Medline Convenience Kits recallZ-3063-2026 · Medline Industries, LP · initiated Jun 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-3063-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 12, 2026
- Posted by the FDA
- Aug 31, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 996692 kits
- Product code
- OIB
- Distribution
- US Nationwide distribution.
Product description
Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUBLE RED CELL PHLEBOTOMY KIT, Catalog Number: DYNDB3006C; 5) PLATELET PHLEBOTOMY KIT, Catalog Number: DYNDB3007C; 6) WHOLE BLOOD KIT, Catalog Number: DYNDB3041D; 7) PHLEBOTOMY, Catalog Number: DYNDB3044D; 8) PHLEBOTOMY, Catalog Number: DYNDB3044F; 9) DONATION - WB, Catalog Number: DYNDB3086D; 10) DONATION - ALYX/TRIMA W/O TOUR, Catalog Number: DYNDB3088B; 11) THERAPEUTIC NFT, Catalog Number: DYNDB3100; 12) BLOOD CULTURE TRAY ADULT, Catalog Number: DYNDH1194A; 13) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1444D; 14) PEDS BLOOD CULTURE KIT, Catalog Number: DYNDH1445C; 15) BLOOD CULTURE KIT, Catalog Number: DYNDH1462B; 16) BLOOD CULTURE ADULT, Catalog Number: DYNDH1619A; 17) BLOOD CULTURE PEDI, Catalog Number: DYNDH1620A; 18) BLOOD CULTURE KIT, Catalog Number: DYNDH1622C; 19) BLOOD CULTURE KIT, Catalog Number: DYNDH1622D; 20) BLOOD CULTURE COLLECTION KIT, Catalog Number: DYNDH1626B; 21) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1694A; 22) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1695; 23) BLOOD CULTURE KIT, Catalog Number: DYNDH1708A; 24) BLOOD CULTURE COLLECTION KIT, Catalog Number: DYNDH1714A; 25) BLOOD CULTURE KIT, Catalog Number: DYNDH1715B; 26) BLOOD CULTURE KIT, Catalog Number: DYNDH1723; 27) NS BLOOD CULTURE 23G KIT, Catalog Number: DYNDH1740A; 28) NS BLOOD CULTURE 23G KIT, Catalog Number: DYNDH1740B; 29) NS BLOOD CULTURE 21G KIT, Catalog Number: DYNDH1741A; 30) NS BLOOD CULTURE 21G KIT, Catalog Number: DYNDH1741B; 31) PEDS BLOOD CULTURE KIT, Catalog Number: DYNDH1771A; 32) BLOOD CULTURE KIT, Catalog Number: DYNDH1805B; 33) BLOOD CULTURE KIT, Catalog Number: DYNDH1817A; 34) BLOOD CULTR KT DIVRSN DEVICE, Catalog Number: DYNDH1866; 35) BLOOD CULTURE KIT, Catalog Number: DYNDH1867; 36) BLOOD CULTURE KIT, Catalog Number: DYNDH1884; 37) PHLEBOTOMY BLOOD CULTURE KIT, Catalog Number: DYNDH1908; 38) BLOOD CULTURE KIT, Catalog Number: DYNDH1912; 39) BLOOD CULTURE KIT, Catalog Number: DYNDH1940; 40) BLOOD CULTURE KIT W/O BOTTLES, Catalog Number: DYNDH1985; 41) BLOOD CULTURE KIT, Catalog Number: DYNDH1988; 42) BLOOD CULTURE, Catalog Number: DYNDH2002; 43) BLOOD CULTURE W/NO IV, Catalog Number: DYNDH2016; 44) BLOOD CULTURE W/IV, Catalog Number: DYNDH2017; 45) BLOOD CULTURE KIT, Catalog Number: DYNDH2023; 46) BLOOD CULTURE KIT, Catalog Number: DYNDH2028; 47) BLOOD CULTURE KIT, Catalog Number: DYNDH2030; 48) BLOOD CULTURE KIT, Catalog Number: DYNDH2048; 49) BLOOD CULTURE KIT, Catalog Number: DYNDH2059; 50) BLOOD CULTURE KIT 450, Catalog Number: DYNDH2074; 51) BLOOD CULTURE-NB, Catalog Number: DYNDH2089
Code information (models, lots, serial numbers)
Show the full code information (38k characters)
1) DYLAB1081, UDI-DI: 10198459199608(each), 40198459199609(case), Lot Number: 25CMC514; 2) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBG614; 3) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBT912; 4) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23HBH298; 5) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23JBN521; 6) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23KBT270; 7) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBD241; 8) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBP477; 9) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25CBI161; 10) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24LBM458; 11) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBL961; 12) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBS200; 13) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24IBO183; 14) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABB637; 15) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABY532; 16) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24CBH932; 17) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 23JBN434; 18) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 23LBH909; 19) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 24CBL044; 20) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23JBC957; 21) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23KBT835; 22) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23LBI822; 23) DYNDB3041D, UDI-DI: 10195327170738(each), 40195327170739(case), Lot Number: 24KBS893; 24) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 24ABH596; 25) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 23KBB360; 26) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 23LBI693; 27) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 25BBA987; 28) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24LBE163; 29) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24DBO764; 30) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24DBO786; 31) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24EBD140; 32) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24GBR601; 33) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24IBR933; 34) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24LBR503; 35) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 23LBM024; 36) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24BBM148; 37) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24CBN122; 38) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24FBC084; 39) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24JBV595; 40) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 23JBL111; 41) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 25ABB075; 42) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24BBO180; 43) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24CBN068; 44) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24DBW006; 45) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24IBC763; 46) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24JBA762; 47) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24JBX189; 48) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 23JBY503; 49) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 23LBM098; 50) DYNDB3100, UDI-DI: 10198459028915(each), 40198459028916(case), Lot Number: 24GBF854; 51) DYNDH1194A, UDI-DI: 10195327432300(each), 40195327432301(case), Lot Number: 23KBQ229; 52) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25ABW691; 53) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25DBO475; 54) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25EBD598; 55) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25FBQ279; 56) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23LBU409; 57) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24CBQ394; 58) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24EBI903; 59) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24GBH145; 60) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23JBI911; 61) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23JBQ284; 62) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24LBU139; 63) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 25BBE915; 64) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 25EBD601; 65) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24ABT708; 66) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24CBH112; 67) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24DBS252; 68) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24FBH711; 69) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23KBH644; 70) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23KBV432; 71) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23LBJ808; 72) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25ABK267; 73) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25BBB147; 74) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24CBU138; 75) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24DBJ003; 76) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24DBP521; 77) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24EBA062; 78) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24EBU164; 79) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24FBD281; 80) DYNDH1462B, UDI-DI: 24FBK195(each), 10195327662462(case), Lot Number: 40195327662463; 81) DYNDH1462B, UDI-DI: 24FBS941(each), 10195327662462(case), Lot Number: 40195327662463; 82) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24GBH144; 83) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24GBX678; 84) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBG621; 85) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBJ952; 86) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBT054; 87) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24JBA557; 88) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25ABK267Z; 89) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 25ABC488; 90) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 25ABO279; 91) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24LBD413; 92) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24ABL303; 93) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24ABY205; 94) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24CBJ142; 95) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24CBT682; 96) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24EBI894; 97) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24FBS943; 98) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24GBT335; 99) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24IBL862; 100) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24IBV950; 101) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24JBB865; 102) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24JBI309; 103) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24KBE071; 104) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23IBL873; 105) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23JBH261; 106) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23KBQ211; 107) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24LBU144; 108) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24LBI541; 109) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24DBO997; 110) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24GBC034; 111) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24KBH535; 112) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 23KBV431; 113) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 25EBD604; 114) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24ABK078; 115) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24BBB587; 116) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24BBV398; 117) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24CBJ143; 118) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24CBU661; 119) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24EBU004; 120) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24FBK924; 121) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24GBH140; 122) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24HBT637; 123) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24JBB866; 124) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24JBF575; 125) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23JBQ525; 126) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23KBW274; 127) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23LBL909; 128) DYNDH1622D, UDI-DI: 10198459234965(each), 40198459234966(case), Lot Number: 25GBA905; 129) DYNDH1622D, UDI-DI: 10198459234965(each), 40198459234966(case), Lot Number: 25HBN203; 130) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24LBI550; 131) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24ABI955; 132) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24BBB590; 133) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24CBJ146; 134) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24DBJ006; 135) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24EBA605; 136) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24GBH142; 137) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24IBL883; 138) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24JBI311; 139) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24JBW009; 140) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23IBI210; 141) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23KBH642; 142) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23KBL369; 143) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23LBD097; 144) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24LBU146; 145) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 25ABK273; 146) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 25BBB149; 147) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24LBI542; 148) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24ABF793; 149) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24ABT705; 150) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24BBI818; 151) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24BBU666; 152) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24DBE739; 153) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24GBH139; 154) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBB026; 155) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBE383; 156) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBL900; 157) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBV952; 158) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24JBF571; 159) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24JBW011; 160) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24KBH537; 161) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23JBC264; 162) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23JBH262; 163) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23KBI725; 164) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23KBQ208; 165) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23LBD095; 166) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23LBF186; 167) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25AMI450; 168) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25BMD501; 169) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25CME092; 170) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24GMH831; 171) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24IMD773; 172) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24IMI239; 173) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24JMC905; 174) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24JMG344; 175) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24KMH750; 176) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24LMC717; 177) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24BBB589; 178) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24BBR118; 179) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24CBR792; 180) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24EBA607; 181) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 23JBP690; 182) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 23LBG924; 183) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 25ABT228; 184) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 25BBM180; 185) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24KBN935; 186) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24LBI551; 187) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24ABM740; 188) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24DBG375; 189) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24DBH805; 190) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24EBA011; 191) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24GBH136; 192) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23IBG944; 193) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23KBL592; 194) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23LBE163; 195) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 25ABT229; 196) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 25BBI372; 197) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24KBN938; 198) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24LBI548; 199) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24ABG625; 200) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24BBB584; 201) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24DBA250; 202) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24FBB687; 203) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24GBL430; 204) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24IBL892; 205) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBG988; 206) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBI307; 207) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBW010; 208) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23KBC138; 209) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23KBT360; 210) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23LBF190; 211) DYNDH1715B, UDI-DI: 10195327407094(each), 40195327407095(case), Lot Number: 24IMF148; 212) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 25ABK268; 213) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 25BBG725; 214) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24LBN286; 215) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24ABX116; 216) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24CBP727; 217) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24CBR793; 218) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24IBL881; 219) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24JBN261; 220) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24JBW018; 221) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23JBO541; 222) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23KBI724; 223) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23LBF189; 224) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24ABM651; 225) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24ABW198; 226) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24DBQ460; 227) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBL839; 228) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBS096; 229) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBW106; 230) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBH643; 231) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBM453; 232) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBV433; 233) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23LBK910; 234) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23LBM124; 235) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25ABC762; 236) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25ABF383; 237) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25BBC863; 238) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25JBA059; 239) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24LBC961; 240) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24EBA610; 241) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24GBT336; 242) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24IBK137; 243) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24JBW017; 244) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24ABM652; 245) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24BBH982; 246) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24CBQ396; 247) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24CBR795; 248) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23IBS931; 249) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23JBD149; 250) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23KBJ502; 251) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23LBF192; 252) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23LBM125; 253) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABC763; 254) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABK266; 255) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABT230; 256) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24KBN936; 257) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24LBI543; 258) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24IBK132; 259) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24JBS020; 260) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24KBH536; 261) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24LBU145; 262) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25ABK265; 263) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25ABO639; 264) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25BBG724; 265) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25CBG204; 266) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBG779; 267) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBQ113; 268) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBS994; 269) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25AMD341; 270) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25AMG536; 271) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25BMB721; 272) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25BMH602; 273) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25CME260; 274) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24DMK121; 275) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24EMK057; 276) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24FMH384; 277) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24GMB446; 278) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24GMG550; 279) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24IMD770; 280) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24IMI240; 281) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMA923; 282) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMC906; 283) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMG338; 284) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24LMA311; 285) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24LMG806; 286) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24ABX115; 287) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24CBI261; 288) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 23LBD096; 289) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 23LBF187; 290) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24LBD415; 291) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23LBF191; 292) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24CBQ393; 293) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24EBA611; 294) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24GBT338; 295) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24IBK136; 296) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24JBW014; 297) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23JBQ200; 298) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23LBE911; 299) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25ABC487; 300) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25ABO282; 301) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25BBC864; 302) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23LBU410; 303) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24ABL891; 304) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24BBI806; 305) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24CBB996; 306) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24CBR796; 307) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24DBP522; 308) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBA612; 309) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBI902; 310) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBT285; 311) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24FBH707; 312) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24FBN853; 313) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24GBB459; 314) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24GBT339; 315) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24HBB568; 316) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24JBB782; 317) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24JBI308; 318) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23JBP541; 319) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23LBB120; 320) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 23LBU408; 321) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24BBR119; 322) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24CBJ147; 323) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24DBJ005; 324) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24EBA588; 325) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 23JBM365; 326) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25AMD344; 327) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25AMG374; 328) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMB195; 329) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMD502; 330) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMH148; 331) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24FMC615; 332) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24FMH383; 333) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24GMG552; 334) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24GMI903; 335) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24HMI304; 336) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMD766; 337) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMG695; 338) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMI237; 339) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24JMA921; 340) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24JMG345; 341) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24KMD105; 342) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24LMC718; 343) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24LMG808; 344) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24ABF789; 345) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24ABT706; 346) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24BBU664; 347) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24DBO706; 348) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24KBU495; 349) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABF388; 350) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABI775; 351) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABO844; 352) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24CBJ647; 353) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24GBL009; 354) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 23JBL322; 355) DYNDH1912, UDI-DI: 10195327476939(each), 40195327476930(case), Lot Number: 23JBK266; 356) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25BMD500; 357) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25BMH149; 358) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25CMH786; 359) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24DMF413; 360) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24EMC278; 361) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24EMK058; 362) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24FME926; 363) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24FMF294; 364) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMB440; 365) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMF833; 366) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMG551; 367) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMI904; 368) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24HMC740; 369) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24HMI305; 370) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMD767; 371) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMG698; 372) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMI238; 373) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMA922; 374) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMG343; 375) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMI885; 376) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMA312; 377) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMC719; 378) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMG811; 379) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24ABF792; 380) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24BBB873; 381) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24BBH981; 382) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24CBI260; 383) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBC202; 384) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBI726; 385) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBQ207; 386) DYNDH1985, UDI-DI: 10195327623043(each), 40195327623044(case), Lot Number: 24DBP083; 387) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24LBI546; 388) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24CBQ392; 389) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24IBK134; 390) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24JBW016; 391) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 25ABO281; 392) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 25ABT231; 393) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24LBD416; 394) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24EBU007; 395) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24IBK135; 396) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24JBS021; 397) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24KBE074; 398) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 26AMC986; 399) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24FBS944; 400) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24GBH203; 401) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24IBR097; 402) DYNDH2017, UDI-DI: 10198459022159(each), 40198459022150(case), Lot Number: 24FBS945; 403) DYNDH2017, UDI-DI: 10198459022159(each), 40198459022150(case), Lot Number: 24GBH204; 404) DYNDH2023, UDI-DI: 10198459050954(each), 40198459050955(case), Lot Number: 24GBQ903; 405) DYNDH2023, UDI-DI: 10198459050954(each), 40198459050955(case), Lot Number: 24IBB027; 406) DYNDH2028, UDI-DI: 10198459035715(each), 40198459035716(case), Lot Number: 24HBI441; 407) DYNDH2030, UDI-DI: 10198459043703(each), 40198459043704(case), Lot Number: 24GBG703; 408) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25ABP888; 409) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25DBP574; 410) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25DBV279; 411) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25EBB578; 412) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25EBI293; 413) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 24IBV953; 414) DYNDH2059, UDI-DI: 10198459238406(each), 40198459238407(case), Lot Number: 25BBG684; 415) DYNDH2074, UDI-DI: 10198459205828(each), 40198459205829(case), Lot Number: 25BBB182; 416) DYNDH2089, UDI-DI: 10198459244070(each), 40198459244071(case), Lot Number: 25DBM846
Devices named in this recall
50 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNDH1622CMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)4 recalls0 reports
- Medline Industries, Inc. DYNDH1723Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)3 recalls0 reports
- Medline Industries, Inc. DYNDH1694AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)3 recalls0 reports
- Medline Industries, Inc. DYNDH1714AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)3 recalls0 reports
- Medline Industries, Inc. DYNDH1866Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)3 recalls0 reports
- Medline Industries, Inc. DYNDH1462BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)3 recalls0 reports
- Medline Industries, Inc. DYNDB3086DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls2 reports
- Medline Industries, Inc. DYNDB3005DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1695Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1619AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1620AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1626BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1715BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1194AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1908Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1708AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1444DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1445CMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1940Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1988Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH2002Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1740BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1741BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH2017Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDB3100Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH2028Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1771AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH2048Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYLAB1081Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH1622DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDH2089Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)2 recalls0 reports
- Medline Industries, Inc. DYNDB3044DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3006CMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3007CMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3088BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3041DMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1741AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1740AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1867Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1884Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1817AMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1912Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1805BMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH1985Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3044FMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH2016Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDB3005FMedline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH2030Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH2023Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
- Medline Industries, Inc. DYNDH2059Medline Industries, Inc. · Blood and urine collection kit (excludes HIV testing)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99334).
