AU B96693Recalls & reported problems
Beckman Coulter, Inc. · AU480 Chemistry Analyzer (with ISE) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Beckman Coulter AU chemistry analyzer has 326 reports: 314 malfunctions and 12 injuries, with no deaths. Most describe incorrect or erratic patient test results, with recent ones often naming sodium, potassium and chloride. It has 2 recalls, one open and one closed; the open one covers a database error that can stop the instrument software from starting on certain Windows computers. Reports fell to 49 in the last 12 months from 85 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 327
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 327 adverse event reports that list this exact device (DI 14987666542972), received between Apr 20, 2018 and Sep 8, 2026 — classified as 12 injuries, 315 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (140 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 3, 2026Class IIOpen, Classified
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
Z-3332-2026 · Root cause: Software design · Recall details · FDA record ↗
- Apr 30, 2025Class IICompleted
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.
Z-1907-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results140
- High Test Results93
- Incorrect, Inadequate or Imprecise Result or Readings89
- Imprecise or Erratic Results5
- Fire2
- High Readings2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions281
- No Consequences Or Impact To Patient30
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Hypernatremia2
- Insufficient Information2
- No Code Available2
- No Known Impact Or Consequence To Patient2
- Electrolyte Imbalance1
Recent adverse event reports
The 25 most recent of 327 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 8, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50679875 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50166002 ↗MalfunctionJun 3, 2026Imprecise or Erratic Results
THE CUSTOMER REPORTED OF GETTING ERRONEOUS ERRATIC PATIENT RESULTS FOR ION SELECTIVE ELECTRODE (ISE) WHICH INCLUDES SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) WITH LOW ANION GAPS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER (PN B96693, SN (B)(6). THE ERRONEOUS PATIENT RESULTS WERE NOT REPORTED OUT OF LABORATORY. THERE WERE NO INJURIES, DEATHS, OR DELAYS/CHANGES IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER REPEATED PATIENT SAMPLES ON ANOTHER ANALYZER AND OBTAINED RESULTS CONSIDERED CORRECT BY THE CUSTOMER. THE CUSTOMER DID NOT PROVIDE PATIENT DATA AND/OR PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25379677 ↗MalfunctionMay 29, 2026Low Test Results
CUSTOMER REPORTED OF GETTING ERRONEOUS LOW PATIENT RESULTS WITH SHORT SAMPLE ERRORS FOR MULTIPLE ANALYTES WHICH INCLUDED ALBUMIN (ALB), BLOOD UREA NITROGEN, CALCIUM, CARBON DIOXIDE, CHLORIDE, CREATININE, GLUCOSE, POTASSIUM, SODIUM, TOTAL BILIRUBIN, PROTEIN, ALANINE AMINOTRANSFERASE, ALKALINE PHOSPHATASE AND ASPARTATE AMINOTRANSFERASE ON THE AU480 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR PATIENT DEMOGRAPHICS, AND THE EXACT NUMBER OF PATIENTS AFFECTED IS UNKNOWN. THE CUSTOMER ONLY PROVIDED AN EXAMPLE FOR TWO (2) PATIENTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25340755 ↗MalfunctionMay 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE GENERATION OF ONE (1) ERRONEOUS LOW PATIENT RESULT FOR MULTIPLE ASSAYS WHICH INCLUDED ALT (ALANINE AMINOTRANSFERASE), GLU (GLUCOSE), CRE (CREATININE), CK (CREATINE KINASE), NA (SODIUM), K (POTASSIUM) AND CL (CHLORIDE) AFTER PM (PREVENTATIVE MAINTENANCE) WAS PERFORMED INVOLVING THE AU480 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25340494 ↗MalfunctionMay 19, 2026High Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH POTASSIUM AND SODIUM PATIENT RESULTS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25222162 ↗MalfunctionMay 15, 2026Low Test Results
THE CUSTOMER REPORTED ERRONEOUS LOW PATIENT RESULTS GENERATED FOR LIPASE (LIP), BICARBONATE (CO2) AND ION SELECTIVE ELECTRODE (ISE) WHICH INCLUDE SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND LATER AMENDED. THE CUSTOMER DID NOT PROVIDE PATIENT DATA PRINTOUTS OR DEMOGRAPHICS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25194588 ↗MalfunctionMay 13, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW SODIUM AND CHLORIDE PATIENT RESULTS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE DEMOGRAPHICS. DATA WAS PROVIDED FOR ONE PATIENT SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25162943 ↗MalfunctionMay 9, 2026Low Test Results
CUSTOMER REPORTED OF GETTING ERRONEOUS PATIENT RESULTS FOR SODIUM AND POTASSIUM INVOLVING THEIR AU480 CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25131647 ↗MalfunctionApr 24, 2026Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED OF GETTING ERRONEOUS PATIENT RESULTS FOR SODIUM, POTASSIUM AND CHLORIDE ON THEIR AU480 CHEMISTRY ANALYZER. THE ERRONEOUS PATIENT RESULTS WERE NOT REPORTED OUT OF LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24980921 ↗MalfunctionApr 23, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER QUESTIONED AN ACETAMINOPHEN RESULT WITH '*' FLAG FOR ONE PATIENT SAMPLE GENERATED ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24971637 ↗MalfunctionApr 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED OBTAINING ERRATIC PATIENT RESULTS ON ION SELECTIVE ELECTRODE (ISE) INCLUDING SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24918134 ↗MalfunctionApr 3, 2026Low Test Results
