Midmark IQvitals Zone, automatically and noninvasively… recallZ-3220-2026 · Midmark Corporation · initiated Aug 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Midmark Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
FDA root cause category: Process change control
What the company is doing
An Urgent Medical Device Recall notification letter dated 8/19/26 was sent to customers. Safety Instructions Until Midmark provides further instructions regarding this recall, use only the Step Deflation Method when obtaining blood pressure measurements. Do not use the Linear Deflation Method. Follow the instructions in Appendix A to verify that Step Deflation is selected before obtaining blood pressure measurements. Please complete, sign and return the Customer Response Form after completing the instructions described in Appendix A of this notification. The affected device(s) may continue to be used for blood pressure measurements provided this safety instruction is followed. Ensure all users of the affected device(s) are informed of this safety instruction. Product Correction Midmark will contact your facility with the next steps required to complete this recall. Additional instructions will be provided at that time. No immediate action is required to contact Midmark at this time. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email [email protected]. Required Distributor Actions 1. Immediately send the attached customer-facing "Urgent Medical Device Recall - IQvitals Zone Linear Deflation Blood Pressure Performance" notification to every impacted end customer to whom you supplied an affected device. Do not wait for Midmark to contact the customer before forwarding the attached safety notification. 2. If you supplied an affected device to another distributor, send that distributor both this first-consignee communication and the attached customer notification. Instruct the downstream distributor to immediately forward only the customer notification to each impacted end customer and provide the final consignee information to you for submission to Midmark. 3. Immediately provide Midmark wi
Recall details
- Recall number
- Z-3220-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 21, 2026
- Posted by the FDA
- Sep 18, 2026
- Recalling firm
- Midmark Corporation
- Quantity
- 33 units
- Product code
- MWI
- FDA submissions
- K161909
- Distribution
- US Nationwide distribution.
Product description
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0360 with Kit Number: 4-000-0724
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model No: 1-200-0360; UDI-DI: 00841709105838; Kit No: 4-000-0724; UDI-DI: 00841709105913; Kit Serial Number: V2788703 V2788046 V2788016 V2788712 V2789302 V2789846 V2789830 V2789837 V2789808 V2789303 V2789304 V2811103 V2811112 V2862288 V2862275 V2869139 V2866577 V2867451 V2869599 V2868801 V2869154 V2868545 V2866984 V2869980 V2869969 V2868130 V2867433 V2868182 V2924525 V2924566 V2924559 V2924539 V2927525; Zone Serial Number: V2788322 V2787552 V2787555 V2788321 V2788905 V2789498 V2787554 V2789488 V2789499 V2788922 V2787553 V2807891 V2807893 V2837260 V2859050 V2868888 V2862992 V2866942 V2868893 V2868239 V2868840 V2868222 V2866940 V2868892 V2869397 V2867491 V2866963 V2868051 V2923639 V2922999 V2922998 V2923640 V2927201
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Midmark IQvitals Zone, w/Masimo SPO2 and WirelessMidmark Corporation · Oximeter1 recall0 reports
- Midmark IQvitals Zone, Wireless with Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K161909 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Midmark IQvitals Zone w/Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall1 report
- Midmark IQvitals Zone, w/Nellcor SPO2Midmark Corporation · Monitor, physiological, patient (without arrhythmia detection or alarms)1 recall0 reports
- Midmark IQvitals Zone, w/Nellcor SPO2 and WirlessMidmark Corporation · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE1 recall0 reports
- Midmark IQvitals Zone, Non-Wireless with Nellcor SPO2Midmark Corporation · Oximeter1 recall0 reports
- Midmark IQvitals Zone, Non-Wireless with Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall0 reports
- Midmark IQvitals Zone, Wireless with Nellcor SPO2Midmark Corporation · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99641).
- Z-3217-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
- Z-3218-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
- Z-3219-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page