BD Intraosseous Vascular Access System recallZ-2872-2026 · Bard Access Systems, Inc. · initiated Jul 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Access Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy

FDA root cause category: Component change control

What the company is doing

On 7/30/2026, recall notices were mailed/emailed: If rotation and locking occurs, the clinician may need to obtain a new IO needle or other form of vascular access device, such as a Peripheral IV catheter. Alternate vascular access approaches, including additional peripheral IV attempts, or central venous access may continue to be pursued based on clinical judgment/protocols. Affected product should not be used to establish IO access. For patients in whom potentially affected intraosseous needles are currently in use/or have been used and are/were functioning properly, no additional steps need to be taken. 1. Destroy all affected needle set kits, or quarantined until disposal. 2. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 3. Complete and return the attached Customer Response Form via email to [email protected] Distributors were also asked to do the following: - A customer letter template is provided to create your own letter to be sent to all customers. - Provide product replacement for all product destroyed by your customer. Questions contact firm at 1-844-823-5433 or [email protected] Firm has acquired new molds with improved dimensional controls. Instructions For Use will be updated with the following to optimize needle / obturator(stylet) / drill handling: Do not pull back on the drill and needle, release the trigger prematurely, or rock/bend the needle during insertion. These actions may result in the needle and drive adapter hubs to rotate and lock together resulting in a bent needle, difficult stylet removal, need for a new device, and/or delay in therapy. Warning: Do not rotate or try to unscrew the drive adapter hub while removing the stylet. These actions may cause the needle and drive adapter hubs to rotate and lock together, resulting in difficult stylet removal, need for a new device, and/or delay in therapy.

Recall details

Recall number
Z-2872-2026
Status
Open, Classified
Initiated by the company
Jul 30, 2026
Posted by the FDA
Aug 14, 2026
Recalling firm
Bard Access Systems, Inc.
Quantity
80,200
Product code
FMI
FDA submissions
K191976, K203193
Distribution
US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG

Product description

BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

REF/UDI-DI/Lot(Expiration): D015151NK/00801741163586/BSLC9275(2026-08-03), BSLC9563(2026-08-13), BSLC9567(2026-08-25), BSLC10103(2026-09-27), BSLC10516(2026-11-01), BSLC10526(2026-11-16), BSLC10872(2026-12-09), BSLC11216(2028-07-02), BSLC11456(2028-07-23); D015251NK/00801741163678/BSLC9467(2026-08-10), BSLC9575(2026-09-20), BSLC10105(2026-10-04), BSLC10522(2026-11-07), BSLC10868(2026-12-02), BSLC11212(2028-06-18), BSLC11454(2028-07-17), BSLC11463(2028-08-07), BSLC11884(2028-09-15), BSLC11890(2028-09-18), BSLC12082(2028-10-08), BSLC12115(2028-10-20), BSLC12123(2028-10-22); D015351NK/00801741163593/BSLC11458(2028-07-24); 00801741163678/BSLC9277(2026-08-04), BSLC9485(2026-08-12), BSLC9569(2026-08-26), BSLC10099(2026-09-07), BSLC10514(2026-10-16), BSLC10518(2026-10-21), BSLC10524(2026-11-08), BSLC10870(2026-12-03), BSLC10874(2028-06-12), BSLC11218(2028-07-07); D015451NK/00801741163609/BSLC9279(2026-08-06), BSLC9565(2026-08-17), BSLC9571(2026-09-07), BSLC10101(2026-09-17), BSLC10456(2026-10-14), BSLC10520(2026-10-22), BSLC10866(2026-11-21), BSLC11214(2028-06-23), BSLC11450(2028-07-09), BSLC11460(2028-07-31), BSLC11873(2028-08-13), BSLC11888(2028-09-25), BSLC12007(2028-10-02), BSLC12105(2028-10-15); D015551NK/00801741163616/BSLC9465(2026-08-07), BSLC9573(2026-09-26), BSLC10107(2026-10-02), BSLC10528(2026-11-20), BSLC11452(2028-07-10)