818734 (each), 40193489818735 (case), Lot Number 21CBD183; Medline Kit Number/SKU DYNJ900357P: UDI/DI 10195327153076 (each), 40195327153077 (case), Lot Number 22EBP698; Medline Kit Number/SKU DYNJ900357Q: UDI/DI 10195327167356 (each), 40195327167357 (case), Lot Number 22LMB942; Medline Kit Number/SKU DYNJ900357Q: UDI/DI 10195327167356 (each), 40195327167357 (case), Lot Number 22KMB647; Medline Kit Number/SKU DYNJ900357Q: UDI/DI 10195327167356 (each), 40195327167357 (case), Lot Number 22JMF453; Medline Kit Number/SKU DYNJ900357Q: UDI/DI 10195327167356 (each), 40195327167357 (case), Lot Number 22IMH536; Medline Kit Number/SKU DYNJ900357Q: UDI/DI 10195327167356 (each), 40195327167357 (case), Lot Number 22IMA885; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 24LBM670; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 24KBQ810; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23LBL536; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23LBA228; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23FBH088; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23EBT957; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23CBW475; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23CBC589; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 23BBU506; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 22HBQ486; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 22FBP791; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 22ABA292; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 21GBE910; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 21EBI351; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 21CBU642; Medline Kit Number/SKU DYNJ905449B: UDI/DI 10193489825268 (each), 40193489825269 (case), Lot Number 21CBN337; Medline Kit Number/SKU DYNJ905449C: UDI/DI 10198459592485 (each), 40198459592486 (case), Lot Number 26BBG330; Medline Kit Number/SKU DYNJ905927B: UDI/DI 10193489773491 (each), 40193489773492 (case), Lot Number 22JBC353Y; Medline Kit Number/SKU DYNJ905927B: UDI/DI 10193489773491 (each), 40193489773492 (case), Lot Number 22JBC353; Medline Kit Number/SKU DYNJ908881F: UDI/DI 10198459705892 (each), 40198459705893 (case), Lot Number 26DBD955; Medline Kit Number/SKU DYNJ909130I: UDI/DI 10198459639807 (each), 40198459639808 (case), Lot Number 26CBI919; Medline Kit Number/SKU DYNJ909130I: UDI/DI 10198459639807 (each), 40198459639808 (case), Lot Number 26BBI071; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22LMD661; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22KME420; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22IMI546; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22GME206; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22DME454; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 22AMD285; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 21HME823; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 21GMC540; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 21FME894; Medline Kit Number/SKU DYNJ907407: UDI/DI 10193489406443 (each), 40193489406444 (case), Lot Number 21EBC851; Medline Kit Number/SKU DYNJ907407A: UDI/DI 10195327335809 (each), 40195327335800 (case), Lot Number 23FMI434; Medline Kit Number/SKU DYNJ907407A: UDI/DI 10195327335809 (each), 40195327335800 (case), Lot Number 23EME041; Medline Kit Number/SKU DYNJ907407A: UDI/DI 10195327335809 (each), 40195327335800 (case), Lot Number 23DMH235; Medline Kit Number/SKU DYNJ907407B: UDI/DI 10195327462208 (each), 40195327462209 (case), Lot Number 23HMF546; Medline Kit Number/SKU CDS984625B: UDI/DI 10193489262247 (each), 40193489262248 (case), Lot Number 21LDB337; Medline Kit Number/SKU CDS984625B: UDI/DI 10193489262247 (each), 40193489262248 (case), Lot Number 21LDB189; Medline Kit Number/SKU CDS984625B: UDI/DI 10193489262247 (each), 40193489262248 (case), Lot Number 21IDA726; Medline Kit Number/SKU CDS984625B: UDI/DI 10193489262247 (each), 40193489262248 (case), Lot Number 21CDC556; Medline Kit Number/SKU CDS984625B: UDI/DI 10193489262247 (each), 40193489262248 (case), Lot Number 21BDA928; Medline Kit Number/SKU CDS985323F: UDI/DI 10889942162932 (each), 40889942162933 (case), Lot Number 21HBO566; Medline Kit Number/SKU CDS985323F: UDI/DI 10889942162932 (each), 40889942162933 (case), Lot Number 21FBT492; Medline Kit Number/SKU CDS985323G: UDI/DI 10193489941616 (each), 40193489941617 (case), Lot Number 22CBB025; Medline Kit Number/SKU CDS985323G: UDI/DI 10193489941616 (each), 40193489941617 (case), Lot Number 21KBZ638; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 23FBH534; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each).40195327091270 (case), Lot Number 23FBF357; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 23EBE981; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 23EBC221; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 22GBT619; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 22GBI155; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 22FBP699; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 22FBP701; Medline Kit Number/SKU CDS985323I: UDI/DI 10195327091279 (each), 40195327091270 (case), Lot Number 22EBN066; Medline Kit Number/SKU CDS985323L: UDI/DI 10198459114588 (each), 40198459114589 (case), Lot Number 24JBA196; Medline Kit Number/SKU DYNJ44978F: UDI/DI 10193489299663 (each), 40193489299664 (case), Lot Number 22CBF121; Medline Kit Number/SKU DYNJ44978F: UDI/DI 10193489299663 (each), 40193489299664 (case), Lot Number 21KBP694; Medline Kit Number/SKU DYNJ44978G: UDI/DI 10195327128319 (each), 40195327128310 (case), Lot Number 23CBH286; Medline Kit Number/SKU DYNJ44978G: UDI/DI 10195327128319 (each), 40195327128310 (case), Lot Number 22DBM896; Medline Kit Number/SKU DYNJ44978G: UDI/DI 10195327128319 (each), 40195327128310 (case), Lot Number 22DBL903; Medline Kit Number/SKU DYNJ44978I: UDI/DI 10195327381066 (each), 40195327381067 (case), Lot Number 23EBO372; Medline Kit Number/SKU DYNJ44978I: UDI/DI 10195327381066 (each), 40195327381067 (case), Lot Number 23DBJ211; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 