MEDLINE Medical Procedure Kits labeled as recallZ-2507-2026 · Medline Industries, LP · initiated May 4, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2507-2026
- Status
- Open, Classified
- Initiated by the company
- May 4, 2026
- Posted by the FDA
- Jun 17, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 2340 units
- Product code
- OJK
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Product description
MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 24BBF438; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23LBR516; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23LBI385; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBW083; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBT137; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBA705; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23EBT206; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBN853; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBG437; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBG444; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBA852; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBA859; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22IBR361; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22IBF066; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22GBA053; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22GBA052; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22GBA054; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22DBM893; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22DBS377; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22OBE639; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22ABB399; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22ABB400; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 24ABX491; Medline Kit Number/SKU CDS983775J: UDI/DI 10195327684761 (each), 40195327684762 (case), Lot Number 24GBG807; Medline Kit Number/SKU CDS983775J: UDI/DI 10195327684761 (each), 40195327684762 (case), Lot Number 24FBI951; Medline Kit Number/SKU CDS983775J: UDI/DI 10195327684761 (each), 40195327684762 (case), Lot Number 24EBH007; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25GMD344; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25EMG590; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25EMC381; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25CMJ064; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25CMC688; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 25BMC192; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 24KBS115; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 24KBF283; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 24JBD646; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 24IBK663; Medline Kit Number/SKU CDS983775K: UDI/DI 10198459086991 (each), 40198459086992 (case), Lot Number 24IBI407; Medline Kit Number/SKU CDS983775L: UDI/DI 10198459518416 (each), 40198459518417 (case), Lot Number 25LBA083; Medline Kit Number/SKU CDS983775L: UDI/DI 10198459518416 (each), 40198459518417 (case), Lot Number 25LBA639; Medline Kit Number/SKU CDS983775L: UDI/DI 10198459518416 (each), 40198459518417 (case), Lot Number 25KBG543; Medline Kit Number/SKU CDS983775L: UDI/DI 10198459518416 (each), 40198459518417 (case), Lot Number 25KBB188; Medline Kit Number/SKU CDS983775L: UDI/DI 10198459518416 (each), 40198459518417 (case), Lot Number 25JBH879; Medline Kit Number/SKU CDS983775M: UDI/DI 10198459658358 (each), 40198459658359 (case), Lot Number 26CBF631; Medline Kit Number/SKU DYNJ84480: UDI/DI 10195327370121 (each), 40195327370122 (case), Lot Number 23LBU479; Medline Kit Number/SKU DYNJ84480: UDI/DI 10195327370121 (each), 40195327370122 (case), Lot Number 23JBY640; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 22LDB219; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 22GDB372; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 22CDA529; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 21LDC134; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 21KDA904; Medline Kit Number/SKU DYNJVB1080: UDI/DI 10193489284492 (each), 40193489284493 (case), Lot Number 21JDA634; Medline Kit Number/SKU DYNJVB1080A: UDI/DI 10195327325695 (each), 40195327325696 (case), Lot Number 23IDA988; Medline Kit Number/SKU DYNJVB1080A: UDI/DI 10195327325695 (each), 40195327325696 (case), Lot Number 23HDA236; Medline Kit Number/SKU DYNJVB1080A: UDI/DI 10195327325695 (each), 40195327325696 (case), Lot Number 23CDB306.
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJVB1080Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. CDS983775IMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJVB1080AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ84480Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. CDS983775JMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. CDS983775KMedline Industries, Inc. · Eye tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98994).
- Z-2479-2026 · Open, ClassifiedMEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 F…
- Z-2480-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
- Z-2481-2026 · Open, ClassifiedMEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
- Z-2482-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
- Z-2483-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C
- Z-2484-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C
- Z-2485-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A
- Z-2486-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit…
- Z-2487-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CH OPEN HEART, ADULT CARDIAC KIT, AHT CVOR CABG, APH EMERGENCY OPEN HEART PACK, Etc. (see recall documents for a full list of products)
- Z-2488-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR M…
- Z-2489-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: BILAT MICRO VASECTOMY REVERSAL, Medline Kit Number/SKU DYNJ68302
- Z-2491-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)
- Z-2492-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
- Z-2493-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CHEST EXPLORATION ECMO, Medline Kit Number/SKU DYNJ63789A; 2) CORONARY/ARTERY BYPASS TRAY, Medline Kit Number/SKU DYNJ17321G; 3) D…
- Z-2494-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A
- Z-2495-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLA…
- Z-2496-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN…
- Z-2497-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as BIOBURDEN TEST KIT, Kit Uro Robot Cystectomy, etc. (see the recall documents for a full list of products)
- Z-2498-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808
- Z-2500-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medlin…
- Z-2501-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline …
- Z-2502-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
- Z-2503-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR P…
- Z-2504-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.
- Z-2505-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, …
- Z-2506-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall doc…
- Z-2508-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ90809…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page