HeartMate 3 recallZ-2518-2026 · Thoratec LLC · initiated May 13, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Thoratec LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to 11 volt Backup Battery failures.
FDA root cause category: Under Investigation by firm
What the company is doing
On 05/13/2026, the firm sent email an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that As of April 27, 2026, Abbott received twenty-four (24) complaints where an advisory warning, Replace Backup Battery, was triggered once certain 11V Backup Batteries were connected to the System Controller. This warning is displayed as a yellow banner on the HeartMate Touch Monitor and System Monitor. This advisory alert is only visible when using HeartMate Touch or the System Monitor in the clinic, it is not visible to the patient. Customers are instructed to: 1. Read this notification thoroughly, complete and sign the form included with this letter, acknowledging that you received and understand this information, and that you will communicate this information to those relevant in your facility. Once complete, return the acknowledgement form to Abbott. 2. Use the online lookup tool (https://www.cardiovascular.abbott/us/en/hcp/product-advisories/heartmatebatteries-2026.html) to confirm if a 11V Lithium-Ion battery at your site is an impacted one. Figure 3 shows an instruction on how to locate the serial numbers of the impacted batteries. 3. Once the impacted batteries are identified, do not use them. 4. Beginning May 15, 2026, a separate email will be sent to your site confirming shipment of replacement batteries. This email will include return shipping labels and a device retrieval form, along with detailed instructions for returning the affected batteries. For Questions - contact your Abbott representative or Abbott Technical Support at 1-800-456-1477 (Business hours: 8 AM EST to 7 PM EST).
Recall details
- Recall number
- Z-2518-2026
- Status
- Open, Classified
- Initiated by the company
- May 13, 2026
- Posted by the FDA
- Jun 18, 2026
- Recalling firm
- Thoratec LLC
- Quantity
- 4
- Product code
- DSQ
- FDA submissions
- P160054
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Canada, Germany, Japan, Netherlands, Poland, Spain.
Product description
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No
Code information (models, lots, serial numbers)
Lot Code: Model Number: 106524US UDI/GTIN 05415067039323 Serial Numbers Distributed in US: See Attachment E for serial numbers distributed.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HeartMate 3™ 106524USThoratec Corporation · Ventricular (assist) bypass1 recall123 deaths207 reports
Product lines
Other devices under the same FDA submission
24 devices cleared under P160054 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HeartMate™ 107754Thoratec Corporation · Ventricular (assist) bypass2 recalls447 reports
- HeartMate 3™ 106531USThoratec Corporation · Ventricular (assist) bypass2 recalls72 reports
- HeartMate Touch™ HMT1100Thoratec Corporation · Ventricular (assist) bypass1 recall50 reports
- Thoratec® HeartMate 3™ 106531LF2Thoratec Corporation · Ventricular (assist) bypass1 recall12 reports
- HeartMate™ 107760Thoratec Corporation · Ventricular (assist) bypass1 recall6 reports
- HeartMate Touch™ HMT1100-RThoratec Corporation · Ventricular (assist) bypass1 recall3 reports
- HeartMate 3™ 106531LF2Thoratec Corporation · Ventricular (assist) bypass1 recall0 reports
- HeartMate™ 106128Thoratec Corporation · Ventricular (assist) bypass1 recall0 reports
- Thoratec® HeartMate 3™ 10005877Thoratec Corporation · Ventricular (assist) bypass22 reports
- HeartMate 3™ 106525USThoratec Corporation · Ventricular (assist) bypass13 reports
- Thoratec® HeartMate 3™ 10005872Thoratec Corporation · Ventricular (assist) bypass10 reports
- HeartMate Touch™ HMT2100Thoratec Corporation · Ventricular (assist) bypass10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99023).
- Z-2519-2026 · Open, ClassifiedBrand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A Product Description: The HeartMate 3" System Controller acts…
- Z-2520-2026 · Open, ClassifiedBrand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Stan…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page