26DBM877; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 26BBC110; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 26ABE787; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25LBG888; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25JBM189; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25IBH680; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25HBO120; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25GBG386; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25EBG327; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25CBV015; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25CBJ988; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25BBK455; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 25ABA475; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24JBF793; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24IBB421; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24HBR106; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24GBQ505; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24EBU081; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24DBL659; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24CBQ253; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 24ABA281; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23JBG066; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23GBN709; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23EBM697; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23CBU232; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23CBB181; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 23ABI042; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22KBR067; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22IBS775; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22HBI921; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22GBA106; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22DBR730; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22CBA492; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 22ABL656; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 21KBU212; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 21KBP732; Medline Kit Number/SKU DYNJ50752: UDI/DI 10889942020614 (each), 40889942020615 (case), Lot Number 21KBC682; Medline Kit Number/SKU DYNJ51102A: UDI/DI 10193489442557 (each), 40193489442558 (case), Lot Number 21HDB157; Medline Kit Number/SKU DYNJ51102A: UDI/DI 10193489442557 (each), 40193489442558 (case), Lot Number 21FDC202; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 23FDB023; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 23BDB687; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 22JDB774; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 22HDB159; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 22BDA713; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 21KDB067; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 21IDC429; Medline Kit Number/SKU DYNJ51102B: UDI/DI 10193489985337 (each), 40193489985338 (case), Lot Number 21IDB498; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 23EBW151; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 23DBF449; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 23CBH198; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 23ABT022; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 22KBK770; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 22IBT104; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 22GBP668; Medline Kit Number/SKU DYNJ53631G: UDI/DI 10195327145439 (each), 40195327145430 (case), Lot Number 22EBL876; Medline Kit Number/SKU DYNJ69553: UDI/DI 10193489952308 (each), 40193489952309 (case), Lot Number 21LBI619; Medline Kit Number/SKU DYNJ69553: UDI/DI 10193489952308 (each), 40193489952309 (case), Lot Number 21HBO585; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 23KBS338; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 23BBF310; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 22IBC321; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 22GBZ696; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 22CBH042; Medline Kit Number/SKU DYNJ69553A: UDI/DI 10195327087456 (each), 40195327087457 (case), Lot Number 22BBR408; Medline Kit Number/SKU DYNJ84181D: UDI/DI 10198459617515 (each), 40198459617516 (case), Lot Number 26CMK427; Medline Kit Number/SKU DYNJ84181D: UDI/DI 10198459617515 (each), 40198459617516 (case), Lot Number 26CME139; Medline Kit Number/SKU DYNJ84181D: UDI/DI 10198459617515 (each), 40198459617516 (case), Lot Number 26AMB018; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 25BBG681; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 24JBC021; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 24IBD289; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 24FBQ450; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 24DBN446; Medline Kit Number/SKU DYNJ84868: UDI/DI 10195327431488 (each), 40195327431489 (case), Lot Number 23IBV037; Medline Kit Number/SKU DYNJ84868A: UDI/DI 10198459334641 (each), 40198459334642 (case), Lot Number 26CBH602; Medline Kit Number/SKU DYNJ84868A: UDI/DI 10198459334641 (each), 40198459334642 (case), Lot Number 25KBD383; Medline Kit Number/SKU DYNJ84868A: UDI/DI 10198459334641 (each), 40198459334642 (case), Lot Number 25JBR763; Medline Kit Number/SKU DYNJ84868A: UDI/DI 10198459334641 (each), 40198459334642 (case), Lot Number 25DBH205; Medline Kit Number/SKU CDS983795G: UDI/DI 10193489591972 (each), 40193489591973 (case), Lot Number 21DKA467; Medline Kit Number/SKU CDS983795G: UDI/DI 10193489591972 (each), 40193489591973 (case), Lot Number 21BKA495; Medline Kit Number/SKU DYNJ903399C: UDI/DI 10193489411393 (each), 40193489411394 (case), Lot Number 21LBM301; Medline Kit Number/SKU DYNJ903399C: UDI/DI 10193489411393 (each), 40193489411394 (case), Lot Number 21JBQ450; Medline Kit Number/SKU DYNJ903399D: UDI/DI 10195327081003 (each), 40195327081004 (case), Lot Number 22BBB874; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 23GBR381; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 23FBH079; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 23EBE991; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 23CBH912; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 22IBI038; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 22IBF943; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 22IBE808; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 22FBV603; Medline Kit Number/SKU DYNJ903399F: UDI/DI 10195327115562 (each), 40195327115563 (case), Lot Number 22EBG782; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24JBL462; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24HBB484; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24FBO697; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24EBJ157; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24BBT947; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 24ABT586; Medline Kit Number/SKU DYNJ903399I: UDI/DI 10195327499624 (each), 40195327499625 (case), Lot Number 23LBF486; Medline Kit Number/SKU DYNJ903399J: UDI/DI 10198459187544 (each), 40198459187545 (case), Lot Number 25IBN080; Medline Kit Number/SKU DYNJ903399J: UDI/DI 10198459187544 (each), 40198459187545 (case), Lot Number 25GBT816; Medline Kit Number/SKU DYNJ903399J: UDI/DI 10198459187544 (each), 40198459187545 (case), Lot Number 25CBT658; Medline Kit Number/SKU DYNJ903399J: UDI/DI 10198459187544 (each), 40198459187545 (case), Lot Number 25ABU820; Medline Kit Number/SKU DYNJ903399J: UDI/DI 10198459187544 (each), 40198459187545 (case), Lot Number 24LBJ207; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23JMG995; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23HMG022; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23GMG363; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23EMH579; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23BMI023; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 23BMB813; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 22LMB376; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 22IMI547; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 22IMI275; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 22GMH460; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 22CMA356; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 21LMG596; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 21IMG034; Medline Kit Number/SKU DYNJ907893: UDI/DI 10193489921694 (each), 40193489921695 (case), Lot Number 21GMC854; Medline Kit Number/SKU DYNJ85910A: UDI/DI 10198459060021 (each), 40198459060022 (case), Lot Number 24GMI984; Medline Kit Number/SKU DYNJ85910A: UDI/DI 10198459060021 (each), 40198459060022 (case), Lot Number 24GME056; Medline Kit Number/SKU DYNJ47720B: UDI/DI 10889942813698 (each), 40889942813699 (case), Lot Number 21FMA687; Medline Kit Number/SKU DYNJ47720B: UDI/DI 10889942813698 (each), 40889942813699 (case), Lot Number 21BMC675; Medline Kit Number/SKU DYNJ47720B: UDI/DI 10889942813698 (each), 40889942813699 (case), Lot Number 21AME352; Medline Kit Number/SKU DYNJ58253A: UDI/DI 10195327381400 (each), 40195327381401 (case), Lot Number 24DBE454; Medline Kit Number/SKU DYNJ58253A: UDI/DI 10195327381400 (each), 40195327381401 (case), Lot Number 24BBM832; Medline Kit Number/SKU DYNJ58253A: UDI/DI 10195327381400 (each), 40195327381401 (case), Lot Number 23IBQ997; Medline Kit Number/SKU DYNJ58253A: UDI/DI 10195327381400 (each), 40195327381401 (case), Lot Number 23GBG184; Medline Kit Number/SKU DYNJ58253A: UDI/DI 10195327381400 (each), 40195327381401 (case), Lot Number 23EBQ417; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 24EBB487; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 24CBC257; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23LBM307; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23KBV679; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23JBO796; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23HBC840; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23EBM759; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 23CBT115; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 22JBG840; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 22JBF786; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 22GBR031; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 22BBI905; Medline Kit Number/SKU DYNJ64305A: UDI/DI 10193489354805 (each), 40193489354806 (case), Lot Number 21LBT867; Medline Kit Number/SKU DYNJ64305B: UDI/DI 10198459068348 (each), 40198459068349 (case), Lot Number 24GBJ335; Medline Kit Number/SKU DYNJ64305C: UDI/DI 10198459107627 (each), 40198459107628 (case), Lot Number 25DBD533; Medline Kit Number/SKU DYNJ64305C: UDI/DI 10198459107627 (each), 40198459107628 (case), Lot Number 25CBE848; Medline Kit Number/SKU DYNJ64305C: UDI/DI 10198459107627 (each), 40198459107628 (case), Lot Number 25BBP437; Medline Kit Number/SKU DYNJ64305C: UDI/DI 10198459107627 (each), 40198459107628 (case), Lot Number 25ABL380; Medline Kit Number/SKU DYNJ64305C: UDI/DI 10198459107627 (each), 40198459107628 (case), Lot Number 24JBH918; Medline Kit Number/SKU DYNJ64305F: UDI/DI 10198459512032 (each), 40198459512033 (case), Lot Number 26CMB989; Medline Kit Number/SKU DYNJ64305F: UDI/DI 10198459512032 (each), 40198459512033 (case), Lot Number 26AMF013; Medline Kit Number/SKU DYNJ64305F: UDI/DI 10198459512032 (each), 40198459512033 (case), Lot Number 25JMJ248; Medline Kit Number/SKU DYNJ64305F: UDI/DI 10198459512032 (each), 40198459512033 (case), Lot Number 25IMG041; Medline Kit Number/SKU DYNJ84023: UDI/DI 10195327327200 (each), 40195327327201 (case), Lot Number 23DBJ094; Medline Kit Number/SKU DYNJT5622: UDI/DI 10198459486678 (each), 40198459486679 (case), Lot Number 25HBO871; Medline Kit Number/SKU DYNJT8637: UDI/DI 10198459745874 (each), 40198459745875 (case), Lot Number 26DBJ715; Medline Kit Number/SKU DYNJ66127: UDI/DI 10193489318722 (each), 40193489318723 (case), Lot Number 