Medline Convenience Kits recallZ-3050-2026 · Medline Industries, LP · initiated Jun 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-3050-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 12, 2026
- Posted by the FDA
- Aug 31, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 58037 kits
- Product code
- OFF
- Distribution
- US Nationwide distribution.
Product description
Medline Convenience Kits: 1) DYKM1414, Catalog Number: DYKM1414; 2) DYKM1414, Catalog Number: DYKM1414; 3) DYNDH1462A, Catalog Number: DYNDH1462A; 4) DYNJ38145F, Catalog Number: DYNJ38145F; 5) DYNJ63199A, Catalog Number: DYNJ63199A
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1) DYKM1414, UDI-DI: 10889942481293(each), 40889942481294(case), Lot Number: 25BBB327; 2) DYKM1414, UDI-DI: 10889942481293(each), 40889942481294(case), Lot Number: 24KBE261; 3) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24ABC301; 4) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24ABT704; 5) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24BBO246; 6) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24BBU663; 7) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24CBJ141; 8) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23JBH263; 9) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23JBO539; 10) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23KBG333; 11) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23KBV434; 12) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23LBM123; 13) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 25ABS139; 14) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 24EBP237; 15) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 24KBE336; 16) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 23JBM582; 17) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 23LBM892; 18) DYNJ63199A, UDI-DI: 10193489311730(each), 40193489311731(case), Lot Number: 24EBG970
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ63199AMedline Industries, Inc. · Central venous catheter tray3 recalls0 reports
- Medline Industries, Inc. DYNJ38145FMedline Industries, Inc. · Central venous catheter tray2 recalls0 reports
- Medline Industries, Inc. DYNDH1462AMedline Industries, Inc. · Central venous catheter tray2 recalls0 reports
- Medline Industries, Inc. DYKM1414Medline Industries, Inc. · Central venous catheter tray2 recalls0 reports
- Medline Industries, Inc. DYNJRA1833AMedline Industries, Inc. · Spinal epidural anesthesia kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99334).
- Z-3048-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, Model Number: DYKMBNDL199A
- Z-3049-2026 · Open, ClassifiedMedline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y, Catalog Number: CDS840396Y; 3) CDS983348AA, Catalog Number: CDS983348AA; 4) CDS983348AB, Cata…
- Z-3051-2026 · Open, ClassifiedMedline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A
- Z-3052-2026 · Open, ClassifiedMedline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437D
- Z-3053-2026 · Open, ClassifiedMedline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375C
- Z-3054-2026 · Open, ClassifiedMedline Convenience Kits: 1) IV START KIT, Catalog Number: DYNDV2520; 2) PERIPHERAL LAB BLOOD CULTURE KIT, Catalog Number: DYNDH1983A; 3) PEDI EMERGENCY TRAY-PURPLE, Catalog Numb…
- Z-3055-2026 · Open, ClassifiedMedline Convenience Kits: 1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295
- Z-3056-2026 · Open, ClassifiedMedline Convenience Kits: 1) IV START KIT LAB DRAW-SH, Model Number: CDS860014P; 2) IV START KIT, Model Number: DYKM1000B; 3) KIT, I V START NSK, Model Number: DYKM2187; 4) KIT,…
- Z-3057-2026 · Open, ClassifiedMedline Convenience Kits: 1) BLOOD CULTURE TRAY PEDIATRIC, Model Number: DYNDH1193; 2) BLOOD CULTURE KIT 2 W/CONNECT, Model Number: DYNDH1487A
- Z-3058-2026 · Open, ClassifiedMedline Convenience Kits: 1) PRE-OP KIT, Catalog Number: DYKS1043F; 2) PRE OP KIT, Catalog Number: DYKS1128C; 3) PRE OP KIT, Catalog Number: DYKS1216I; 4) PRE OP KIT, Catalog Nu…
- Z-3059-2026 · Open, ClassifiedMedline Convenience Kits: 1) LABOR CDS, Model Number: CDS981762B; 2) VAG DELIVERY ADMISSION BAG, Model Number: DYKA1054F
- Z-3060-2026 · Open, ClassifiedMedline Convenience Kits: 1) C-SECTION PREP, Model Number: DYNJ32702B
- Z-3061-2026 · Open, ClassifiedMedline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number: DYKM2751A; 3) CATARACT PREOP KIT, Model Number: DYKS1583A; 4) CATARACT PREOP KIT, Model N…
- Z-3062-2026 · Open, ClassifiedMedline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD BLOCK KIT, Model Number: DYKM1881A; 3) TRIMA KIT, Model Number: DYNDB3077C
- Z-3063-2026 · Open, ClassifiedMedline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUB…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page