Medline Convenience Kits recallZ-3054-2026 · Medline Industries, LP · initiated Jun 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-3054-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 12, 2026
- Posted by the FDA
- Aug 31, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 434,595 kits
- Product code
- LRO
- FDA submissions
- K213481
- Distribution
- US Nationwide distribution.
Product description
Medline Convenience Kits: 1) IV START KIT, Catalog Number: DYNDV2520; 2) PERIPHERAL LAB BLOOD CULTURE KIT, Catalog Number: DYNDH1983A; 3) PEDI EMERGENCY TRAY-PURPLE, Catalog Number: ACC010157C; 4) PEDS EMERGENCY TRAY- GREEN, Catalog Number: ACC010158C; 5) PEDI EMERGENCY TRAY-BLUE, Catalog Number: ACC010159C; 6) PEDS EMERGENCY TRAY-YELLOW, Catalog Number: ACC010160C; 7) PEDS EMERGENCY TRAY-WHITE, Catalog Number: ACC010161C; 8) PEDS EMERGENCY TRAY-ORANGE, Catalog Number: ACC010162C; 9) PEDS EMERGENCY TRAY-RED, Catalog Number: ACC010163C; 10) CVHP DRAWER -IV, Catalog Number: ACC010308A; 11) PEDI EMERGENCY TRAY-BLUE, Catalog Number: ACC010316A; 12) PEDS EMERGENCY TRAY- GREEN, Catalog Number: ACC010318A; 13) PEDS EMERGENCY TRAY-ORANGE, Catalog Number: ACC010319A; 14) DRAWER THREE - IV SET UP, Catalog Number: ACC010361; 15) DRAWER 1, Catalog Number: ACC010374A; 16) DRAWER 3 LAB DRAW & IV START, Catalog Number: ACC010417A; 17) DRAWER #6, Catalog Number: ACC010477B; 18) DRAWER #4, Catalog Number: ACC010479A; 19) ADULT DRAWER 4, Catalog Number: ACC010679; 20) DRAWER 2, Catalog Number: ACC010690; 21) DRAWER 2, Catalog Number: ACC010726; 22) DRAWER 2, Catalog Number: ACC010726A; 23) RRMC CODE CART DRAWER 3, Catalog Number: ACC010770; 24) DRAWER 2, Catalog Number: ACC010875; 25) BOTTOM DRAWER, Catalog Number: ACC010972; 26) CODE CART DRAWER TWO IV SUPPLI, Catalog Number: ACC010977; 27) DRAWER 4 CIRCULATION, Catalog Number: ACC011088; 28) NON-SUTURED SHEATH #113197 KIT, Catalog Number: DYKM1495; 29) SUTURED SHEATH W/SCI #113198, Catalog Number: DYKM1496; 30) LABOR TRIAGE KIT - LF #13232, Catalog Number: DYKM1564; 31) FINGER THORACOSTOMY PCC KIT, Catalog Number: DYKM2676A; 32) UVC KIT, Catalog Number: DYKM2677A; 33) CRIC KIT, Catalog Number: DYKM2678A; 34) SCOPE KIT 3, Catalog Number: DYKM2750; 35) SURGICAL KIT, Catalog Number: DYKM2752; 36) PRE OP KIT, Catalog Number: DYKS1022B; 37) MAC KIT, Catalog Number: DYKS1089B; 38) PRE OP KIT, Catalog Number: DYKS1091A; 39) PRE OP KIT, Catalog Number: DYKS1128B; 40) PRE OP KIT, Catalog Number: DYKS1325; 41) UCLA RAPID RESPONSE KIT, Catalog Number: DYLAB1097; 42) BSSM-WHOLE BLOOD PHLEBOTOMY, Catalog Number: DYNDB3061; 43) WHOLE BLOOD PHLEBOTOMY, Catalog Number: DYNDB3064; 44) WHOLE BLOOD/DOUBLE RED, Catalog Number: DYNDB3078C; 45) ACCESS KIT, Catalog Number: DYNDH1159A; 46) ADLT BLOOD CULTURE COLLECT KIT, Catalog Number: DYNDH1245; 47) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1362; 48) PEDIATRIC BLOOD CULTURE KIT, Catalog Number: DYNDH1363; 49) STANFORD TIME TO POSITIVITY, Catalog Number: DYNDH1591; 50) BLOOD CULTURE PINK PERIPHERAL, Catalog Number: DYNDV1875; 51) BLOOD CULTURE BLUE PERIPHERAL, Catalog Number: DYNDV1876; 52) NICU LINE PACK, Catalog Number: DYNJ63410; 53) STUDENT PRACTICE BAG W LOGO, Catalog Number: EDUC1043; 53)
Code information (models, lots, serial numbers)
Show the full code information (27k characters)
1) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2023122690; 2) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024020790; 3) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024032790; 4) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024052390; 5) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024071190; 6) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024081490; 7) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024101090; 8) DYNDV2520, UDI-DI: 10653160384045(each), 00653160384048(case), Lot Number: 2024111990; 9) DYNDH1983A, UDI-DI: 10653160388920(each), 00653160388923(case), Lot Number: 2025073190; 10) ACC010157C, UDI-DI: 10195327492649(each), 40195327492640(case), Lot Number: 23JDB105; 11) ACC010158C, UDI-DI: 10195327492786(each), 40195327492787(case), Lot Number: 23JDB122; 12) ACC010159C, UDI-DI: 10195327492779(each), 40195327492770(case), Lot Number: 23JDB078; 13) ACC010160C, UDI-DI: 10195327492823(each), 40195327492824(case), Lot Number: 23JDA433; 14) ACC010160C, UDI-DI: 10195327492823(each), 40195327492824(case), Lot Number: 23JDB266; 15) ACC010161C, UDI-DI: 10195327492816(each), 40195327492817(case), Lot Number: 23JDA587; 16) ACC010161C, UDI-DI: 10195327492816(each), 40195327492817(case), Lot Number: 23JDB267; 17) ACC010161C, UDI-DI: 10195327492816(each), 40195327492817(case), Lot Number: 23JDC145; 18) ACC010162C, UDI-DI: 10195327492793(each), 40195327492794(case), Lot Number: 23JDB106; 19) ACC010163C, UDI-DI: 10195327492809(each), 40195327492800(case), Lot Number: 23JDB268; 20) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 25DDA404; 21) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 25EDB078; 22) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 25FDA297; 23) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 25FDB295; 24) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 25GDA555; 25) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 24FDB461; 26) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 24FDB645; 27) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 23KDC062; 28) ACC010308A, UDI-DI: 10889942424597(each), 40889942424598(case), Lot Number: 23LDB762; 29) ACC010316A, UDI-DI: 10889942831746(each), 40889942831747(case), Lot Number: 25CDB196; 30) ACC010316A, UDI-DI: 10889942831746(each), 40889942831747(case), Lot Number: 24DDA506; 31) ACC010316A, UDI-DI: 10889942831746(each), 40889942831747(case), Lot Number: 24EDA984; 32) ACC010316A, UDI-DI: 10889942831746(each), 40889942831747(case), Lot Number: 24LDA823; 33) ACC010318A, UDI-DI: 10889942831760(each), 40889942831761(case), Lot Number: 24EDC133; 34) ACC010318A, UDI-DI: 10889942831760(each), 40889942831761(case), Lot Number: 23KDA881; 35) ACC010319A, UDI-DI: 10889942831777(each), 40889942831778(case), Lot Number: 24DDA262; 36) ACC010319A, UDI-DI: 10889942831777(each), 40889942831778(case), Lot Number: 24DDA507; 37) ACC010319A, UDI-DI: 10889942831777(each), 40889942831778(case), Lot Number: 24EDB665; 38) ACC010319A, UDI-DI: 10889942831777(each), 40889942831778(case), Lot Number: 23JDB228; 39) ACC010361, UDI-DI: 10889942944989(each), 40889942944980(case), Lot Number: 24ADA597; 40) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 25CDB687; 41) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 25EDB273; 42) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 25FDA380; 43) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24ADA326; 44) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24DDC115; 45) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24EDB585; 46) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24FDA451; 47) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24JDA313; 48) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 24LDA107; 49) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 23KDB464; 50) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 23KDC441; 51) ACC010374A, UDI-DI: 10193489893977(each), 40193489893978(case), Lot Number: 23LDB221; 52) ACC010417A, UDI-DI: 10195327230043(each), 40195327230044(case), Lot Number: 25BDA421; 53) ACC010417A, UDI-DI: 10195327230043(each), 40195327230044(case), Lot Number: 25FDB683; 54) ACC010417A, UDI-DI: 10195327230043(each), 40195327230044(case), Lot Number: 24JDA799; 55) ACC010417A, UDI-DI: 10195327230043(each), 40195327230044(case), Lot Number: 24KDB853; 56) ACC010417A, UDI-DI: 10195327230043(each), 40195327230044(case), Lot Number: 23JDB187; 57) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 25CDA651; 58) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 25EDB703; 59) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 25GDB405; 60) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24ADB209; 61) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24BDA093; 62) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24BDA446; 63) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24BDC015; 64) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24CDB041; 65) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24DDB732; 66) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24EDA124; 67) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 24KDA544; 68) ACC010477B, UDI-DI: 10195327131524(each), 40195327131525(case), Lot Number: 23LDA976; 69) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 25DDA606; 70) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 25EDB678; 71) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 25FDA196; 72) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 25HDA602; 73) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 24DDB117; 74) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 24EDA750; 75) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 24FDB890; 76) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 24HDA095; 77) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 24IDA021; 78) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 23KDB046; 79) ACC010479A, UDI-DI: 10195327131517(each), 40195327131518(case), Lot Number: 23KDC470; 80) ACC010479A, UDI-DI: 23LDB379(each), 10195327131517(case), Lot Number: 40195327131518; 81) ACC010679, UDI-DI: 25ADB289(each), 10195327396268(case), Lot Number: 40195327396269; 82) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 25CDA527; 83) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 25EDC013; 84) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 24ADA519; 85) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 24ADC040; 86) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 24JDA657; 87) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 24LDB799; 88) ACC010679, UDI-DI: 10195327396268(each), 40195327396269(case), Lot Number: 23GDB618; 89) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 25GDA245; 90) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 24CDB296; 91) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 24DDB313; 92) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 24EDA327; 93) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 24GDB902; 94) ACC010690, UDI-DI: 10195327433307(each), 40195327433308(case), Lot