22LMF981; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23LBI470; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23JBO003; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23JBF442; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23IBI912; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23FBK760; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23DBS638; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23DBD696; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23CBX453; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23CBL738; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23ABM813; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 23ABL453; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22JBP836; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22JBN994; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22FBI827; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22FBI831; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22DBH688; Medline Kit Number/SKU DYNJ907640A: UDI/DI 10195327092948 (each), 40195327092949 (case), Lot Number 22NBC390; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 25KBA598; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 25EBU310; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 25CBS456; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 25ABT749; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 24KBP147; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 24IBM827; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 24HBN686; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 24EBS841; Medline Kit Number/SKU DYNJ907640B: UDI/DI 10195327600877 (each), 40195327600878 (case), Lot Number 24CBA274; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 25DBQ664; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 25ABO087; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24KBI101; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24IBQ134; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24HBN945; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24FBI934; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24CBT603; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 24BBS358; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 23LBD162; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 23JBD350; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 23IBA180; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 23GBE725; Medline Kit Number/SKU DYNJ907033C: UDI/DI 10195327342623 (each), 40195327342624 (case), Lot Number 23BBR075; Medline Kit Number/SKU DYNJ42513C: UDI/DI 10193489305883 (each), 40193489305884 (case), Lot Number 21IBF732; Medline Kit Number/SKU DYNJ42513C: UDI/DI 10193489305883 (each), 40193489305884 (case), Lot Number 21GBU348; Medline Kit Number/SKU DYNJ42513C: UDI/DI 10193489305883 (each), 40193489305884 (case), Lot Number 21DBN337; Medline Kit Number/SKU DYNJ42513C: UDI/DI 10193489305883 (each), 40193489305884 (case), Lot Number 21ABJ702; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 26DMB652; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 25KMF663; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 25JMF966; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 25FMD566; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 25CMD681; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 24LMH198; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 24KMG072; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 24JMH713; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 24HMD698; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 23LME318; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 23JMB492; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 23GMG228; Medline Kit Number/SKU DYNJ04795K: UDI/DI 10195327401177 (each), 40195327401178 (case), Lot Number 23EMD236; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24HMF782; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24GME990; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24EMF954; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24CMG571; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24BBH452; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 24ABD849; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23KBW450; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23JBU872; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23IBK902; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23EBA300; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23EBA301; Medline Kit Number/SKU DYNJ0384702T: UDI/DI 10195327049683 (each), 40195327049684 (case), Lot Number 23BBN374; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 22OBK883; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 22BBR332; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 21HBT623; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 21HBA033; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 21FBA008; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 21ABU331; Medline Kit Number/SKU DYNJ902172D: UDI/DI 10193489454246 (each), 40193489454247 (case), Lot Number 21ABS927; Medline Kit Number/SKU DYNJ903038G: UDI/DI 10193489308464 (each), 40193489308465 (case), Lot Number 22LBB087; Medline Kit Number/SKU DYNJ903038G: UDI/DI 10193489308464 (each), 40193489308465 (case), Lot Number 22KBT094; Medline Kit Number/SKU DYNJ903038G: UDI/DI 10193489308464 (each), 40193489308465 (case), Lot Number 22HBR645; Medline Kit Number/SKU DYNJ903038G: UDI/DI 10193489308464 (each), 40193489308465 (case), Lot Number 22HBD859; Medline Kit Number/SKU DYNJ903038G: UDI/DI 10193489308464 (each), 40193489308465 (case), Lot Number 21HBI242; Medline Kit Number/SKU DYNJ909739: UDI/DI 10195327464691 (each), 40195327464692 (case), Lot Number 24EBB777; Medline Kit Number/SKU DYNJ909739: UDI/DI 10195327464691 (each), 40195327464692 (case), Lot Number 24CBR861; Medline Kit Number/SKU DYNJ909739: UDI/DI 10195327464691 (each), 40195327464692 (case), Lot Number 24BBL055; Medline Kit Number/SKU DYNJ909739: UDI/DI 10195327464691 (each), 40195327464692 (case), Lot Number 24BBK951; Medline Kit Number/SKU