Skip to content

Brand Name: GEM Premier 5000 recallZ-2824-2026 · Instrumentation Laboratory · initiated May 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.

FDA root cause category: Software Design Change

What the company is doing

Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.

Recall details

Recall number
Z-2824-2026
Status
Open, Classified
Initiated by the company
May 8, 2026
Posted by the FDA
Jul 29, 2026
Recalling firm
Instrumentation Laboratory
Quantity
67
Product code
CHL
FDA submissions
K203790
Distribution
US Nationwide distribution.

Product description

Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Lot Code: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in United Kingdom: 67 affected BCR wands on the GEM Premier 5000 in the United Kingdom. GEM Premier 5000 Serial Nos.: 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 24011137, 25081799, 25081797, 25091839, 25091840, 25091864, 25091870, 25091868, 25091873, 25091880, 25091876, 25101886, 25101889, 25101887, 25101891, 25101907, 25101904, 25101912, 25101905, 25101917, 25101921, 25101926, 25101929, 25101934, 25101927, 25111968, 25121986, 25121985, 25121981, 25121984, 25121994, 25121995, 25121982, 25121993, 25121991, 25121999, 25121996, 16030255, 17060782, 17060784, 17060785, 17060789, 17060790, 17060797, 17111126, 17121258, 18011414, 20054515, 20054516 and 22082271. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024019255 UDI: 08426950870227 Units distributed in the United Kingdom: 67

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

21 devices cleared under K203790 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99115).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page