MEDLINE Medical Procedure Kits labeled as recallZ-2508-2026 · Medline Industries, LP · initiated May 4, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2508-2026
- Status
- Open, Classified
- Initiated by the company
- May 4, 2026
- Posted by the FDA
- Jun 17, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 3521 units
- Product code
- OJH
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Product description
MEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ908097B; 4) CRANI, Medline Kit Number/SKU DYNJ908097C; 5) CRANI, Medline Kit Number/SKU DYNJ908097D; 6) DR TUCKER ANT TOTAL HIP, Medline Kit Number/SKU DYNJ912029; 7) JEWISH CRANI PACK, Medline Kit Number/SKU DYNJ55637B; 8) L SCOPE PACK, Medline Kit Number/SKU DYNJ43930A; 9) LAMI PACK, Medline Kit Number/SKU DYNJ67767; 10) LAVH NEW BRAUNFELS, Medline Kit Number/SKU DYNJ61136F; 11) SOI KNEE ARTHROSCOPY, Medline Kit Number/SKU DYNJV0172Q; 12) T AND A PACK, Medline Kit Number/SKU DYNJ86229.
Code information (models, lots, serial numbers)
Show the full code information (9k characters)
Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 23FBQ975; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 23FBF166; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 23DBM780; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22KBM832; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22JBY524; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22GBR089; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22FBR153; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22FBC633; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22EBB478; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22BBX706; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22BBU902; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 22ABN864; Medline Kit Number, SKU DYNJ908097: UDI/DI 10193489973846 (each), 40193489973847 (case), Lot Number 21IBO640; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 24EBC216; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 24CBU878; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 23LBV779; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 23JBV724; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 23JBQ243; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 23JBQ188; Medline Kit Number, SKU DYNJ908097A: UDI/DI 10195327507077 (each), 40195327507078 (case), Lot Number 23IBS778; Medline Kit Number, SKU DYNJ908097B: UDI/DI 10198459043499 (each), 40198459043490 (case), Lot Number 24IBU230; Medline Kit Number, SKU DYNJ908097B: UDI/DI 10198459043499 (each), 40198459043490 (case), Lot Number 24HBF200; Medline Kit Number, SKU DYNJ908097B: UDI/DI 10198459043499 (each), 40198459043490 (case), Lot Number 24GBJ113; Medline Kit Number, SKU DYNJ908097B: UDI/DI 10198459043499 (each), 40198459043490 (case), Lot Number 24GBB611; Medline Kit Number, SKU DYNJ908097C: UDI/DI 10198459129773 (each), 40198459129774 (case), Lot Number 24LBI207; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 26BBN470; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25LBO007; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25KBT991; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25EBW026; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25EBA043; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25EBA044; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25CBG395; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25CBC209; Medline Kit Number, SKU DYNJ908097D: UDI/DI 10198459237782 (each), 40198459237783 (case), Lot Number 25BBH141; Medline Kit Number, SKU DYNJ912029: UDI/DI 10198459600197 (each), 40198459600198 (case), Lot Number 26ABG299; Medline Kit Number, SKU DYNJ55637B: UDI/DI 10889942886418 (each), 40889942886419 (case), Lot Number 21HMB506; Medline Kit Number, SKU DYNJ55637B: UDI/DI 10889942886418 (each), 40889942886419 (case), Lot Number 21CMB156; Medline Kit Number, SKU DYNJ55637B: UDI/DI 10889942886418 (each), 40889942886419 (case), Lot Number 21BMA535; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 26BBQ722; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25LBT225; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25JBV630; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25JBA301; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25GBG702; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25FBI032; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25EBV203; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 25BBU777; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 24LBN665; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 24JBH892; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 24FBR472; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 24DBU491; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 24ABU986; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23JBH124; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23HBQ073; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23GBR436; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23FBS680; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23EBU086; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 23BBK246; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22LBE875; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22JBD615; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22HBT661; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22GMH469; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22EMH976; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 22AME279; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21LMD434; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21JMC555; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21JMB332; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21HMG155; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21GMC422; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21FMF742; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21EMF262; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21DMD509; Medline Kit Number, SKU DYNJ43930A: UDI/DI 10889942624065 (each), 40889942624066 (case), Lot Number 21AMC934; Medline Kit Number, SKU DYNJ67767: UDI/DI 10193489430615 (each), 40193489430616 (case), Lot Number 21ABW707; Medline Kit Number, SKU DYNJ67767: UDI/DI 10193489430615 (each), 40193489430616 (case), Lot Number 21ABM797; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 26ABL734; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 25ABO431; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 24LBH715; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 24KBB752; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 24JBJ191; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 24DBP178; Medline Kit Number, SKU DYNJ61136F: UDI/DI 10195327606169 (each), 40195327606160 (case), Lot Number 24BBF170; Medline Kit Number, SKU DYNJV0172Q: UDI/DI 10193489336238 (each), 40193489336239 (case), Lot Number 21KBS950; Medline Kit Number, SKU DYNJ86229: UDI/DI 10195327604967 (each), 40195327604968 (case), Lot Number 24GMJ630; Medline Kit Number, SKU DYNJ86229: UDI/DI 10195327604967 (each), 40195327604968 (case), Lot Number 24CMG797.
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ67767Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908097Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908097AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43930AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJV0172QMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ86229Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ61136FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908097BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908097CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908097DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ55637BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98994).
- Z-2479-2026 · Open, ClassifiedMEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 F…
- Z-2480-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
- Z-2481-2026 · Open, ClassifiedMEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
- Z-2482-2026 · Open, ClassifiedMEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
- Z-2483-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C
- Z-2484-2026 · Open, ClassifiedMEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C
- Z-2485-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A
- Z-2486-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit…
- Z-2487-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CH OPEN HEART, ADULT CARDIAC KIT, AHT CVOR CABG, APH EMERGENCY OPEN HEART PACK, Etc. (see recall documents for a full list of products)
- Z-2488-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR M…
- Z-2489-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: BILAT MICRO VASECTOMY REVERSAL, Medline Kit Number/SKU DYNJ68302
- Z-2491-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)
- Z-2492-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
- Z-2493-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CHEST EXPLORATION ECMO, Medline Kit Number/SKU DYNJ63789A; 2) CORONARY/ARTERY BYPASS TRAY, Medline Kit Number/SKU DYNJ17321G; 3) D…
- Z-2494-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A
- Z-2495-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) KIT UROLOGY URETHROPLASTY, Medline Kit Number/SKU DYKMBNDL117C; 2) URETHROPLASTY, Medline Kit Number/SKU DYNJ908978; 3) URETHROPLA…
- Z-2496-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN…
- Z-2497-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as BIOBURDEN TEST KIT, Kit Uro Robot Cystectomy, etc. (see the recall documents for a full list of products)
- Z-2498-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: DR FERNANDEZ, Medline kit number/SKU DYNJ68808
- Z-2500-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medlin…
- Z-2501-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline …
- Z-2502-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
- Z-2503-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR P…
- Z-2504-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.
- Z-2505-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, …
- Z-2506-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall doc…
- Z-2507-2026 · Open, ClassifiedMEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.