Skip to content

CARESCAPE Central Station recallZ-3143-2026 · GE Medical Systems Information Technologies Inc · initiated Jul 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems Information Technologies Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

FDA root cause category: Under Investigation by firm

What the company is doing

On July 30, 2026, the firm sent Urgent Medical Device Correction letters to customers with affected units. This letter informs the customer of the issue and provides actions to be taken by the customer/user for continued use of the device. GE HealthCare will provide a User Manual Addendum that provides guidance regarding the prolonged startup condition.

Recall details

Recall number
Z-3143-2026
Status
Open, Classified
Initiated by the company
Jul 30, 2026
Posted by the FDA
Sep 11, 2026
Recalling firm
GE Medical Systems Information Technologies Inc
Quantity
29,570 (23,632 US; 5,938 OUS)
Product code
DQK
FDA submissions
K162012
Distribution
Worldwide - US Nationwide distribution.

Product description

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Code information (models, lots, serial numbers)

5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K162012 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page