Systems/units containing affected software/SQL recallZ-3332-2026 · Beckman Coulter Mishima K.K. · initiated Aug 3, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Mishima K.K. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

FDA root cause category: Software design

What the company is doing

On 08/03/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via postal mail and email informing customers that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCsmay output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. Customers are instructed to: 1. Be sure to create a new index once or more per day at an appropriate time. Refer to 5.1 Start the System of the IFU to complete this action. 2. Beckman Coulter recommends posting this letter on or near the affected analyzers. Resolution - Beckman Coulter has identified the root cause and will resolve the issue by applying a Software solution to the impacted systems. BEC service teamwill contact customers to schedule the Software solution application, and a Field Service Engineer (FSE) will perform the on-site work. For questions - contact Customer Support Center at: http://www.beckmancoulter.com

Recall details

Recall number
Z-3332-2026
Status
Open, Classified
Initiated by the company
Aug 3, 2026
Posted by the FDA
Oct 1, 2026
Recalling firm
Beckman Coulter Mishima K.K.
Quantity
492 units
Product code
JJE
FDA submissions
K981743
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.

Product description

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Code information (models, lots, serial numbers)

Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates