Skip to content

BD Intraosseous Vascular Access System D015351NKRecalls & reported problems

Bard Access Systems, Inc. · BD Intraosseous Needle Set Kit · Class 2

All models of BD Intraosseous Vascular Access System (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741163593), received between Jun 27, 2025 and Feb 23, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IOpen, Classified

    After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy

    Z-2872-2026 · Root cause: Component change control · Recall details · FDA record ↗

  2. Jun 20, 2022Class IOpen, Classified

    BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage as the stylet is removed; 3) Metal discs in the powered driver may stick rendering the driver unusable.

    Z-1397-2022 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove1
  • Material Deformation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2026Material Deformation; Material Twisted/Bent

    IT WAS REPORTED AN IO NEEDLE BENT DURING A RECENT TRAINING EVENT IN A CADAVER LAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24423211 ↗
  • MalfunctionJun 27, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT 35MM NEEDLE NOT SEPARATING AFTER INSERTION. THIS WAS A TYPICAL INSERTION. NOTHING WAS ABNORMAL DURING THE INSERTION PROCESS. THE DRILL WORKED PERFECTLY, AND INSERTION WAS ON THE PATIENT'S TIBIA. THERE WAS NO PATIENT HARM. THE NEEDLES WERE USED IN A CADAVER LAB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22347822 ↗

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page