Midmark IQvitals Zone, automatically and noninvasively… recallZ-3217-2026 · Midmark Corporation · initiated Aug 21, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Midmark Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.

FDA root cause category: Process change control

What the company is doing

An Urgent Medical Device Recall notification letter dated 8/19/26 was sent to customers. Safety Instructions Until Midmark provides further instructions regarding this recall, use only the Step Deflation Method when obtaining blood pressure measurements. Do not use the Linear Deflation Method. Follow the instructions in Appendix A to verify that Step Deflation is selected before obtaining blood pressure measurements. Please complete, sign and return the Customer Response Form after completing the instructions described in Appendix A of this notification. The affected device(s) may continue to be used for blood pressure measurements provided this safety instruction is followed. Ensure all users of the affected device(s) are informed of this safety instruction. Product Correction Midmark will contact your facility with the next steps required to complete this recall. Additional instructions will be provided at that time. No immediate action is required to contact Midmark at this time. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email [email protected]. Required Distributor Actions 1. Immediately send the attached customer-facing "Urgent Medical Device Recall - IQvitals Zone Linear Deflation Blood Pressure Performance" notification to every impacted end customer to whom you supplied an affected device. Do not wait for Midmark to contact the customer before forwarding the attached safety notification. 2. If you supplied an affected device to another distributor, send that distributor both this first-consignee communication and the attached customer notification. Instruct the downstream distributor to immediately forward only the customer notification to each impacted end customer and provide the final consignee information to you for submission to Midmark. 3. Immediately provide Midmark wi

Recall details

Recall number
Z-3217-2026
Status
Open, Classified
Initiated by the company
Aug 21, 2026
Posted by the FDA
Sep 18, 2026
Recalling firm
Midmark Corporation
Quantity
1 units
Product code
MWI
FDA submissions
K161909
Distribution
US Nationwide distribution.

Product description

Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0320 with Kit Number: 4-000-0622

Code information (models, lots, serial numbers)

Model No: 1-200-0320; UDI-DI: 00841709105791; Kit No: 4-000-0622; UDI-DI: 00841709105876; Kit Serial Number: V2900924; Zone Serial Number: V2899738

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

6 devices cleared under K161909 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99641).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page