Skip to content

BD Intraosseous Vascular Access System D015151NKRecalls & reported problems

Bard Access Systems, Inc. · BD Intraosseous Needle Set Kit · Class 2

All models of BD Intraosseous Vascular Access System (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
1
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741163586), received between Mar 11, 2022 and Apr 16, 2026 — classified as 1 death, 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (4 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IOpen, Classified

    After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy

    Z-2872-2026 · Root cause: Component change control · Recall details · FDA record ↗

  2. Jun 20, 2022Class IOpen, Classified

    BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage as the stylet is removed; 3) Metal discs in the powered driver may stick rendering the driver unusable.

    Z-1397-2022 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove4
  • Defective Component3
  • Product Quality Problem2
  • Activation Problem1
  • Failure to Infuse1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Unintended Loss of Vascular Access1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Product Quality Problem; Difficult to Remove

    IT WAS REPORTED BY CUSTOMER THAT DIFFICULTY SEPARATING NEEDLE AND OBTURATOR. NO FURTHER DETAILS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH 3 DEVICES. THIS REPORT ADDRESSES ALL 3 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903677 ↗
  • MalfunctionApr 16, 2026Product Quality Problem; Difficult to Remove

    IT WAS REPORTED BY THE CUSTOMER THAT DIFFICULTY SEPARATING NEEDLE AND OBTURATOR. NO FURTHER DETAILS PROVIDED. IT WAS REPORTED THIS OCCURRED WITH 3 DEVICES. THIS REPORT ADDRESSES ALL 3 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24896215 ↗
  • DeathFeb 5, 2026Difficult to Remove

    IT WAS REPORTED BY THE CUSTOMER THAT IV ACCESS WAS ATTEMPTED WITH NO SUCCESS. IO ESTABLISHMENT WAS ATTEMPTED. NUMEROUS ATTEMPTS WERE MADE TO REMOVE THE STYLET FROM THE CATHETER ONCE ESTABLISHED IN THE PATIENT¿S PROXIMAL TIBIA WITH NO SUCCESS. THE PROVIDERS WERE UNABLE TO DISLODGE THE STYLET FROM THE CATHETER, CAUSING SIGNIFICANT DELAY AND THE INABILITY TO ADMINISTER APPROPRIATE MEDICATION(S.) THE SIGNIFICANT DELAY OF MEDICATION ADMINISTRATION HINDERED THE CARDIAC ARREST RESUSCITATION ATTEMPT. THE PATIENT FROM THE INCIDENT WAS PRONOUNCED DECEASED AT THE EMERGENCY DEPARTMENT AFTER RESUSCITATION ATTEMPT.

    Patient problems: Unintended Loss of Vascular Access

    MAUDE report 24272431 ↗
  • InjuryMar 17, 2025Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    IT WAS REPORTED THAT PATIENT NEEDED A IO THEY WERE ABLE TO ESTABLISH ONE IN THE FEMUR WITH THE RED 15MM BUT HAD TROUBLE SEPARATING THE NEEDLE AFTER DRILLING. IT WAS REPORTED THIS OCCURRED WITH TWO DEVICES. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21621487 ↗
  • MalfunctionAug 18, 2022Activation Problem; Activation Problem

    IT WAS REPORTED BY THE CUSTOMER "INCIDENT OCCURRED ON NEWBORN BABY IN OPERATING ROOM AFTER MOTHER GAVE BIRTH VIA CAESAREAN SECTION. NEWBORN WAS TRANSFERRED TO A FACILITY." CUSTOMER SHARED THAT A 15MM IO NEEDLE WAS USED. NO OTHER INFORMATION WAS PROVIDED. ADVERSE IMPACT TO PATIENT IS UNKNOWN AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15255396 ↗
  • MalfunctionApr 29, 2022Defective Component; Defective Component

    IT WAS REPORTED "IN 2 INSTANCES AFTER INSERTION INTO THE BONE, THE TROCAR COULD NOT BE REMOVED." THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251637 ↗
  • MalfunctionApr 29, 2022Defective Component; Defective Component

    IT WAS REPORTED "IN 2 INSTANCES AFTER INSERTION INTO THE BONE, THE TROCAR COULD NOT BE REMOVED." THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251631 ↗
  • MalfunctionApr 29, 2022Defective Component; Defective Component

    IT WAS REPORTED AFTER INSERTION INTO THE BONE, THE TROCAR COULD NOT BE REMOVED. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251627 ↗
  • MalfunctionMar 11, 2022Failure to Infuse; Failure to Infuse

    IT WAS REPORTED CANNOT INFUSE THROUGH THE T CONNECTOR PORT WHEN THE SYRINGE IF FULLY THREADED ON THE CONNECTOR. IT WAS STATED THE DEVICE WAS NOT USED ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13736061 ↗

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page