- Z-3048-2026 · Open, ClassifiedMedline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, Model Number: DYKMBNDL199A
- Z-3049-2026 · Open, ClassifiedMedline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y, Catalog Number: CDS840396Y; 3) CDS983348AA, Catalog Number: CDS983348AA; 4) CDS983348AB, Cata…
- Z-3050-2026 · Open, ClassifiedMedline Convenience Kits: 1) DYKM1414, Catalog Number: DYKM1414; 2) DYKM1414, Catalog Number: DYKM1414; 3) DYNDH1462A, Catalog Number: DYNDH1462A; 4) DYNJ38145F, Catalog Number:…
- Z-3051-2026 · Open, ClassifiedMedline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A
- Z-3052-2026 · Open, ClassifiedMedline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437D
- Z-3053-2026 · Open, ClassifiedMedline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375C
- Z-3054-2026 · Open, ClassifiedMedline Convenience Kits: 1) IV START KIT, Catalog Number: DYNDV2520; 2) PERIPHERAL LAB BLOOD CULTURE KIT, Catalog Number: DYNDH1983A; 3) PEDI EMERGENCY TRAY-PURPLE, Catalog Numb…
- Z-3055-2026 · Open, ClassifiedMedline Convenience Kits: 1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295
- Z-3056-2026 · Open, ClassifiedMedline Convenience Kits: 1) IV START KIT LAB DRAW-SH, Model Number: CDS860014P; 2) IV START KIT, Model Number: DYKM1000B; 3) KIT, I V START NSK, Model Number: DYKM2187; 4) KIT,…
- Z-3057-2026 · Open, ClassifiedMedline Convenience Kits: 1) BLOOD CULTURE TRAY PEDIATRIC, Model Number: DYNDH1193; 2) BLOOD CULTURE KIT 2 W/CONNECT, Model Number: DYNDH1487A
- Z-3058-2026 · Open, ClassifiedMedline Convenience Kits: 1) PRE-OP KIT, Catalog Number: DYKS1043F; 2) PRE OP KIT, Catalog Number: DYKS1128C; 3) PRE OP KIT, Catalog Number: DYKS1216I; 4) PRE OP KIT, Catalog Nu…
- Z-3059-2026 · Open, ClassifiedMedline Convenience Kits: 1) LABOR CDS, Model Number: CDS981762B; 2) VAG DELIVERY ADMISSION BAG, Model Number: DYKA1054F
- Z-3060-2026 · Open, ClassifiedMedline Convenience Kits: 1) C-SECTION PREP, Model Number: DYNJ32702B
- Z-3061-2026 · Open, ClassifiedMedline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number: DYKM2751A; 3) CATARACT PREOP KIT, Model Number: DYKS1583A; 4) CATARACT PREOP KIT, Model N…
- Z-3062-2026 · Open, ClassifiedMedline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD BLOCK KIT, Model Number: DYKM1881A; 3) TRIMA KIT, Model Number: DYNDB3077C
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page