Number: 24JDA372; 95) ACC010726, UDI-DI: 10195327486402(each), 40195327486403(case), Lot Number: 24ADA800; 96) ACC010726, UDI-DI: 10195327486402(each), 40195327486403(case), Lot Number: 24BDB414; 97) ACC010726, UDI-DI: 10195327486402(each), 40195327486403(case), Lot Number: 23JDA881; 98) ACC010726A, UDI-DI: 10198459043666(each), 40198459043667(case), Lot Number: 24GDA002; 99) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24ADB026; 100) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24ADB445; 101) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24ADB860; 102) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24ADC225; 103) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24BDB043; 104) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24CDA646; 105) ACC010770, UDI-DI: 10195327576998(each), 40195327576999(case), Lot Number: 24CDC111; 106) ACC010875, UDI-DI: 10198459113659(each), 40198459113650(case), Lot Number: 25DDB627; 107) ACC010875, UDI-DI: 10198459113659(each), 40198459113650(case), Lot Number: 24JDA648; 108) ACC010972, UDI-DI: 10198459303289(each), 40198459303280(case), Lot Number: 25EDA655; 109) ACC010977, UDI-DI: 10198459308642(each), 40198459308643(case), Lot Number: 25DDA901; 110) ACC011088, UDI-DI: 10198459589577(each), 40198459589578(case), Lot Number: 26BDA548; 111) DYKM1495, UDI-DI: 10889942639748(each), 40889942639749(case), Lot Number: 23KBE256; 112) DYKM1496, UDI-DI: 10889942636082(each), 40889942636083(case), Lot Number: 24DBH449; 113) DYKM1496, UDI-DI: 10889942636082(each), 40889942636083(case), Lot Number: 24JBB668; 114) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 25DBK369; 115) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 25GBH209; 116) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24LBR594; 117) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24ABG494; 118) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24CBB243; 119) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24CBD763; 120) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24DBI378; 121) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24EBR566; 122) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24GBX258; 123) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 24JBX634; 124) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 23IBF967; 125) DYKM1564, UDI-DI: 10889942754571(each), 40889942754572(case), Lot Number: 23LBI141; 126) DYKM2676A, UDI-DI: 10195327522988(each), 40195327522989(case), Lot Number: 25BBE461; 127) DYKM2676A, UDI-DI: 10195327522988(each), 40195327522989(case), Lot Number: 24CBH132; 128) DYKM2676A, UDI-DI: 10195327522988(each), 40195327522989(case), Lot Number: 24CBT727; 129) DYKM2676A, UDI-DI: 10195327522988(each), 40195327522989(case), Lot Number: 23LBN569; 130) DYKM2677A, UDI-DI: 10195327522995(each), 40195327522996(case), Lot Number: 23LBU445; 131) DYKM2677A, UDI-DI: 10195327522995(each), 40195327522996(case), Lot Number: 24CBP143; 132) DYKM2677A, UDI-DI: 10195327522995(each), 40195327522996(case), Lot Number: 24DBF109; 133) DYKM2677A, UDI-DI: 10195327522995(each), 40195327522996(case), Lot Number: 23JBP672; 134) DYKM2678A, UDI-DI: 10195327523008(each), 40195327523009(case), Lot Number: 25BLA823; 135) DYKM2678A, UDI-DI: 10195327523008(each), 40195327523009(case), Lot Number: 25DLA868; 136) DYKM2678A, UDI-DI: 10195327523008(each), 40195327523009(case), Lot Number: 24CLA476; 137) DYKM2678A, UDI-DI: 10195327523008(each), 40195327523009(case), Lot Number: 24LLA140; 138) DYKM2678A, UDI-DI: 10195327523008(each), 40195327523009(case), Lot Number: 23LLA884; 139) DYKM2750, UDI-DI: 10198459112690(each), 40198459112691(case), Lot Number: 24JBB774; 140) DYKM2752, UDI-DI: 10198459118401(each), 40198459118402(case), Lot Number: 24KBK304; 141) DYKM2752, UDI-DI: 10198459118401(each), 40198459118402(case), Lot Number: 24KBG695; 142) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 25AMF164; 143) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24AMA227; 144) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24AMA782; 145) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24AMH643; 146) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24CMD293; 147) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24DMF883; 148) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24EMJ967; 149) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24GMH029; 150) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24GMK146; 151) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24JMH465; 152) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24KMJ204; 153) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 24LME558; 154) DYKS1022B, UDI-DI: 10193489432848(each), 40193489432849(case), Lot Number: 23JMJ533; 155) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 25CMA215; 156) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24BMB722; 157) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24DMA601; 158) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24EMB032; 159) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24GMH030; 160) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24GMK147; 161) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24HMI067; 162) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 24LMD918; 163) DYKS1089B, UDI-DI: 10193489306286(each), 40193489306287(case), Lot Number: 23KMF447; 164) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 25AMG863; 165) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 25BMC075; 166) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 23LMH153; 167) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 24EMF530; 168) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 24IMC383; 169) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 24LMF652; 170) DYKS1091A, UDI-DI: 10193489574319(each), 40193489574310(case), Lot Number: 23JMG324; 171) DYKS1128B, UDI-DI: 10193489983043(each), 40193489983044(case), Lot Number: 23LBU122; 172) DYKS1128B, UDI-DI: 10193489983043(each), 40193489983044(case), Lot Number: 23LBH912; 173) DYKS1325, UDI-DI: 10193489983036(each), 40193489983037(case), Lot Number: 23JBP437; 174) DYKS1325, UDI-DI: 10193489983036(each), 40193489983037(case), Lot Number: 23KBL343; 175) DYKS1325, UDI-DI: 10193489983036(each), 40193489983037(case), Lot Number: 23LBH895; 176) DYKS1325, UDI-DI: 10193489983036(each), 40193489983037(case), Lot Number: 23LBP617; 177) DYLAB1097, UDI-DI: 10198459346316(each), 40198459346317(case), Lot Number: 25FLA513; 178) DYLAB1097, UDI-DI: 10198459346316(each), 40198459346317(case), Lot Number: 25GLA297; 179) DYLAB1097, UDI-DI: 10198459346316(each), 40198459346317(case), Lot Number: 25HLA636; 180) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 25CBJ821; 181) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24ABY315; 182) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24CBK867; 183) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24DBL293; 184) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24FBA828; 185) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24JBI833; 186) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 24JBX194; 187) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 23KBJ574; 188) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 23KBV200; 189) DYNDB3061, UDI-DI: 10889942143733(each), 40889942143734(case), Lot Number: 23LBI823; 190) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 25ABB074; 191) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 25BBP541; 192) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24ABW521; 193) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24CBA610; 194) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24DBE339; 195) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24FBC085; 196) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24IBR193; 197) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 24JBQ751; 198) DYNDB3064, UDI-DI: 10889942215782(each), 40889942215783(case), Lot Number: 23KBQ427; 199) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 25ABE987; 200) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 25BBI439; 201) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 25CBN467; 202) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24KBJ660; 203) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24LBE165; 204) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24ABE056; 205) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24ABG510; 206) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24CBG613; 207) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24CBT164; 208) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24EBR367; 209) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24FBJ717; 210) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24HBG037; 211) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 24JBB679; 212) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 23IBS706; 213) DYNDB3078C, UDI-DI: 10193489273328(each), 40193489273329(case), Lot Number: 23JBP441; 214) DYNDH1159A, UDI-DI: 10889942881659(each), 40889942881650(case), Lot Number: 24BBR188; 215) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25KME694; 216) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25AMD339; 217) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25AMI448; 218) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25BMB198; 219) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25BMD503; 220) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 25BMH151; 221) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23LMH919; 222) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24AMF220; 223) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24AMJ457; 224) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24BMB874; 225) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24BMI381; 226) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24CMC934; 227) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24CMG933; 228) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24CMK091; 229) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24DMF489; 230) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24EMA066; 231) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24EMH953; 232) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24GMB441; 233) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24GMF372; 234) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24GMG553; 235) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24GMI902; 236) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24IMD769; 237) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24JMI883; 238) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23JMC106; 239) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23JME609; 240) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23JMJ056; 241) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23KMA305; 242) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 23KMJ911; 243) DYNDH1245, UDI-DI: 10889942604722(each), 40889942604723(case), Lot Number: 24CMK091Z; 244) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 25ABC490; 245) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 25ABT234; 246) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 25BBB152; 247) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24LBD414; 248) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24LBN281; 249) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24ABI528; 250) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24ABS533; 251) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24BBJ514; 252) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24CBA543; 253) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24CBK879; 254) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24CBQ864; 255) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24DBE149; 256) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24DBR058; 257) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24FBD953; 258) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24GBO300; 259) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24IBL629; 260) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24JBE586; 261) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24JBQ159; 262) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 24KBE076; 263) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 23JBW105; 264) DYNDH1362, UDI-DI: 10888277797086(each), 40888277797087(case), Lot Number: 23LBE015; 265) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 25DBH954; 266) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 25GBD824; 267) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 23LBQ779; 268) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 24ABY394; 269) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 24CBF326; 270) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 24DBL545; 271) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 24EBI891; 272) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 24FBN485; 273) DYNDH1363, UDI-DI: 10888277795464(each), 40888277795465(case), Lot Number: 23KBA941; 274) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 23LMH149; 275) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 24BMF431; 276) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 24GMF832; 277) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 24IME659; 278) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 24IMI021; 279) DYNDH1591, UDI-DI: 10193489497236(each), 40193489497237(case), Lot Number: 24JMA966; 280) DYNDV1875, UDI-DI: 10888277860209(each), 40888277860200(case), Lot Number: 24AME774; 281) DYNDV1876, UDI-DI: 10888277860032(each), 40888277860033(case), Lot Number: 24CMG621; 282) DYNDV1876, UDI-DI: 10888277860032(each), 40888277860033(case), Lot Number: 24CMH613; 283) DYNDV1876, UDI-DI: 10888277860032(each), 40888277860033(case), Lot Number: 24DMH848; 284) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 25CDA509; 285) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 25CDA643; 286) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 25CDB663; 287) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 24ADB665; 288) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 24IDB069; 289) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 24IDB669; 290) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 23LDA067; 291) DYNJ63410, UDI-DI: 10193489737431(each), 40193489737432(case), Lot Number: 23LDA811; 292) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24ADA018; 293) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24DDB177; 294) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24DDB234; 295) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24GDB033; 296) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24HDA202; 297) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24HDA894; 298) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 24HDB180; 299) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 23LDA532; 300) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 23LDB267; 301) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 23LDB288; 302) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 23LDB289; 303) EDUC1043, UDI-DI: 10195327425531(each), 40195327425532(case), Lot Number: 23LDB290; 304) EDUC1043A, UDI-DI: 10198459334559(each), 40198459334550(case), Lot Number: 25ELA281
Devices named in this recall
47 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ63410Medline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. ACC010690Medline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. DYNDH1245Medline Industries, Inc. · General surgery tray4 recalls0 reports