DYNJ909739: UDI/DI 10195327464691 (each), 40195327464692 (case), Lot Number 23IBU602; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 26CBG700; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 26BBE962; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 25JBX249; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 25IBB837; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 25CBG845; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 25BBI584; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 24LBA425; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 24KBM823; Medline Kit Number/SKU DYNJ909739B: UDI/DI 10198459150029 (each), 40198459150020 (case), Lot Number 24JBR776; Medline Kit Number/SKU DYNJ53443K: UDI/DI 10195327678210 (each), 40195327678211 (case), Lot Number 24GBW400; Medline Kit Number/SKU DYNJ53443K: UDI/DI 10195327678210 (each), 40195327678211 (case), Lot Number 24EMH930; Medline Kit Number/SKU DYNJ53443K: UDI/DI 10195327678210 (each), 40195327678211 (case), Lot Number 24DMI589; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23LBA317; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23JBQ173; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23GBQ781; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23FBB029; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23CBM203; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23BBA362; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 23ABF633; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 22JBJ594; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 22EBS240; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 22CBN170; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 22ABD619; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 21KBT107; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 21KBI681; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 21JBL916; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 21HBT471; Medline Kit Number/SKU DYNJ37262D: UDI/DI 10193489849813 (each), 40193489849814 (case), Lot Number 21FBR615; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 23BMB539; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22LMJ217; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22FMD179; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22FMD180; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22EMG632; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22DMA722; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22BMA081; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 22AMA617; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 21GMC054; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 21DMD964; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 21DMA229; Medline Kit Number/SKU DYNJ60348A: UDI/DI 10193489498127 (each), 40193489498128 (case), Lot Number 21BMD783; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 26CBA638; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 26ABT836; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25LBF426; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25KBD398; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25JBH387; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25IBL558; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25HBE457; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25FBP770; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25DBR847; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25DBH803; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25CBH890; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25BBV008; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25BBH914; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 25ABG058; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24KBO531; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24JBI783; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24IBH575; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24HBB493; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24FBL983; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24EBJ858; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24DBR078; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24CBN311; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24BBQ174; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 24BBF640; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23LBS287; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23JBW332; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23GBR443; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23GBG559; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23FBO011; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23EBL970; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23CBF150; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23CBF677; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 23ABJ600; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22JMD510; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22HMG437; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22GMH470; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22EMB738; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22DMD210; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22DMB491; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22CMB058; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 22AMC875; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21KMG046; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21JMB429; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21IMB476; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21HMB159; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21FMD876; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21DME229; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21CMD892; Medline Kit Number/SKU DYNJ43931C: UDI/DI 10193489431605 (each), 40193489431606 (case), Lot Number 21AMC903; Medline Kit Number/SKU CDS981546F: UDI/DI 10195327036881 (each), 40195327036882 (case), Lot Number 22GBM080; Medline