ACUMED 3070-27016Recalls & reported problems

Acumed LLC · 2.7mm x 16mm NL Hexalobe Screw · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10806378126079), received between Apr 1, 2026 and May 6, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 28, 2026Open, Classified

    Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

    Z-2532-2026 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Material Fragmentation1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 6, 2026Material Fragmentation

    IT WAS REPORTED: 2 2.7MM NL SCREWS FROM AL NEXT BROKE DURING THE CASE. THE FIRST SCREW (16MM) BROKE WHILE THE SCREW WAS BEING USED TO PULL THE PLATE TO THE BONE. THE HEAD OF THE SCREW BROKE IN HALF AND THEN FELL THROUGH THE SLOT IN THE PLATE. THE SECOND SCREW HEAD SNAPPED OFF THE SCREW BEFORE IT WAS EVEN DOWN TO THE PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25104615 ↗
  • InjuryApr 1, 2026Break

    IT WAS REPORTED: 3 OF THE 2.7MM NL SCREWS FROM THE ACULOC NEXT SET BROKE WHEN DRILLING IN THE PROXIMAL OBLONG FOR THE LONGER OF THE NEW NEXT WRIST SPANNING PLATES. PATIENT WAS 40-YEAR-OLD MALE. GOOD DIAPHYSEAL BONE. WE ARE WONDERING IF THERE NEEDS TO BE A DIFFERENT LARGER DRILL OR A TAP AVAILABLE FOR DIAPHYSEAL BONE. THE SCREWS DID NOT BREAK RIGHT BELOW THE SCREW HEAD; IT WAS A FEW THREADS DOWN. BASICALLY, WHENEVER THE BONE STARTED GIVING SOME RESISTANCE THEY SNAPPED, BUT IT WAS FELLOW WITH THE SURGEON AND HE WAS VERY CONTROLLED AND WASN'T OVER TORQUING. SCREWS SHOULD NOT HAVE BEHAVED LIKE THAT EVEN WITH SOME RESISTANCE. IT WAS VERY ODD. WE LEFT TWO IN THE OBLONG WITH THE HEADS SHEARED OFF AND A THIRD IN A SCREW HOLE DISTAL I BELIEVE. WE HAD TWO SCREWS PROXIMAL THAT WE LOCKED AROUND THOSE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24758788 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →