Acumed Hexalob Screw various sizes recallZ-2532-2026 · Acumed LLC · initiated May 28, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Acumed LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

FDA root cause category: Device Design

What the company is doing

On May 28, 2026 Acumed issued a Urgent Medical Device Recall Notification to affected consignees via phone calls and email. Acumed asked consignees to take the following actions: 1)IMMEDIATELY stop using and/or distributing the product and complete the following actions: 2) Perform a physical count of your inventory. Identify any inventory of the batch/lot numbers listed in the table above. 3) Remove the product segregated in step 1, such that they are taken out of service (i.e., quarantined). 3) Verbally acknowledge (then respond via email) to [email protected] the recall notice. 4) You will be contacted by Field Inventory department who will directly remove the screws and conduct a count reconciliation. 5) This notice needs to be passed on to all those who need to be aware within your organization or facilities where the affected devices have been transferred.

Recall details

Recall number
Z-2532-2026
Status
Open, Classified
Initiated by the company
May 28, 2026
Posted by the FDA
Jun 23, 2026
Recalling firm
Acumed LLC
Quantity
5,272 units
Product code
HRS
FDA submissions
K251132, K251296
Distribution
US Nationwide distribution in the states of AR, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MS, NC, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI.

Product description

Acumed Hexalob Screw various sizes: REF Numbers: : 3070-27008 2.7mm x 8mm NL Hexalobe Screw; 3070-27009 2.7mm x 9mm NL Hexalobe Screw; 3070-27010 2.7mm x 10mm NL Hexalobe Screw; 3070-27011 2.7mm x 11mm NL Hexalobe Screw; 3070-27012 2.7mm x 12mm NL Hexalobe Screw; 3070-27013 2.7mm x 13mm NL Hexalobe Screw; 3070-27014 2.7mm x 14mm NL Hexalobe Screw; 3070-27015 2.7mm x 15mm NL Hexalobe Screw; 3070-27016 2.7mm x 16mm NL Hexalobe Screw; 3070-27017 2.7mm x 17mm NL Hexalobe Screw; 3070-27018 2.7mm x 18mm NL Hexalobe Screw; 3076-30010 3.0mm x 10mm NL Hexalobe Screw; 3076-30012 3.0mm x 12mm NL Hexalobe Screw; 3076-30014 3.0mm x 14mm NL Hexalobe Screw

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

3070-27008/Lots: 668261/UDI: 10806378125997; 3070-27009/Lots: 680797/UDI: 10806378126000; 3070-27010/Lots: 667231, 682194/UDI: 10806378126017; 3070-27011/Lots: 680621, 680737/UDI: 10806378126024; 3070-27012/Lots: 667233, 668266, 668704, 682195/UDI: 10806378126031; 3070-27013/Lots:668706, 680400/UDI: 10806378126048; 3070-27014/Lots: 667236, 668705, 680935/UDI: 10806378126055; 3070-27015/Lots: 667237, 680936/UDI: 10806378126062; 3070-27016/Lots: 668269, 682390/UDI: 10806378126079; 3070-27017/Lot: 681575/UDI: 10806378126086; 3070-27018/Lot: 667169, 668260, 682482/UDI: 10806378126093; 3076-30010/Lots: 667202, 668271/UDI: 10806378126109; 3076-30012/Lots: 667242, 668273/UDI: 10806378126116; 3076-30014/Lots: 667205, 668272/UDI: 10806378126123

Devices named in this recall

14 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K251132, K251296 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.