- Medline Industries, Inc. ACC010770Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYKM1564Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYKS1089BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNDH1591Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010374AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010479AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010679Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. EDUC1043Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010726AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010875Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE DYNDH1363Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- MEDLINE DYNDH1362Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDV1875Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010308AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYKM1496Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. ACC010316AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYKS1022BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKS1091AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKS1325Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKS1128BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010477BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010417AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010726Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010157CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010159CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010158CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010162CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010163CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010161CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010160CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM2676AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM2677AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM2678AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM2750Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM2752Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010977Medline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE DYNDH1983AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNDV1876Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYKM1495Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010318AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010319AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNDH1159AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ACC010361Medline Industries, Inc. · General surgery tray1 recall0 reports
Other devices under the same FDA submission
319 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC7675Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99334).
- Z-3048-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, Model Number: DYKMBNDL199A
- Z-3049-2026 · Open, ClassifiedMedline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y, Catalog Number: CDS840396Y; 3) CDS983348AA, Catalog Number: CDS983348AA; 4) CDS983348AB, Cata…
- Z-3050-2026 · Open, ClassifiedMedline Convenience Kits: 1) DYKM1414, Catalog Number: DYKM1414; 2) DYKM1414, Catalog Number: DYKM1414; 3) DYNDH1462A, Catalog Number: DYNDH1462A; 4) DYNJ38145F, Catalog Number:…
- Z-3051-2026 · Open, ClassifiedMedline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A
- Z-3052-2026 · Open, ClassifiedMedline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437D
- Z-3053-2026 · Open, ClassifiedMedline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375C
- Z-3055-2026 · Open, ClassifiedMedline Convenience Kits: 1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295
- Z-3056-2026 · Open, ClassifiedMedline Convenience Kits: 1) IV START KIT LAB DRAW-SH, Model Number: CDS860014P; 2) IV START KIT, Model Number: DYKM1000B; 3) KIT, I V START NSK, Model Number: DYKM2187; 4) KIT,…
- Z-3057-2026 · Open, ClassifiedMedline Convenience Kits: 1) BLOOD CULTURE TRAY PEDIATRIC, Model Number: DYNDH1193; 2) BLOOD CULTURE KIT 2 W/CONNECT, Model Number: DYNDH1487A
- Z-3058-2026 · Open, ClassifiedMedline Convenience Kits: 1) PRE-OP KIT, Catalog Number: DYKS1043F; 2) PRE OP KIT, Catalog Number: DYKS1128C; 3) PRE OP KIT, Catalog Number: DYKS1216I; 4) PRE OP KIT, Catalog Nu…
- Z-3059-2026 · Open, ClassifiedMedline Convenience Kits: 1) LABOR CDS, Model Number: CDS981762B; 2) VAG DELIVERY ADMISSION BAG, Model Number: DYKA1054F
- Z-3060-2026 · Open, ClassifiedMedline Convenience Kits: 1) C-SECTION PREP, Model Number: DYNJ32702B
- Z-3061-2026 · Open, ClassifiedMedline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number: DYKM2751A; 3) CATARACT PREOP KIT, Model Number: DYKS1583A; 4) CATARACT PREOP KIT, Model N…
- Z-3062-2026 · Open, ClassifiedMedline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD BLOCK KIT, Model Number: DYKM1881A; 3) TRIMA KIT, Model Number: DYNDB3077C
- Z-3063-2026 · Open, ClassifiedMedline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUB…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page