Brand Name: GEM Premier 7000 with iQM3 recallZ-2822-2026 · Instrumentation Laboratory · initiated May 8, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
FDA root cause category: Software Design Change
What the company is doing
Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
Recall details
- Recall number
- Z-2822-2026
- Status
- Open, Classified
- Initiated by the company
- May 8, 2026
- Posted by the FDA
- Jul 29, 2026
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 4 US
- Product code
- CHL
- FDA submissions
- K223608
- Distribution
- US Nationwide distribution.
Product description
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.
Code information (models, lots, serial numbers)
Lot Code: GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: 4 affected BCR wands on the GEM Premier 7000 in the US. GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024015859 UDI: 08426950869634 Units distributed in US: 4
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- GEM Premier 7000 with iQM3 00000015279Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- Wand, Bar Code 00024015859Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K223608 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GEM Premier 7000 with iQM3 PAK (BG/ISE/GL/COOX 300 Tests) 00077430010Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- GEM Premier 7000 with iQM3 PAK (BG/ISE/GL/COOX 150 Tests) 00077415010Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- GEM Premier 7000 with iQM3 PAK (BG/ISE/GL/COOX 75 Tests) 00077407510Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- GEM Premier 7000 with iQM3 PAK (7k BG/ISE/COOX 150 Tests) 00077415008Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- GEM Premier 7000 with iQM3 PAK (BG/ISE/COOX 75 Tests) 00077407508Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
- GEM Premier 7000 with iQM3 PAK (BG/COOX 150 Tests) 00077415004Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99115).
- Z-2821-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment…
- Z-2823-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
- Z-2824-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
- Z-2825-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
- Z-2826-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page