Kit Number/SKU CDS981546F: UDI/DI 10195327036881 (each), 40195327036882 (case), Lot Number 22OBC361; Medline Kit Number/SKU CDS981546F: UDI/DI 10195327036881 (each), 40195327036882 (case), Lot Number 22ABS175; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 25DBJ433; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 25BBE279; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 24KBC789; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 24GBK279; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 24FBI142; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 24BBQ492; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23KBT549; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23JBD267; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23HBL944; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23GBG489; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23GBC448; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23DBC176; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 23BBD996; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 22JBH243; Medline Kit Number/SKU DYNJ81142: UDI/DI 10195327087548 (each), 40195327087549 (case), Lot Number 22DBB713; Medline Kit Number/SKU DYNJ81142A: UDI/DI 10198459472770 (each), 40198459472771 (case), Lot Number 26BBT553; Medline Kit Number/SKU DYNJ81142A: UDI/DI 10198459472770 (each), 40198459472771 (case), Lot Number 25LBB920; Medline Kit Number/SKU DYNJ81142A: UDI/DI 10198459472770 (each), 40198459472771 (case), Lot Number 25JBH283; Medline Kit Number/SKU DYNJ81142A: UDI/DI 10198459472770 (each), 40198459472771 (case), Lot Number 25HBA710; Medline Kit Number/SKU DYNJ907777: UDI/DI 10193489871814 (each), 40193489871815 (case), Lot Number 21DBS166; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 25LMG361; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 25GBB784; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 25CBH360; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 25BBE083; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 24FBP292; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 24DBH901; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 23KBD047; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 23IBP471; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 23AMD016; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 23AMC175; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 22LME215; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 22HMC652; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 22DMJ030; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 21LME917; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 21LMA492; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 21JMF565; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 21CMA636; Medline Kit Number/SKU DYNJ50832: UDI/DI 10889942027941 (each), 40889942027942 (case), Lot Number 21BMC060; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 25DBG095; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 25CBD398; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 24KBH831; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 24IBE079; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 24EBA963; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 24ABL743; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 23KBD039; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 23CBP385; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 23BBQ859; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 23BBK217; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 22JBK273; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 22EBJ322; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 21KBA357; Medline Kit Number/SKU DYNJ69733: UDI/DI 10193489974089 (each), 40193489974080 (case), Lot Number 21HBL834; Medline Kit Number/SKU DYNJ38054D: UDI/DI 10889942692118 (each), 40889942692119 (case), Lot Number 21IMG474; Medline Kit Number/SKU DYNJ38054D: UDI/DI 10889942692118 (each), 40889942692119 (case), Lot Number 21HMC986; Medline Kit Number/SKU DYNJ38054D: UDI/DI 10889942692118 (each), 40889942692119 (case), Lot Number 21EME051; Medline Kit Number/SKU DYNJ38054D: UDI/DI 10889942692118 (each), 40889942692119 (case), Lot Number 21AMA386.
Devices named in this recall
88 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. CDS984860FMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ906926BMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ901133AMedline Industries, Inc. · Gynecological laparoscopic kit3 recalls0 reports
- Medline Industries, Inc. DYNJ43931CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906931CMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ69733Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ51102BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0384702TMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ53820AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ53631GMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907407AMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907407BMedline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. DYNJ909739Medline Industries, Inc. · Gynecological laparoscopic kit2 recalls0 reports
