AU B96692Recalls & reported problems

Beckman Coulter, Inc. · AU480 Chemistry Analyzer · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14987666542965), received between Jan 21, 2021 and Apr 22, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 3, 2026Class IIOpen, Classified

    Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

    Z-3332-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Apr 30, 2025Class IICompleted

    A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.

    Z-1907-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Low Test Results2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Low Test Results

    THE CUSTOMER REPORTED CONTINUED G RESULT FLAG ERRORS ON THEIR AU 480 CLINICAL CHEMISTRY ANALYZER (PN: B96692, SN: (B)(6) AND LOW RESULTS DURING FENTANYL TESTS, WHICH WERE NOT RESOLVED THROUGH DILUTIONS OR ROUTINE MAINTENANCE. THERE WAS NO ADVERSE EVENT, NO HARM, INJURY, EXPOSURE, DEATH, OR CONFIRMED CHANGE/DELAY ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA, AND / OR PATIENT DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955935 ↗
  • MalfunctionApr 8, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED ERRATIC PATIENT RESULTS GENERATED FOR MULTIPLE ANALYTES WHICH INCLUDED UALB (URINE ALBUMIN), ALP (ALKALINE PHOSPHATASE), CALA (CALCIUM), CRE (CREATININE), GGT (GAMMA-GLUTAMYL TRANSFERASE), GLUC (GLUCOSE), TP (TOTAL PROTEIN), UREA, D-DIM (D- DIMER), UA (URIC ACID) AND OBSERVED SAMPLE PROBE DILUTING SAMPLES ON SYSTEM AU480 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS PATIENT RESULTS WERE NOT REPORTED OUT OF LABORATORY. THERE WAS NO CHANGE IN PATIENT TREATMENT IN ASSOCIATION WITH THIS EVENT. QUALITY CONTROL WAS WITHIN THE LABORATORY¿S ESTABLISHED RANGES PRIOR TO EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11638336 ↗
  • MalfunctionJan 21, 2021Low Test Results; Low Test Results

    THE CUSTOMER REPORTED INTERMITTENT LOW GLUCOSE (GLU), CREATININE (CRE), AND ALANINE AMINOTRANSFERASE (ALT) PATIENT RESULTS ON THE AU480 CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO CHANGE IN PATIENT TREATMENT IN ASSOCIATION WITH THIS EVENT. QUALITY CONTROL (QC) WERE WITHIN LABORATORY¿S ESTABLISHED RANGE PRIOR TO AND AFTER THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11206333 ↗

FDA clearance and classification

The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AU B96692

Has the AU B96692 been recalled?

Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 completed by the company but not yet closed by the FDA. The most recent, Z-3332-2026 (Class II), began on Aug 3, 2026: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the AU B96692?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jan 21, 2021 and Apr 22, 2026. In those reports, the most common specific device problems are “Low Test Results” (2) and “Incorrect, Inadequate or Imprecise Result or Readings” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AU B96692?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AU B96692 FDA approved?

Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.

Who makes the AU B96692, and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand AU, model B96692). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices

All Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices →