MAC recallZ-2876-2026 · GE Medical Systems Information Technologies Inc · initiated Jun 29, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems Information Technologies Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
FDA root cause category: Software design
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 6/29/2026 via US mail. The notice explained the issue, potential risk, and requested the following: 'Actions to be taken by Customer/User You can continue to use your device. Pending corrections from GE HealthCare, after pressing the Order Refresh icon do not rely on automatic closure of the Preview and or Review screen when: " patient has been disconnected, and " at least 2 minutes of system inactivity If this occurs, do the following to clear all patient information data: " On the Preview screen, select Reject > Start New Patient " On the Review screen, select Done > Start New Patient NOTE: Be certain Start New Patient is selected to clear all patient information data. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 30123 Digital Response Form or print, fill out manually, scan, and email to [email protected]." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-2876-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 29, 2026
- Posted by the FDA
- Aug 4, 2026
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Quantity
- 163 units
- Product code
- DPS
- FDA submissions
- K173830
- Distribution
- Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.
Product description
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU219400034SA, VU219400045SA, VU219400055SA, VU220270027SA, VU220510048SA, VU220510078SA, VU221100161SA, VU221100181SA, VU221150116SA, VU222120212SA, VU222120220SA, VU222120239SA, VU222120243SA, VU222120249SA, VU222120250SA, VU222120253SA, VU222120258SA, VU222120260SA, VU222120261SA, VU222120266SA, VU222120269SA, VU222140004SA, VU222200296SA, VU222520114SA, VU223120014SA, VU223120041SA, VU223420119SA, VU224050158SA, VU224070091SA, VU224200028SA, VU224420159SA, VU224520140SA, VU225170260SA, VU225420050SA, VU225520037SA, VU226040025SA, VU419320042SA, VU421490135SA, VU421490151SA, VU421490153SA, VU421490155SA, VU421490159SA, VU422460150SA, VU422460194SA, VU422460203SA, VU422460303SA, VU422480164SA, VU422500239SA, VU422500240SA, VU422500241SA, VU422500242SA, VU422500244SA, VU422500245SA, VU422500246SA, VU422500247SA, VU422500248SA, VU422500249SA, VU422500250SA, VU422500251SA, VU422500254SA, VU422500260SA, VU422500261SA, VU422500264SA, VU422500265SA, VU422500266SA, VU422500267SA, VU422500269SA, VU422510038SA, VU422510049SA, VU422510060SA, VU422510061SA, VU423050080SA, VU423050178SA, VU423220064SA, VU423220225SA, VU425130020SA, VU425130022SA, VU425130024SA, VU425130084SA, VU425130085SA, VU421490060SA, VU421490063SA, VU421490065SA, VU421490068SA, VU421490069SA, VU421490097SA, VU421490122SA, VU422460206SA, VU422460209SA, VU422460215SA, VU422460218SA, VU422460219SA, VU422460302SA, VU422460304SA, VU422460269SA, VU422460273SA, VU422460274SA, VU422460278SA, VU422460282SA, VU422460284SA, VU422480071SA, VU422460306SA, VU422460285SA, VU422480070SA, VU422460287SA, VU422460301SA, VU422480066SA, VU422480078SA, VU422480083SA, VU422480084SA, VU422480085SA, VU422480086SA, VU422480087SA, VU422480088SA, VU422480099SA, VU422480110SA, VU422480111SA, VU422480120SA, VU422480121SA, VU422480125SA, VU422480127SA, VU422480128SA, VU422480129SA, VU425130087SA, VU224200105SA, VU222120088SA, VU223130071SA, VU223220135SA, VU223220174SA, VU219490019SA, VU226050047SA, VU225460222SA, VU225460237SA, VU225470012SA, VU225470018SA, VU225470022SA, VU225470023SA, VU225470024SA, VU225470030SA, VU225470034SA, VU225470035SA, VU225470038SA, VU221050014SA, VU223310039SA, VU223450002SA, VU223310012SA, VU222190155SA, VU225060039SA, VU225060040SA, VU225060046SA, VU225060038SA, VU226154224SA, VU225500096SA, VU225500106SA, VU225500109SA, VU225500019SA, VU225500031SA, VU225500076SA.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAC VUGe Medical Systems Information Technologies, Inc. · Electrocardiograph1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K173830 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page