ACUMED 3070-27015Recalls & reported problems
Acumed LLC · 2.7mm x 15mm NL Hexalobe Screw · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10806378126062), received between Feb 12, 2026 and Apr 1, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 28, 2026Open, Classified
Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.
Z-2532-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Material Fragmentation1
Patient problems
- Device Embedded In Tissue or Plaque3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2026Break
IT WAS REPORTED: 3 OF THE 2.7MM NL SCREWS FROM THE ACULOC NEXT SET BROKE WHEN DRILLING IN THE PROXIMAL OBLONG FOR THE LONGER OF THE NEW NEXT WRIST SPANNING PLATES. PATIENT WAS ~40 YEAR OLD MALE. GOOD DIAPHYSEAL BONE. WE ARE WONDERING IF THERE NEEDS TO BE A DIFFERENT LARGER DRILL OR A TAP AVAILABLE FOR DIAPHYSEAL BONE. THE SCREWS DID NOT BREAK RIGHT BELOW THE SCREW HEAD, IT WAS A FEW THREADS DOWN. BASICALLY WHENEVER THE BONE STARTED GIVING SOME RESISTANCE THEY SNAPPED, BUT IT WAS FELLOW WITH THE SURGEON AND HE WAS VERY CONTROLLED AND WASN'T OVER TORQUING. SCREWS SHOULD NOT HAVE BEHAVED LIKE THAT EVEN WITH SOME RESISTANCE. IT WAS VERY ODD. WE LEFT TWO IN THE OBLONG WITH THE HEADS SHEARED OFF AND A THIRD IN A SCREW HOLE DISTAL I BELIEVE. WE HAD TWO SCREWS PROXIMAL THAT WE LOCKED AROUND THOSE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24758789 ↗InjuryApr 1, 2026Break
IT WAS REPORTED: 3 OF THE 2.7MM NL SCREWS FROM THE ACULOC NEXT SET BROKE WHEN DRILLING IN THE PROXIMAL OBLONG FOR THE LONGER OF THE NEW NEXT WRIST SPANNING PLATES. PATIENT WAS 40-YEAR-OLD MALE. GOOD DIAPHYSEAL BONE. WE ARE WONDERING IF THERE NEEDS TO BE A DIFFERENT LARGER DRILL OR A TAP AVAILABLE FOR DIAPHYSEAL BONE. THE SCREWS DID NOT BREAK RIGHT BELOW THE SCREW HEAD; IT WAS A FEW THREADS DOWN. BASICALLY, WHENEVER THE BONE STARTED GIVING SOME RESISTANCE THEY SNAPPED, BUT IT WAS FELLOW WITH THE SURGEON AND HE WAS VERY CONTROLLED AND WASN'T OVER TORQUING. SCREWS SHOULD NOT HAVE BEHAVED LIKE THAT EVEN WITH SOME RESISTANCE. IT WAS VERY ODD. WE LEFT TWO IN THE OBLONG WITH THE HEADS SHEARED OFF AND A THIRD IN A SCREW HOLE DISTAL I BELIEVE. WE HAD TWO SCREWS PROXIMAL THAT WE LOCKED AROUND THOSE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24758786 ↗InjuryFeb 12, 2026Material Fragmentation
IT WAS REPORTED THAT THE SCREW BROKE IN THE DISTAL END OF THE SLOT IN THE WRIST SPANNING PLATE. IT SHEARED OFF RIGHT AT THE BONE SURFACE WITH ABOUT ONE THREAD LEFT UNDER THE HEAD OF THE SCREW. THE SURGEON MENTIONED IT WAS DENSE BONE AND HE LEFT THE BROKEN SCREW IN THE BONE AND MENTIONED THAT HE WOULD REMOVE IT USING A TREPHINE WHEN HE REMOVES THE WSP. HE MENTIONED THAT HE WASN'T TOO CONCERNED ABOUT THE REMOVAL BECAUSE THE BONE QUALITY WAS VERY GOOD AND THE PATIENT IS YOUNG. HE WAS ABLE TO INSERT A SECOND NL SCREW INTO THE PROXIMAL PORTION OF THE SAME SLOT, SO THE BREAKAGE DIDN'T AFFECT HIS FIXATION.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24328622 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page