Skip to content

ACUMED 3070-27015Recalls & reported problems

Acumed LLC · 2.7mm x 15mm NL Hexalobe Screw · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10806378126062), received between Feb 12, 2026 and Apr 1, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 28, 2026Open, Classified

    Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

    Z-2532-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Material Fragmentation1

Patient problems

  • Device Embedded In Tissue or Plaque3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2026Break

    IT WAS REPORTED: 3 OF THE 2.7MM NL SCREWS FROM THE ACULOC NEXT SET BROKE WHEN DRILLING IN THE PROXIMAL OBLONG FOR THE LONGER OF THE NEW NEXT WRIST SPANNING PLATES. PATIENT WAS ~40 YEAR OLD MALE. GOOD DIAPHYSEAL BONE. WE ARE WONDERING IF THERE NEEDS TO BE A DIFFERENT LARGER DRILL OR A TAP AVAILABLE FOR DIAPHYSEAL BONE. THE SCREWS DID NOT BREAK RIGHT BELOW THE SCREW HEAD, IT WAS A FEW THREADS DOWN. BASICALLY WHENEVER THE BONE STARTED GIVING SOME RESISTANCE THEY SNAPPED, BUT IT WAS FELLOW WITH THE SURGEON AND HE WAS VERY CONTROLLED AND WASN'T OVER TORQUING. SCREWS SHOULD NOT HAVE BEHAVED LIKE THAT EVEN WITH SOME RESISTANCE. IT WAS VERY ODD. WE LEFT TWO IN THE OBLONG WITH THE HEADS SHEARED OFF AND A THIRD IN A SCREW HOLE DISTAL I BELIEVE. WE HAD TWO SCREWS PROXIMAL THAT WE LOCKED AROUND THOSE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24758789 ↗
  • InjuryApr 1, 2026Break

    IT WAS REPORTED: 3 OF THE 2.7MM NL SCREWS FROM THE ACULOC NEXT SET BROKE WHEN DRILLING IN THE PROXIMAL OBLONG FOR THE LONGER OF THE NEW NEXT WRIST SPANNING PLATES. PATIENT WAS 40-YEAR-OLD MALE. GOOD DIAPHYSEAL BONE. WE ARE WONDERING IF THERE NEEDS TO BE A DIFFERENT LARGER DRILL OR A TAP AVAILABLE FOR DIAPHYSEAL BONE. THE SCREWS DID NOT BREAK RIGHT BELOW THE SCREW HEAD; IT WAS A FEW THREADS DOWN. BASICALLY, WHENEVER THE BONE STARTED GIVING SOME RESISTANCE THEY SNAPPED, BUT IT WAS FELLOW WITH THE SURGEON AND HE WAS VERY CONTROLLED AND WASN'T OVER TORQUING. SCREWS SHOULD NOT HAVE BEHAVED LIKE THAT EVEN WITH SOME RESISTANCE. IT WAS VERY ODD. WE LEFT TWO IN THE OBLONG WITH THE HEADS SHEARED OFF AND A THIRD IN A SCREW HOLE DISTAL I BELIEVE. WE HAD TWO SCREWS PROXIMAL THAT WE LOCKED AROUND THOSE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24758786 ↗
  • InjuryFeb 12, 2026Material Fragmentation

    IT WAS REPORTED THAT THE SCREW BROKE IN THE DISTAL END OF THE SLOT IN THE WRIST SPANNING PLATE. IT SHEARED OFF RIGHT AT THE BONE SURFACE WITH ABOUT ONE THREAD LEFT UNDER THE HEAD OF THE SCREW. THE SURGEON MENTIONED IT WAS DENSE BONE AND HE LEFT THE BROKEN SCREW IN THE BONE AND MENTIONED THAT HE WOULD REMOVE IT USING A TREPHINE WHEN HE REMOVES THE WSP. HE MENTIONED THAT HE WASN'T TOO CONCERNED ABOUT THE REMOVAL BECAUSE THE BONE QUALITY WAS VERY GOOD AND THE PATIENT IS YOUNG. HE WAS ABLE TO INSERT A SECOND NL SCREW INTO THE PROXIMAL PORTION OF THE SAME SLOT, SO THE BREAKAGE DIDN'T AFFECT HIS FIXATION.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24328622 ↗

Other Screw, fixation, bone devices