Brand Name: GEM Premier 5000 recallZ-2826-2026 · Instrumentation Laboratory · initiated May 8, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
FDA root cause category: Software Design Change
What the company is doing
Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
Recall details
- Recall number
- Z-2826-2026
- Status
- Open, Classified
- Initiated by the company
- May 8, 2026
- Posted by the FDA
- Jul 29, 2026
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 3
- Product code
- CHL
- FDA submissions
- K203790
- Distribution
- US Nationwide distribution.
Product description
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples. Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system. iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during, and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls (QC). Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. The GEM Premier 5000 system makes use of potentiometric sensors to measure pCO2, pH, Na+, K+, Cl-, and Ca++. It uses amperometric sensors to measure pO2, glucose, and lactate concentrations. Blood conductivity is the method used to measure hematocrit. CO-Oximetry and total bilirubin measurements involve chemically lysing the whole blood sample and then utilizing a broad spectrum spectrophotometer to evaluate the sample at a variety of wavelengths.
Code information (models, lots, serial numbers)
Lot Code: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: 3 affected BCR wands on the GEM Premier 5000 in Argentina. GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: 3
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
Other devices under the same FDA submission
21 devices cleared under K203790 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/COOX 75 TEST 00055407504Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 600 TEST 00055360010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 600 TEST 00055360011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 150 TEST 00055415010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 300 TEST 00055430011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 450 TEST 00055445011Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/COOX 600 TEST 00055360008Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/COOX 75 TEST 00055407508Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 75 TEST 00055407510Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX/TBILI 75 TEST 00055407511Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99115).
- Z-2821-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment…
- Z-2822-2026 · Open, ClassifiedBrand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality co…
- Z-2823-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
- Z-2824-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
- Z-2825-2026 · Open, ClassifiedBrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Product Description: The GEM Premier 5000 system provides fast, accurate, quantitative…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page