A CUSTOMER REPORTED THAT THEIR AU480 CLINICAL CHEMISTRY ANALYZER PRODUCED LOW PATIENT RESULTS FOR CREATININE, SODIUM, AND POTASSIUM. THE SAME SAMPLE WAS RE-TESTED AND YIELDED NORMAL RESULTS. ALTHOUGH THE INITIAL LOW RESULTS WERE RELEASED, THEY WERE SUBSEQUENTLY AMENDED AFTER THE REPEAT TESTING. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24778625 ↗MalfunctionMar 31, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE AU480 CLINICAL CHEMISTRY ANALYZER GENERATED FOUR (4) ERRONEOUS ERRATIC PATIENT RESULTS FOR ION SELECTIVE ELECTRODE (ISE) WHICH INCLUDES SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) WITH LOW ANION GAPS AND ISE DRAIN OVERFLOWING. THE CUSTOMER DID NOT PROVIDE PATIENT PRINTOUT OR DEMOGRAPHICS. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746357 ↗MalfunctionMar 26, 2026High Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH ION-SELECTIVE ELECTROLYTES (ISE) PATIENT RESULTS, INCLUDING SODIUM (NA), CHLORIDE (CL) AND POTASSIUM (K), ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. DATA WAS PROVIDED FOR THREE (3) PATIENT SAMPLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24700463 ↗MalfunctionMar 25, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS WITH MULTIPLE FLAGS FOR SODIUM (NA), POTASSIUM (K), CHLORIDE (CL), BLOOD UREA NITROGEN (BUN), CREATININE (CREA) AND C-REACTIVE PROTEIN (CRP) ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THE SAMPLE WAS REPEATED WITH NORMAL RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT SHARED THE RESULTS WITH ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24693407 ↗MalfunctionMar 25, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR SODIUM AND TOTAL PROTEIN ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691514 ↗MalfunctionMar 23, 2026Low Test Results
THE CUSTOMER REPORTED GENERATION OF ERRONEOUS LOW SODIUM (NA), AND CHLORIDE (CL) PATIENT RESULTS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER (PN B96693, SN (B)(6). THERE WAS NO INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE ANY PATIENT DATA AND/OR PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663936 ↗MalfunctionMar 20, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER QUESTIONED PATIENT RESULTS FOR SODIUM (NA), CHLORIDE, POTASSIUM (K) AND CO2 RESULTS DUE TO LOW ANION GAP RESULTS INVOLVING THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS, OR DATA AND THE EXACT NUMBER OF PATIENTS AFFECTED IS UNKNOWN. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654617 ↗MalfunctionMar 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED GETTING ERRATIC PATIENT RESULTS FOR ION SELECTIVE ELECTRODE (ISE) INCLUDING SODIUM, POTASSIUM AND CHLORIDE INVOLVING THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24613132 ↗MalfunctionMar 11, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE GENERATION OF ERRATIC PATIENT RESULTS FOR MULTIPLE ANALYTESM INCLUDING, LIPID PANELS AND CMP (COMPREHENSIVE METABOLIC PANEL) WITH CALIBRATION ISSUES, RB (REAGENT BLANK) DATA ERROR, INCORRECT CUVETTE ERRORS AND OTHER MAINTENANCE ALERTS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE CUSTOMER PROVIDED DATA FOR ONE PATIENT SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24571486 ↗MalfunctionMar 9, 2026Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED OF GETTING ERRATIC PATIENT RESULTS FOR SODIUM AND CHLORIDE INVOLVING THE AU480 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND LATER CORRECTED. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24550822 ↗MalfunctionMar 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED A SAMPLE PROBE ERROR ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. AS RESULT, FIVE PATIENTS HAD ERRATIC RESULTS ON ION SELECTIVE ELECTRODE (ISE) INCLUDING SODIUM, POTASSIUM AND CHLORIDE. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24503775 ↗MalfunctionFeb 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED ERRATIC ERRONEOUS PATIENT RESULTS ON MULTIPLE ASSAYS WHICH INCLUDES UREA NITROGEN (UREA), CREATININE (CRE) AND SODIUM (NA), INVOLVING THE AU480 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442312 ↗MalfunctionFeb 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED OF GETTING ERRONEOUS ERRATIC PATIENT RESULTS FOR CARBON DIOXIDE (CO2), SODIUM (NA), AND CHLORIDE (CL) WITH LOW ANION GAPS ON THEIR AU480 CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS, AND THE EXACT NUMBER OF PATIENTS AFFECTED IS UNKNOWN. THE CUSTOMER PROVIDED A VERBAL EXAMPLE OF FALSE RESULTS FOR ONE (1) PATIENT. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24388530 ↗
FDA clearance and classification
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the AU B96693
Has the AU B96693 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 completed by the company but not yet closed by the FDA. The most recent, Z-3332-2026 (Class II), began on Aug 3, 2026: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the AU B96693?
The FDA's MAUDE database lists 327 adverse event reports for this device, received between Apr 20, 2018 and Sep 8, 2026. In those reports, the most common specific device problems are “Low Test Results” (140), “High Test Results” (93) and “Incorrect, Inadequate or Imprecise Result or Readings” (89); the most common patient problems are “Hypernatremia” (2) and “Electrolyte Imbalance” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AU B96693?
None of the reports are classified as deaths, 12 as injuries and 315 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AU B96693 FDA approved?
Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.
Who makes the AU B96693, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand AU, model B96693). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls677 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death645 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE680 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page