- Medline Industries, Inc. CDS984625BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ44978FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS980810QMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ42513CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903038GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ66127Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ64305AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907407Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903399CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ51102AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ902172DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ68426Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ60348AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ61437Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS983795GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ54963AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ905924BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ905927BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900357OMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ905449BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ37256FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ37262DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907777Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907893Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS985323GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ69553Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ906047AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS981546FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903399DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ69553AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ81142Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS985323IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907640AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903399FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ44329IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ44978GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900357PMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ900357QMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ69307AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ84023Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907033CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ44978IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ58253AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ61437AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ04795KMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ909637Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ84868Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903399IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ909989Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ61437BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ37256GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ907640BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ53443KMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS984860IMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ80875FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ80875GMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ85910AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ64305BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ905924CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ64305CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS985323LMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ909739BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ903399JMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ61437CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ905924DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ906047CMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ44329JMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ84868AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ81142AMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ50752Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ50832Medline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. CDS985323FMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ37256DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ38054DMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
- Medline Industries, Inc. DYNJ47720BMedline Industries, Inc. · Gynecological laparoscopic kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98994).
- Z-2479-2026 · Open, ClassifiedMEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 F…
- Z-2480-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
- Z-2481-2026 · Open, ClassifiedMEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
- Z-2482-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
- Z-2483-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C
- Z-2484-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C
- Z-2485-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A
- Z-2486-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit…
- Z-2487-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CH OPEN HEART, ADULT CARDIAC KIT, AHT CVOR CABG, APH EMERGENCY OPEN HEART PACK, Etc. (see recall documents for a full list of products)
- Z-2488-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR M…
- Z-2489-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: BILAT MICRO VASECTOMY REVERSAL, Medline Kit Number/SKU DYNJ68302
- Z-2491-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)
- Z-2492-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
- Z-2493-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CHEST EXPLORATION ECMO, Medline Kit Number/SKU DYNJ63789A; 2) CORONARY/ARTERY BYPASS TRAY, Medline Kit Number/SKU DYNJ17321G; 3) D…
- Z-2494-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A
- Z-2495-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLA…
- Z-2496-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN…
- Z-2497-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as BIOBURDEN TEST KIT, Kit Uro Robot Cystectomy, etc. (see the recall documents for a full list of products)
- Z-2498-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808
- Z-2500-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medlin…
- Z-2501-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline …
- Z-2502-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
- Z-2504-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.
- Z-2505-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, …
- Z-2506-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall doc…
- Z-2507-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC…
- Z-2508-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ90809…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page