MAC VURecalls & reported problems
Ge Medical Systems Information Technologies, Inc. · MAC VU360 V3.00 SW UPGRADE · Class 2
All models of MAC (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K173830); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jun 29, 2026Class IIOpen, Classified
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Z-2876-2026 · Root cause: Software design · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K173830 on Sep 18, 2018 (decision: Substantially Equivalent). Submitted by Ge Medical Systems Information Technologies, Inc. as “MAC VU360 Resting ECG Analysis System”. The FDA review took 274 days. FDA record ↗
The FDA regulates product code DPS (electrocardiograph) under 21 CFR 870.2340 as a Class II device. 21 CFR 870.2340 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K173830. One submission often covers many models, so check the product and lot details in each record.
- Z-0231-2026 · Aug 22, 2025 · Open, ClassifiedMAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
- Z-0543-2021 · Oct 27, 2020 · TerminatedMAC VU360, Electrocardiograph
- Z-2060-2019 · Jun 28, 2019 · TerminatedMAC VU360, Model Number 2030360-001, Electrocardiograph
Common questions about the MAC VU
Has the MAC VU been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2876-2026 (Class II), began on Jun 29, 2026: Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the MAC VU FDA approved?
It was cleared through 510(k) K173830 on Sep 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the MAC VU MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the MAC VU, and is it still sold?
The FDA's device database lists Ge Medical Systems Information Technologies, Inc. as the company responsible for it (brand MAC, model VU). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Electrocardiograph devices
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- Cardiac Workstation Cardiac Workstation 7000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls16 reports
- Cardiac Workstation Cardiac Workstation 5000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls1 report
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
- XScribe Cardiopulmonary Kit 41000-030-62Mortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Q-Stress QS6-AXTDSMortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Welch Allyn, Inc. 901007Welch Allyn, Inc. · Electrocardiograph2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page